Gel Warmer Test; Single Cwd; Mode Transition Checks; General Information - GE LOGIQ P9 Service Manual

Hide thumbs Also See for LOGIQ P9:
Table of Contents

Advertisement

GEDRAFT
D
5604324, R
IRECTION
EVISION
4-3-6-6

Gel Warmer Test

1) Insert Gel Bottle into Gel Warmer.
2) Power ON the warmer.
3) Verify Gel gets warm.
NOTE:
Depending on environment, Gel requres 30 minute or more to warm up.
4-3-6-7

Single CWD

1) Connect P8D probe. Select vascular application like "Carotid". And start scan.
2) Verify that CW timeline image appears.
4-3-7

Mode Transition Checks

4-3-7-1

General Information

1) [R1.X.X] Select UTILITY -> SYSTEM -> User Configurable Key.
[R2.X.X or later] Select UTILITY -> APPLICATION -> Settings.
2) [R1.X.X] Select "M" in 2 list of "Mode Rotary Line Key".
[R2.X.X or later] Select "M" in 2 list of "User Configurable Mode Rotary".
3) [R1.X.X] Select "PW" in 3 list of "Mode Rotary Line Key".
[R2.X.X or later] Select "PW" in 3 list of "User Configurable Mode Rotary".
4) [R1.X.X] Select "CF" in 5 list of "Mode Rotary Line Key".
[R2.X.X or later] Select "CF" in 5 list of "User Configurable Mode Rotary".
5) Press "Save".
6) Select UTILITY -> SYSTEM -> User Configurable Key.
7) Select "B-Flow" in 4 list of "User Defined Hard Key".
8) Select "Reset" in 3 list of "User Defined Touch Button".
11
DRAFT (J
Chapter 4 - Functional Checks
24, 2019)
ANUARY
LOGIQ P9/P7
S
M
ERVICE
ANUAL
4-24