Sensor specifications
Sensor
Pressure
Flow
(High Flow Kit)
Flow (Low Flow Kit)
Conductivity
pH
UV
Temperature
1
2
3
4
5
6
7
8
ÄKTA ready Operating Instructions 28960345 AD
Range
-1 to 10 bar g
0 to 510 l/h
2
0 to 175 l/h
4
0 to 199.99 mS/cm
6
pH 0 to 14
-7 to 7 AU
0 to 47°C
Max. error, valid within acceptance range under operation conditions.
Flow meter cells for High Flow Kit and Low Flow Kit are designed with different flow path ge-
ometries
Valid for max. temperature difference of 10 ºC between liquid and ambient temperature.
Temperature compensation optional. The standard setting of the cell constant for the sensor
is 12.5 cm -1 and is not to be changed when replacing flow kits.
Valid at actual temperature (no temperature compensation).
Temperature compensation not available.
Cell length 1.8 mm ±0.2 mm.
Cell length 1.8 mm ±0.2 mm.
14 Reference information
Acceptance range
0 to 5 bar g
7.5 to 510 l/h
3 to 175 l/h
0 to 150 mS/cm
pH 2 to 12
0.01 to 1.0 AU
2ºC to 40ºC
14.1 Specifications
Max. measure-
1
ment error
±0.20 bar g
±5% actual value
3
at flow ≥ 40 l/h
±10% actual value
at flow < 40 l/h
±5% actual value
at flow ≥ 20 l/h
±10% actual value
at flow < 20 l/h
5
±5% full scale
±0.2 pH at calibra-
tion temperature
Drift: ±0.025 pH
units/h (20ºC, pH 4)
Flow rate sensitivi-
ty: ±0.1 pH unit
Deviation from lin-
, 7 , 8
earity ±5%
±3ºC
205