15. Alarms Overview

The RigRat provides audible and visible alarm notification system, plus it combines local alarms on the
device with real-time remote wireless alarm notification to enhance worker safety. Local alarms include
audible buzzer alarm, visible alarm via bright LED lights, and an alarm notification on the display. The
audible alarm can be programmed or selectively turned on or off. The LED alarm indicators cannot be
turned off.

Alarm Signals

During each measurement period, the gas concentration is compared with the programmed alarm
limits for Low, High, TWA and STEL alarm. If the concentration exceeds any of the preset limits, the
alarms are activated immediately to warn of the alarm condition.
In addition, the RigRat alarms if the battery voltage is low.
When the low battery alarm occurs, it is recommended that you promptly charge the battery in a non-
hazardous location.

Alarm Signal Summary

Normal Operation Mode
Type
Red LED
Over Range
5 beeps/s
High
3 beeps/s
Low
2 beeps/s
STEL
1 beep/s
TWA
1 beep/s
Unit Failure
Fail
Negative
PID Lamp
Remote Alarm
External Alarm
Pump
Battery critical
BW RigRat User's Guide
Yellow LED
Green
LED
600ms/s
400ms/s
200ms/s
200ms/s
200ms/s
3 beeps/s
200ms/s
2 beeps/s
200ms/s
1 beeps/s
200ms/s
1 beeps/s
200ms/s
1 beeps/s
200ms/s
1 beeps/s
200ms/s
1 beeps/s
200ms/s
1 beeps/s
200ms/s
Buzzer
LCD Display
"+OL"
High alarm icon
and reading
Low alarm icon
and reading
STEL alarm icon
and reading
TWA alarm icon
and reading
"FAIL"
"-OL"
"LAMP"
Units in alarm in
network detail
screen
External signal
state in
"Input/output
state" screen
Blink pump icon
Battery critical
icon
62
Reading
Notes
Blink red
Blink red
Blink red
Blink red
Blink red
Instrument fatal
error
Blink red
Sensor generic
fail
Blink red
Blink red
PID lamp fail
Remote device
alarm in closed
loop network,
4~20mA input,
digital input
Battery voltage
less than 3 to
3.5V for 4s, unit
will power off in
15 minutes.
The voltage