Switch

RigRat supports three separate SPST PhotoMOS relay output switch connections for applications
where other equipment needs to be controlled (lights, sirens, etc.).
Relay Definitions:
Relay 1
Any gas sensor failed, or STEL/TWA/+OL/-OL alarms
Relay 2
Any high alarms
Relay 3
Any low alarms
IMPORTANT! This is a fixed configuration that can only be modified via Device Configurator app.
Note: Driving high current or voltage through the switch will damage the product.
Note: The default option is NO (normally open). It can be set to NC (normally closed) via the Device
Configurator app. Note that NC consumes more battery current.
PhotoMOS relay output switch parameters
Must meet IS parameters if used in hazardous area!
Item
Voltage
Max power consumption on switch
Continuous load current
On resistance
BW RigRat User's Guide
Value
30V
AC/DC
0.75W
100mA
1Ω typical