MAN0905_Issue 7_07/17
Searchpoint Optima Plus with optional HART
output
®
Operating Instructions
2108M0550
10. Maintenance using
SHC1 Handheld Interrogator
10.8 Investigation of Faults or Warnings
If the unit is signalling Fault or Warning on the 4-20 mA output, the problem should be investigated and
remedied as soon as possible. Some types of faults and warnings are latching and will need to be cleared
by a Reset.
1. Select Diagnose then Active Faults, or Diagnose then Active Warnings to view any faults or warnings.
Note: If no active faults or warnings are present, the 4-20 mA output signal may be due to a latched fault or
warning. Proceed to step 3.
2. Use the Problem Solving section of this manual (Chapter 14) to rectify the issue.
3. Select Diagnose then Soft Reset to clear any latched warnings or faults.
Note: If any configuration changes have been made, wait 15 seconds before doing a soft reset or cycling
the power.
4. Press Escape to return to the main menu.
47