Honeywell Searchpoint Optima Plus Operating Instructions Manual page 17

With optional hart output
Hide thumbs Also See for Searchpoint Optima Plus:
Table of Contents

Advertisement

MAN0905_Issue 7_07/17
European Duct Mounting Kit – Metric Threads (2104B0349)
Junction
Box
Junction
Box
Bracket
Duct
Mounting
Plate
US Duct Mounting Kit – NPT Threads (2442-0016)
This kit enables ¾ NPT versions of Searchpoint Optima Plus to be conveniently mounted inside a duct.
Electrical connections are made to the detector via a junction box which is mounted outside of the duct.
Test gas can be applied to the detector via the gassing points provided.
Note: the detector must have the Remote Gassing Cell option factory fitted.
2. Introduction
This kit enables M25 versions of Searchpoint
Gassing
Optima Plus to be conveniently mounted inside
Points
a duct. Electrical connections are made to the
detector via a junction box which is mounted
Searchpoint
outside of the duct. Test gas can be applied to the
Optima
detector via the gassing points provided.
Note: the detector must have the Remote Gassing
Cell option factory fitted.
17
Searchpoint Optima Plus with optional HART
output
®
Operating Instructions
2108M0550

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents