12. Commissioning using
Read this chapter if HART
Searchpoint Optima Plus.
Note: The instructions that follow assume that a suitable HART
Searchpoint Optima Plus and that the user is logged in with Level 1 access privileges. See Chapter 11 "Use
of the HART
Communications" for details.
®

12.1 Introduction

This chapter explains several operations that may be required during commissioning. The user should work
through First Time Switch On, User Configuration and Configure Device Identification Information, before
selecting further operations as required by the specific installation.
First Time Switch On
User Configuration
Configure Device Identification Information
Fix the 4-20 mA Output
Test 4-20 mA Loop Integrity
Calibrate the 4-20 mA Loop
Configure Internal Alarm Threshold
Simulation
Clear Faults or Warnings
Caution: After changing any parameter, poll the Searchpoint Optima Plus and verify that the received
value is correct.
Note: The instructions that follow assume that a suitable HART
Searchpoint Optima Plus and that the user is logged in with Level 1 access privileges. See Chapter 11 "Use
of the HART
Communications" for details.
®
HART
Communications
®
communications are being used to communicate with
®
Checks that are required on first power up
How to change settings from the default values: -
Change the Password – How to change the password for Level 1 access
Configure HART
output level to 1 mA so that HART
Set the Time and Date - How to adjust the real time clock to local time
Configure Warning, Inhibit and Overrange Signal Levels - How to
change the signal levels from the default settings
How to set text fields that may be useful for device identification
How to force the mA output to set levels in order to calibrate the controller
or test the system
How to check that there is sufficient compliance in the loop to signal alarm
and overrange
How to calibrate the current output (note this is not a gas calibration)
How to change the internal threshold used to simulate Alarm and log gas
events
How to test the system by simulating Alarm, Warning and Fault conditions
How to clear faults or warnings that may be generated during
commissioning
Searchpoint Optima Plus with optional HART
host (handheld or PC) is connected to the
®
Operation during Fault - How to change the Fault
®
will operate during Fault
®
host (handheld or PC) is connected to the
®
52
output
®
Operating Instructions
2108M0550