Reassembling the Battery Charger and Conditioner

Step 8

Removing the Air Filters

The air filters are located on the left and right side of the device.
Step 1
Step 2
Air Filter
Step 3
Step 4
Step 5
Step 6
Reassembling the Battery Charger and Conditioner
Reassembly is the reverse of the disassembly procedure.

Following Reassembly

Once you have reassembled the Instrument, you must perform a safety and performance
check on the Instrument. Refer to "Maintaining the Instrument" and "Testing the Instrument".
242
Remove the right-hand cooling fan.
Note: When reconnecting the two-pin plug, ensure that the red wire faces the heat
sink.
Open the housing (see "Opening the Housing" on page 239).
Remove the plug from the five-pin connector on the circuit board. This allows
better access to the inside of the charger.
Remove the two screws holding the air filter.
Remove the air filter cover.
Remove the filter mat and clean the dust out, or exchange the filter mat if required.
Insert the mat and re-fit the cover and screws.
Air Filter
Repairing the Instrument