Test Map

Test Map
The test map shows which tests are required in which situations.
Service Event
(When performing...)
Installation of M3/M4 with no display
connected to the VGA output
Installation of M3/M4 with a display
connected to the VGA output
Installation of M3/M4 with a recorder
connected to the serial interface
Repairs where the power supply in the M3/
M4 is replaced
Repairs where the monitor has been
dropped
All other M3046A repairs, Hardware or
Software Upgrades
Preventative Maintenance
Safety and Performance
140
Test Blocks Required
......Complete these tests)
Perform Visual and Power On test
blocks
Perform Visual, Power On and
Safety (1) test blocks
Perform Visual, Power On and
Safety (1) test blocks. Perform test
print on recorder.
Perform Power On and Safety (2)
Test blocks
Perform Power On and Safety (2)
and (3) Test blocks
Perform Power On test block
Perform all Preventative
Maintenance Tests
Perform all Safety and Performance
Test blocks.