Patient Safety Checks

S(1) Part 1: System Enclosure Leakage Current - NC (normal condition)

L (N)
N (L)
Expected test results:
Measures leakage current of exposed metal parts of Instrument under Test (IUT) and between
parts of the system within the patient environment; normal and reversed polarity using S2.
Safety test according IEC 60601-1 / UL2601-1

S(1) Part 2: System Enclosure Leakage current - Single Fault (open earth)

L (N )
S 1
N (L )
S 4
160
S2
S1
S4
(*)
PE
MD
Normal condition maximum leakage current x1 ≤ 100µA
S 2
P E
M D
Medical electrical system
Instrument under test
(*) Not present in Class 2.
(**) Can be multiple different connections to different equipment at same time.