Removing The Connector Board - Philips M3001A Service Manual

M3/m4 monitors,measurement server,measurement server extensions
Table of Contents

Advertisement

Disassembly for the Monitor
Step 7
Step 8

Removing the Connector Board

To remove the Connector Board, do the following:
Step 1
Step 2
Step 3
228
Press down the retaining lever to release the backlight tube:
Press Lever Down to Release
Slide out the tube and repeat for the second tube.
Remove the chassis as described previously in this chapter.
Remove the Power Supply as described previously in this chapter.
The Connector Board (VGA) is fixed with a screw and can no longer be taken out
by removing the Connector Board cover and pulling the Connector Board out. The
Plastic Housing must be completely removed to allow the Connector Board to be
unscrewed from the side of the Monitor:
Backlight Tube
Repairing the Instrument