Removing The Co 2 Scrubber - Philips M3001A Service Manual

M3/m4 monitors,measurement server,measurement server extensions
Table of Contents

Advertisement

Removing the CO
To remove the CO
Step 1
Step 2
Step 3
Step 4
Repairing the Instrument
Disassembly Procedures for the M3015A Measurement Server Extension
Scrubber
2
Scrubber, do the following:
2
Locate the CO
Scrubber in the Extension.
2
Being careful not to touch anything else in the Extension, use tweezers to pull the
body of the CO
Scrubber out of the bracket
2
.
Holding the body of the CO
Extension intake tube from the scrubber end and remove the CO
Extension.
Dispose of the CO
Scrubber according to local legal requirements for low volume
2
chemical waste.
Scrubber with your fingers, carefully disconnect the
2
Scrubber from the
2
235

Advertisement

Table of Contents
loading

This manual is also suitable for:

M3000aM3046aM3015aM3016a

Table of Contents