Restoring The Factory Settings - Miele G 7883 Operating Instructions Manual

Laboratory glassware washer
Table of Contents

Advertisement

^ Turn the program selector to "STOP"
f .
^ Turn the machine off.
^ Press and hold "Display" ß and
"Start" 6.
lab washer on with the "On/Off"
button.
The displays shows the current
program status "P...".
The "Fill/Drain" indicator lights.
^ Press and hold "Drying" 0 until "E33"
appears in the display.
^ Turn the program selector to any
position (except "STOP").
The display shows either "00" or "PP".
"00" = factory setting; electronic was
not reprogrammed.
"PP" = electronic was reprogrammed.
^ If "PP" is displayed press "Display" ß
to change from "PP" to "00".
^ Press the "Start" 6 button. "SP"
appears in the display.
^ Press the "Start" 6 button again.
All factory settings are now restored.
42
If the hardness level setting in the
water softener had been changed,
this will reset to the factory setting of
340 ppm (19 gr/gal). This will need
to be reset to suit the hardness level
in your area.
turn the
Changes in programs "Custom",
"Inorganica" and "Organica" cannot
be reset.