For future servicing, please
document any alterations to the
standard factory settings.
Program selector positions that have
not been allocated to a function are
indicated in the middle part of the
display by a bar (-).
The programming level is called up via
the "Stop" f position.
34
Before performing any of the special
functions that follow, the programming
mode of the labwasher controller needs
to be accessed.
^ Turn the program selector to "Stop"
f.
^ Turn the machine off.
^ Press and hold the "Display" ß and
"Start" 6 buttons. At the same time,
turn the machine on with the "On/Off"
button.
^ The display will show the current
program status "P...".
The "Fill/Drain" indicator will
illuminate.
^ Press 0 twice. "E02" (programming
level 2) appears in the display.