Chapter 9: Troubleshooting
351 - Software Invalid State Error
Message: "Software invalid state error. Please power cycle the Analyzer. If the problem persists, export errors
and contact GE Analytical Instruments for further instructions and analysis."
Explanation: There are a variety of causes for this error. This error should be very rare. Analysis of the exported
errors will give the specific line of code that triggered the error in order to suggest the resolution.
6401 - Operating System Error
Message: "Operating system error. Please power cycle the Analyzer. If the problem persists, export errors and
contact GE Analytical Instruments for further instructions and analysis."
Explanation: This is a "catch all" for any errors not specifically defined elsewhere. These should be very rare and
would relate to threads or potential communication issues. There could be a large number of causes that would
be difficult to address in this section. Analysis of the exported errors will give the specific line of code that
triggered the error in order to suggest the resolution.
6402 - Peristaltic Pump Life Error
Message: "The peristaltic pumps have exceeded their recommended life. Replace P1-SH and P1-R."
Explanation: Based on the history of usage, the Analyzer estimates that the life of the peristaltic pump in the
Sample Handling and Reactor modules has expired. Order two new peristaltic pumps from GE now.
6403 - Tubing Life Error
Message: "The tubing has exceeded its recommended life. Replace the tubing in all pinch valves."
Explanation: Based on the history of usage, the Analyzer estimates that the pinch valve tubing life has expired.
Order tubing from GE now to avoid down time and more costly failure of pinch valves.
6405 - Peristaltic Pump Life Warning
Message: "The estimated pump life is less than 2 weeks. Order new peristaltic pumps to replace P1-SH and P1-R."
Explanation: Based on the history of usage, the Analyzer estimates that only two weeks of peristaltic pump life
remains. Order two new peristaltic pumps from GE now to avoid down time when the pumps expire.
6406 - Tubing Life Warning
Message: "The estimated tubing life is less than 2 weeks. Order new tubing."
Explanation: Based on the history of usage, the Analyzer estimates that only two weeks of pinch valve tubing
remains. The tubing must be replaced to protect the operation of the pinch valves. Order tubing from GE now to
avoid down time and more costly failures of pinch valves.