Chapter 6: Analyzer Operation
Maintenance — Display information about the Analyzer's consumables levels, system information, and
error messages. (Refer to Chapter 8, "Maintenance" for more information.)
Cal/Ver— Calibrate the Analyzer and perform calibration verification. Also, view past calibration and
verification data. (Refer to Chapter 5, "Calibration and Verification" for more information.)
You can also press the Start Analysis button, located on the lower portion of the screen, to initiate
measurements. This button is also visible from any other tab.

Setting Up Measurements (The Setup Tab)

Use the Setup tab to define the measurement settings for the Analyzer. The Setup tab includes the following
options:
Online — Set up a selected number of online measurements on each (one or more) enabled sample
streams and the check standard (if it is enabled).
Check Standard — Set the Analyzer to check the performance of the Analyzer by performing one or
more number of measurements on the check standard, and then stopping.
Grab — Select up the number of measurements (up to 20) for the sample in the vial port. The Analyzer
will then calculate the average of those measurements, and also report precision (Relative Standard
Deviation) of those measurements.
Rinse — Set the Analyzer to make measurements from the Check Standard port until the operator stops
the analysis. This function is designed to allow the user to rinse down the Analyzer by placing the check
standard tube in low-TOC water. When operating in Rinse mode, the measurements are not stored in
the Analyzer's Data History in order to ensure that this data is not confused with measurements of real
samples.