Record and assess the highest value determined in the repeat measurement in the exist-
ing report 'Device leakage current / report'.
This value must not exceed the permissible leakage current values according to
DIN VDE 0751-1:2001-10, Table F.1, line "Device leakage current for devices according to
remarks 1 and 3", of 2.5 mA.
Measure and record the current line voltage. If the measured line voltage deviates from
the nominal voltage, then correct the measured value to the value corresponding to a
measurement at the nominal value of the line voltage. Document this in addition.
Document the measuring procedure (differential measurement or direct measurement)
and the measuring instrument used (designation and serial number).
NOTE

Replacement for damaged or lost screws

Damaged or lost screws may be replaced only with screw types that meet specifications,
made of steel per DIN 267 and that have the specified hardness rating.
CAUTION

Cleaning

Prior to cleaning the unit, the unit must always be switched off or disconnected from line
power.
Clean the unit with just water or with a lukewarm, aqueous solution of water and a
common household cleaning agent.
Do not use souring powders and no organic solvents or cleaning agents, e.g. cleaning
fluid, alcohol, spot remover, etc.
Water spraying is not permitted.
For further information, see the operator manual "Cleaning/Disinfecting".
POLYMOBIL III
Evaluate the results by comparing the first measured value to the
corresponding values documented during preceding mainte-
nance procedures or safety checks.
A sudden or unexpected increase of the measured values may
indicate that a fault has occurred in the primary circuit of the
power supply (damaged insulation, damage caused by humidity,
defective interference suppressor, etc.) - even if the limit value of
2.5 mA is not exceeded.
Replacement of screws.
All Allen screws used must basically have the strength class 8.8.