Functional checks
Fig. 3
Checking the control elements on the double slot diaphragm.
•
Check the light localizer in each case at the control console and at the double slot
diaphragm.
To do this press the button
±
20 s
•
Check the turning ability of the double slot diaphragm
•
By turning the diaphragm setting knobs (8/Fig. 3) check the easy movement and
uniformity of the format setting (leaf pairs).
Checking the exposure release with high-voltage
•
Close the double slot diaphragm.
Set 60 kV, 10 mAs.
•
Pull out the exposure release switch (9/Fig. 4) completely.
•
Release exposure.
The radiation indicator (10/Fig. 4) lights up for the duration of the exposure, an acoustic
signal sounds at the same time.
•
Check that the cable of the exposure release switch (9/Fig. 4) rolls back in without
difficulty after it is pulled out.
Siemens AG
Medical Solutions
7
8
1s.
Fig. 4
(6/Fig. 4 or 7/Fig. 3); the light localizer lamp lights up for
RXB8-115.034.02
Rev. 03
04.05
6
10
5
±
45°.
Page 3 of 6
CS PS 24
4 - 3
9
4
4
POLYMOBIL III