4 Operation
4.3 Intelligent packing
Perform the packing
Follow the steps below to perform the packing. Use the UNICORN method created above.
Step
1
2
3
80
Action
Connect position 1 or 2 according to the method on the Column valve
module to the top (on the adaptor) and bottom of the AxiChrom column.
Connect the IP port to the horizontal nipple on the column top (not on the
adaptor). See
Illustration, on page
nections to position 1.
WARNING
Before connecting a column, read the instructions for
use of the column.
Tip:
In the Intelligent Packing phase of your packing method, click on Show
Column Connection to find an illustration of the connections.
Run your packing method in the System Control module, follow the instruc-
tions in dialog boxes on the screen.
Note:
Make sure that you press Continue when the adapter meets the consolidated
bed surface. If Continue is not pressed in time, the resin can be damaged,
and the delta column pressure alarm pauses the packing method.
Evaluate and save the packing of the column in the Evaluation module, see
the Getting started page in the Evaluation module, or the UNICORN Evalu-
ation Manual.
77. Note that the illustration shows con-
ÄKTA pilot 600 User Manual 29274559 AA