GE AKTA pilot 600 User Manual page 105

Table of Contents

Advertisement

Step
1
2
3
4
5
6
7
ÄKTA pilot 600 User Manual 29274559 AA
Action
Clean the conductivity flow cell before calibration (see
the conductivity flow cell, on page
Disconnect the tubing from the bottom of the conductivity monitor and re-
place it with a Luer-to-SNAP connector.
Disconnect the tubing from the top of the conductivity monitor and replace
it with a piece of tubing leading to the waste container.
Select System:Settings in UNICORN System Control module.
In the Instructions list, select Conductivity:Cond Temp Compensation.
Set the Compensation factor to 0 and click OK.
Select System:Calibrate from the main System control menu.
5.4.2 Calibrate the conductivity monitor
96).
W
5 Maintenance
5.4 Calibration procedures
Section 5.3.1 Clean
105