peak of C3+. This may or may not be labeled but is not used in
calculations. There could be trace amounts of other components
in the calibration blend as indicated by gates with no component
label. This is shown below between the CO2 and C2 peaks.
17F
Place the vertical line of the cursor over the small tick mark on
C2, and verify the time in the upper right hand corner of the
chrom is approximately 220 seconds. It is not critical that it be
exactly 220 seconds but should be within 3 or 4 seconds.
Step 18
Verify the Stream Sequence
While the stream sequence was verified during the Start-Up Wizard, the
user may want to see if the streams they set to run automatically are
For a stream to automatically sequence, it must be enabled and be
present in the stream sequence. The calibration stream will function
without being enabled. Additionally, the stream should not be enabled or
be present in the stream sequence unless the user prefers to run cycles
on the calibration stream.
A reason that the user may not want the calibration stream enabled is
the fact that when enabled, a stream displays information (Unnormalized
Total, Superior CV, etc.) on the Operation screen. This information
represents process stream information and not calibration stream
information and can lead to confusion.
After confirming the streams, close the Stream Sequence screen.