Section 19 - Appendix B–Filter Board Options PWA 79562-01 and PWA 105521-01
19.16

Test Filter Option (Alternate Procedure)

This procedure should only be used when your Full Scale Range Option is 05, 06, 15, 16,
or 17 (see
Section
8.2) and the full scale amplitude calculated in
procedure
(Section
19.15) is greater than 3.5V or too small to set up accurately on your
function generator.
Note: Both channels A and B must be calibrated before performing this test (see
15). Test equipment should be set up as shown in step two of
are used, test with barriers in place.
1.
Configure the main board as follows to change the Integrator Stage to a Gain
Stage:
Channel A:
Channel B:
2.
Adjust the frequency of the function generator to the -3dB frequency of the filter.
For High Pass Filters, the -3dB frequency is the same as the corner frequency
chosen in
Section 19.7 or
times the corner frequency chosen in
the function generator for a full scale sine wave signal with a -7.5 Vdc offset:
Full Scale Option
05
06
15
16
17
3.
Verify that the front panel bargraph for the channel being tested reads 65% to
75% of full scale range.
4.
If you are testing a Band Pass Filter, repeat steps two and three at the second -
3dB frequency.
Install
W6B,C
Remove
W6A
Install
W3D,E
Remove
W3F
19.8. For Low Pass Filters, the -3dB frequency is 1.06
0-1 in/s
0-2 in/s
0-25 mm/s
0-50 mm/s
0-100 mm/s
step 5b
Section
Section 19.10 or
19.11. Set the amplitude of
Full Scale Input Voltage
Vpk
0.500
1.00
0.492
0.984
1.969
of the main
Section
19.15. If barriers
Vrms
0.354
0.707
0.348
0.696
1.392
83