3300/25 Dual Accelerometer Input Monitor Operation Manual
9. Meter Scale Replacement
The monitor meter scales can be replaced for operation with different full scale ranges.
The replacement meter scales are located in the back of this manual.
1.
Set the monitor full scale range. (See the Appendix A, Main Board Options.)
2.
Cut the meter scale from the back of this manual. Be sure to cut along the marked
outline so the meter scale will fit properly.
3.
Open the front panel by loosening the two monitor retaining screws and sliding the
front panel to the right.
4.
Remove the old meter scale.
5.
Insert the new meter scale in the front panel.
6.
Close the front panel.
24