Calibrate Channels - GE Bently Nevada 3300/25 Operation Manual

Dual accelerometer input monitor
Table of Contents

Advertisement

Before you begin Channel Calibration, you must choose the full scale range and set the
appropriate jumpers
calibrate channels you will need these instruments and tools
• multimeter
• function generator
• screwdriver
Use this procedure to calibrate both channels.
Calibrating a channel involves these main steps:
• Set the filter option to No Filters
• Connect and adjust the test instruments
• Calibrate the channels
• Prepare to bring the monitor on line

15. Calibrate Channels

(See Section
18.6). Calibrate both channels of your monitor. To
Section 15 - Calibrate Channels
37

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents