20.13

Test Filter Option (Alternate Procedure)

This procedure should only be used when your Full Scale Range Option is 05, 06, 15, 16,
or 17 (see
Section
8.2) and the full scale amplitude calculated in step 5b of the main
procedure
(Section
20.12) is greater than 3.5V or too small to set up accurately on your
function generator.
Note: Both channels A and B must be calibrated before performing this test (see
15). Test equipment should be set up as shown in step two of
are used, test with barriers in place.
1.
Configure the main board as follows to change the Integrator Stage to a Gain
Stage:
Channel A:
Channel B:
2.
Adjust the frequency of the function generator to the -3dB frequency of the filter.
For High Pass Filters, the -3dB frequency is the corner frequency chosen in
Section 20.7 or
frequency chosen in
for a full scale sine wave signal with a -7.5 Vdc offset:
Full Scale Option
05
06
15
16
17
3.
Verify that the front panel bargraph for the channel being tested reads 65% to
75% of full scale range.
4.
If you are testing a Band Pass Filter, repeat steps two and three at the second -
3dB frequency.
Section 20 - Appendix C–Filter Board Options PWA 148921-01
Install
W6B,C
Remove
W6A
Install
W3D,E
Remove
W3F
20.8. For Low Pass Filters, the -3dB frequency is the corner
Section
20.10. Set the amplitude of the function generator
Full Scale Input Voltage
0-1 in/s
0-2 in/s
0-25 mm/s
0-50 mm/s
0-100 mm/s
Section
Vpk
0.500
1.00
0.492
0.984
1.969
Section
20.12. If barriers
Vrms
0.354
0.707
0.348
0.696
1.392
133