Sense:correction Subsection - HP ESA-E Series Programmer's Manual

Esa spectrum analyzers
Table of Contents

Advertisement

SENSe:CORRection Subsection

Perform Amplitude Correction
[:SENSe]:CORRection:CSET:ALL[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET:ALL[:STATe]?
Turns On or Off the amplitude corrections. When turned On, only the
correction sets that were turned on are enabled. When turned Off, all of
the correction sets are disabled.
Factory Preset
and *RST:
Off
Remarks:
To turn On or Off an individual correction set, use:
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe].
Front Panel
Access:
AMPLITUDE/Y Scale, Corrections, Corrections On Off
Set Amplitude Correction Data
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA
<freq>,<rel_ampl>{,<freq>,<rel_ampl>}
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA?
Sets the amplitude correction data. These frequency/amplitude
corrections will be applied to the displayed data to correct for system
losses/gains outside the analyzer. Four different sets of correction data
can be stored.
Example:
:CORR:CSET1:DATA
900E6,0.3,1.0E9,0.35,1.3E9,0.2
Range:
200 points per set
Default Unit:
There are no units on the frequency and amplitude
pairs. They must be entered in hertz (Hz) and decibels
(dB).
Remarks:
CSET number equivalents to front panel access
definitions are as follows:
CSET1 is Antenna
CSET2 is Cable
CSET3 is Other
CSET4 is User
Front Panel
Access:
AMPLITUDE/Y Scale, Corrections, Modify
Chapter 5
Language Reference
SENSe Subsystem
5-77