Language Reference
SENSe Subsystem
SENSe:CHPower Subsection
Set Channel Power Number of Averages
[:SENSe]:CHPower:AVERage:COUNt <integer>
Set the channel power measurement number of averages.
Remarks:
This command is used for measurements in the
MEASURE
Front Panel
Access:
none
Channel Power Averaging On/Off
[:SENSe]:CHPower:AVERage[:STATe] OFF|ON|0|1
Turn on or off the channel power measurement averaging.
Factory Preset
and *RST:
Off
Remarks:
This command is used for measurements in the
MEASURE
Front Panel
Access:
Meas Setup, Avg Number On Off
Set Channel Power Integration Bandwidth
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration <freq>
Set the frequency span (bandwidth) over which to integrate power.
Remarks:
This command is used for measurements in the
MEASURE
Front Panel
Access:
Meas Setup, Integration BW
Set Channel Power Span
[:SENSe]:CHPower:FREQuency:SPAN <freq>
Set the measurement channel power span of the analyzer.
Remarks:
This command is used for measurements in the
MEASURE
Front Panel
Access:
Meas Setup, Chan Pwr Span
5-76