Field Mountable Videographic Recorder
SM500F
Thermocouple Type
Ni-Cr/Ni-Al (K)
Brown Blue
Ni-Cr/Cu-Ni (E)
Nicrisil/Nisil (N)
Orange Blue Orange Orange Red Orange
Pt/Pt-Rh (R and S)
White Blue Green
Pt-Rh/Pt-Rh (B)
Cu/Cu-Ni (T)
White Blue
Fe/Con (J)
Yellow Blue Black
Fe/Con (DIN 43710)
Table 2.1 Thermocouple Compensating Cable
2.5 Power Supply Connections – Fig. 2.10
Line
Neutral
+
Fig. 2.10 Power Supply Connections
Note. Power supply terminal screws must be tightened to a torque of 0.8 Nm (7 lbf.in).
IM/SM500F Issue 10.1
BS1843
ANSI MC 96.1
+
–
Case
+
Red
Yellow Red Yellow
–
Black Red Green
–
Blue
Blue
White Red Black
–
Fuse,
315 mA, Type T
Warning. Use Fuse Rating 315 mA (max.) Type T
A AC Supply
Fuse,
1.5 A, Type
T
Warning. Use Fuse Rating 1.5 A (max.) Type T
B DC Supply
Compensating Cable
DIN 43714
–
Case
+
Red
Green Green Green White Green *
–
Red
White White Orange White Orange *
–
Red
Blue
Red
Brown Brown Brown White Brown *
Red
DIN 43710
–
Blue/red
L
85 V min. to 265 V max. 50/60 Hz
N
+
10 V min. to 36 V max.DC
2 Installation
BS4937 Part No.30
–
Case
+
–
Violet White
–
Pink
White
–
Grey
White
Blue
Blue
Black White Black *
* Case Blue for intrinsically safe circuits
Blue
Blue
–
Case
Violet *
Pink *
Grey *
–
17