Philips V24CT User's Reference Manual page 373

Cms patient monitoring system volume 1
Table of Contents

Advertisement

Procedure
Preventative Maintenance
Replace backlight tube
assembly
Cleaning procedures
Full performance checks
including safety checks.
Check Calibration of NBP &
Sidestream CO
2
Check Calibration of all
parameters except NBP &
Sidestream CO
2
Frequency
2.5 years of continuous use
or as needed
Perform the Preventive
Maintenance Procedure for
the M1180A #A01/A02 Wall
Mount Assembly every year.
At least every two years,
after any repairs, or as
needed.
Once a year
Every 2 years
Monitor Maintenance
Source of Information
See "Testing and
Maintaining the CMS" in the
"CMS Service Guide" or
"Testing and Maintaining
the V24 and V26"in the "V24
and V26 Service Guide".
See "Testing and
Maintaining the CMS" in the
"CMS Service Guide" or
"Testing and Maintaining
the V24 and V26"in the "V24
and V26 Service Guide".
See chapter 3 of the CMS
Service Guide "Maintaining
the CMS".
See chapter 3 of the CMS
Service Guide "Maintaining
the CMS".
Maintenance 13-7