Parameter Settings Transfer Messages - Philips V24CT User's Reference Manual

Cms patient monitoring system volume 1
Table of Contents

Advertisement

Parameter Settings Transfer Messages

The following messages related to Parameter Settings Transfer can
appear on the display:
Message
(Parameter name)
no settings
transfer capability
(Parameter name)
check settings''
Check Patient
Category and
(parameter name)
settings
(Parameter name)
unable to store
settings in module
Parameter Settings Transfer Messages
Condition
The indicated parameter
module is not a T
module.
No settings are stored in
the module, or the
settings are incorrectly
stored.
The patient category
stored with the settings
in the module is
different from that of the
monitor.
The parameter module
was removed before the
settings could be stored,
or there is a malfunction
in the module. (This
message can appear with
Settings Transfer On or
Off.)
Action required
Insert a module that has
a T on the front.
Enter the parameter
Task Window and check
the settings. The
settings are
automatically stored
when you exit the Task
Window.
Either change the
patient category for the
monitor or change the
parameter settings.
Press
Silence Reset
re-insert the module and
check/adjust the
settings in the
parameter Task
Window. Exit the Task
Window and leave the
module in the rack for
about 10 seconds before
you remove it. If the
same message appears,
replace the module.