Heart Rate
(HR) / Pulse
Cardiotach Range
HR Alarm Limit Range
Pleth/Pressure Pulse Alarm Limit
Range
The expected range can be estimated by performing a manual pulse
count.

RESP

Respirotach Range
High Alarm Limit Range
Low Alarm Limit Range
In general, use the Adult or Pediatric category if rates below 50 rpm are
expected; use Neonatal if expected rate is above 100.

Pressure

The algorithm responsible for respiration artifact suppression is used in
all three patient categories for the 10 and 30 scales. In the Pediatric and
Adult categories, the algorithm is also used for the 60 scale.
SpO
Default for averaging time is 10 seconds for Adult, Pedi and Neo.
2
Note—
Adult/Pedi (rpm)
0 to 120
10 to 100
0 to 95
Default can be changed in Configuration Mode.
Changing the Patient Category
Adult (bpm)
Pedi/Neo (bpm)
15 to 300
15 to 350
15 to 250
15 to 300
30 to 250
30 to 300
Neo (rpm)
0 to 180
30 to 150
0 to 145
Setting up your Monitor 3-51