Alarm Display; Alarm Functions On The Control Panel - Philips V24CT User's Reference Manual

Cms patient monitoring system volume 1
Table of Contents

Advertisement

Alarm Display

Alarm Display
The alarm messages are displayed across the top of the screen. When
you switch the power on, the system alarm capability is switched off.
This is indicated by an ALARMS SUSPENDED or ALARMS SUSPND 1, 2
or 3 MIN message at the top right corner of the screen, depending on the
alarm suspended mode which is configured on the monitor. The Alarm
Suspend symbol( ) on the control panel is also illuminated.
Note—
for approximately 3 seconds. If there is no sound, please check the
volume settings.
If there is more than one red alarm, yellow alarm or INOP message, you
will see an up arrow (
displayed in rotation every 2 seconds. If your patient has more than one
alarm, you can see a list of all these active alarms in the Alarm Messages
Task Window. To access this Task Window press
Alarms
Warning
Warning
Alarms and INOPs are not active in the configuration and service
modes, but are active in monitoring and demo modes.
Alarm
Functions
on the
Control
Panel
5-2
Alarm Functions
When you switch the monitor on, a loudspeaker test is conducted
followed by
Alarm Messages
Alarm Lamps
Silence/Reset key
Suspend key
) next to the message. The messages are
.
Alarms/
Silence/
Volume
Reset
Main
Suspend
Screen
Alarms Suspended Lamp
or
Alarms/Volume
Other
Monitor
Patients
Setup
Change
Screen