GE LOGIQ Totus Basic Service Manual page 407

Table of Contents

Advertisement

D
5936427, R
IRECTION
EV
8-6-4-2
Removal Procedure
Unscrew 4 screws from the Ultrasound Bezelless
1.
23.8inch LCD T2.
Use the Hook Remover to remove the Monitor Rear
2.
Cover 23.8 inch.
Unscrew 8 screws to remove the Monitor Front Cover
3.
23.8inch T2.
8-6-4-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform "Functional Checks after FRU replacement (Debrief Guideline)" as described in
12 on page 8-223
GE HealthCare CONFIDENTIAL
. 1
Step
Section 8-6 - Replacing Monitor and Monitor Arm
LOGIQ T
Basic Service Manual
OTUS
Corresponding graphic
Section 8-
8 - 81

Advertisement