Hide thumbs Also See for LOGIQ S6:
Table of Contents

Advertisement

Quick Links

Technical
Publication
Direction 5162630
R
3
EVISION
GE Healthcare
LOGIQ™ S6 Service Manual
©
Copyright
2006-2008 by General Electric Co.
GE Healthcare

Advertisement

Table of Contents
loading

Summary of Contents for GE LOGIQ S6

  • Page 1 GE Healthcare Technical Publication Direction 5162630 EVISION GE Healthcare LOGIQ™ S6 Service Manual © Copyright 2006-2008 by General Electric Co.
  • Page 3 • WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN WARNUNG ANWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND NACHVOLLZIEHEN KÖNNEN. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLä GE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
  • Page 4 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL • ESTE MANUAL DE SERVICIO Só LO ESTÁ DISPONIBLE EN INGLé S. • SI ALGú N PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLé S, LA TRADUCCIÓN ES RESPONSABILIDAD DEL CLIENTE.
  • Page 5 ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ∆ΥΝΟΥΣ. • EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A FIGYELMEZTETÉS KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
  • Page 6 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL • TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKé M JAZYCE. • V Př íPADě , ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTř EBUJE Ná VOD V JINé M JAZYCE, JE ZAJIšTě Ní Př EKLADU DO ODPOVíDAJíCíHO JAZYKA ú...
  • Page 7 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL • ÐIS EKSPLOATAVIMO VADOVAS YRA IÐLEISTAS TIK ANGLØ KALBA. • JEI KLIENTO PASLAUGØ TEIKËJUI REIKIA VADOVO KITA KALBA – NE ANGLØ, VERTIMU PASIRÛPINTI TURI KLIENTAS. • NEMËGINKITE ATLIKTI ÁRANGOS TECHNINËS PRIEÞIÛROS DARBØ, NEBENT ÁSPËJIMAS...
  • Page 8 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL • Да н н о е рук о в о дс т в о п о о бс луж и в а н и ю ПРЕДОСТАВЛЯЕТСЯ т о ль к о н а а н г ли й с к о м Яз ы к е .
  • Page 9 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL...
  • Page 10 GE Medical Systems personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 11 The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Medical Systems. GE Medical Systems may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
  • Page 12 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank.
  • Page 13 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Revision History Revision Date Reason for change May 17, 2006 Initial Release Nov.1, 2007 BT08 release Nov.6, 2008 BEP4 release List of Effected Pages PAGES REVISION PAGES REVISION...
  • Page 14 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL...
  • Page 15: Table Of Contents

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Table of Contents Important Precautions CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
  • Page 16 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL General Console Requirements........2 - 2 Console Environmental Requirements .
  • Page 17 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL CHAPTER 4 Functional Checks Overview............4 - 1 Purpose for Chapter 4 .
  • Page 18 Block Diagrams and Theory......... . .5 - 14 Block Diagram LOGIQ S6 (For forward production of BT08 or later) ..5 - 14 Block Diagram LOGIQ S6 (For original HW config systems, including initial release, and upgraded BT08 systems) .
  • Page 19 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL CHAPTER 6 Service Adjustments Overview............6 - 1 Purpose of this chapter 6 .
  • Page 20 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview ............8 - 1 Purpose of Chapter 8 .
  • Page 21 BT08 with BECOMP3 Parts ........9 - 7 Renewal Parts List for Original LOGIQ S6 (BT06) ......9 - 9 Original LOGIQ S6 : Equipment Models Covered in this Chapter .
  • Page 22 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Safety Test Overview ......... .10 - 12 Grounding Continuity .
  • Page 23: Introduction

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This Chapter describes important issues related to safety servicing this ultrasound machine. The service provider must read and understand all the information presented here before installing or servicing a unit.
  • Page 24: Typical Users Of The Basic Service Manual

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 1-1-4 Typical Users of the Basic Service Manual • Service Personnel (installation, maintenance, etc.). • Hospital’s Service Personnel • Architects (Some parts of Chapter 2 - Pre-Installation) 1-1-5 LOGIQ™...
  • Page 25: Purpose Of Operator Manual(S)

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 1-1-6 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the LOGIQ™ S6 and also kept near the unit for quick reference.
  • Page 26: Important Conventions

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Model Designations. This manual covers the LOGIQ™ S6 scanners. Icons. Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
  • Page 27 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 1-2-2 Standard Hazard Icons Important Information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could possibly cause harm.
  • Page 28: Product Icons

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-5 Warnings Label/Symbol Purpose/Meaning Location...
  • Page 29 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Table 1-5 Warnings Label/Symbol Purpose/Meaning Location Note “CAUTION - Dangerous voltage” (the lightning flash with arrowhead in Various equilateral triangle) is used to indicate electric shock hazards. “General Warning Sign”...
  • Page 30 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Table 1-5 Warnings Label/Symbol Purpose/Meaning Location Note This symbol indicates that waste Rear Panel BT08 or later electrical and electronic equipment must not be disposed of as unsorted municipal waste and must be collected separately.
  • Page 31: Introduction

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
  • Page 32: Electrical Safety

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 1-3-4 Electrical Safety To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The system is equipped with a three-conductor AC power cable. This must be plugged into an approved electrical outlet with safety ground.
  • Page 33 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 1-3-5 Label Locations NOTE: For the symbols shown in the illustration below, refer to previous pages in this chapter. NOTE: Background color of the triangle caution marks may be white depending on the system shipment timing.
  • Page 34 M A N U F A C T U R E D   : F R E Q U E N C Y                     50/60H z 東京都日 野市旭 が丘4 −7− 127 GE Yok oga wa M edi cal Systems Europe/Asia/China/ Japan 100V Console...
  • Page 35: Lockout/Tagout Requirements (For Usa Only)

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 1-3-6 Dangerous Procedure Warnings Warnings, such as the example below, precede potentially dangerous procedures through our this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN DANGER THIS EQUIPMENT.
  • Page 36: Electromagnetic Compatibility

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 1-4 EMC, EMI, and ESD 1-4-1 Electromagnetic Compatibility Electro Magnetic Compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface.
  • Page 37: Customer Assistance

    1-262-524-5698 Customer Answer Center Fax: +1-414-647-4125 Europe Tel: GE Ultraschall Deutschland GmbH& Co. KG +33 0 130 831 300 - CARDIAC BeethovenstraBe 239 +33 0 130 831 300 - GENERAL IMAGING Postfach 11 05 60, D-42655 Solingen Fax: +49 212 2802 431...
  • Page 38 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 1-5-2 Contact Information (cont’d) NOTE: If this equipment does not work as indicated in the Operator Manual(s), contact your support center. Have the system ID number available when you call.
  • Page 39: Pre Installation

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Chapter 2 Pre Installation Section 2-1 Overview 2-1-1 Purpose of this chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ™ S6.
  • Page 40: General Console Requirements

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 2-2 General Console Requirements 2-2-1 Console Environmental Requirements Table 2-9 Environmental Requirements for LOGIQ™ S6 Scanners Operational Storage Transport 10 - 35 -10 - 50 -10 - 50...
  • Page 41: Electrical Requirements

    Electrical Requirements NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Line: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 42 2-2-2-6 Unit Power Plug If the unit arrives without the power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-7 Power Stability Requirements Voltage drop-out Max 10 ms.
  • Page 43: Emi Limitations

    RF leakage. Or, if a label has been found in such a position, move the label. The interconnect cables are grounded and require ferrite beads and other shielding. Use GE specified Also, cable length, material, and routing are all important; do not change from what is harnesses and peripherals specified.
  • Page 44: Probes Environmental Requirements

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 2-2-4 Probes Environmental Requirements Table 2-13 Operation and storage Temperatures for 2D Probes Operational Storage Transport 10 - 40 -10 - 60 -40 - 60 Temperature 50 - 104...
  • Page 45: Facility Needs

    All electrical work on these product must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
  • Page 46: Required Features

    Emergency oxygen supply • Storage for linens and equipment • Nearby waiting room, lavatory, and dressing room • Dual level lighting (bright and dim) • Lockable cabinet ordered by GE for its software and service manual Section 2-3 - Facility Needs...
  • Page 47: Recommended And Alternate Ultrasound Room Layout

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 2-3-4 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a minimal floor plan for ultrasound equipment: LOGIQ LOGIQ Figure 2-3 RECOMMENDED ULTRASOUND ROOM LAYOUT Section 2-3 - Facility Needs...
  • Page 48 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 2-3-5 Networking Pre-installation Requirements 2-3-5-1 Purpose of DICOM Network Function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network. Examples of DICOM services include the transfer of images to workstations for viewing or transferring images to remote printers.
  • Page 49 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 2-3-5-2 DICOM Option Pre-installation Requirements (cont’d) LOGIQ™ S6 IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default ROUTER1...
  • Page 50 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 2-12 Section 2-3 - Facility Needs...
  • Page 51: Installation

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Chapter 3 Installation Section 3-1 Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are references to a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
  • Page 52: Installation Warnings

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-1-3 Installation Warnings 1.) Since the LOGIQ™ S6 weighs approximately 179 kg. (395 lb) without options, preferably two people should unpack it. Two people are also preferable for installing any additional bulky items.
  • Page 53 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1.
  • Page 54 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment (cont’d) NOTE: Check the shipping container for special instructions. Verify that the container is intact. In some cases a secondary container may be used. If so, ask the carrier for unpacking instructions.
  • Page 55: Safety Reminders

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-2-1 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE DANGER AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT TOUCH THE UNIT! Two people should unpack the unit because of its weight.
  • Page 56: Moving Into Position

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-2-2 Moving into Position Do not lift the unit by the Keyboard. CAUTION Do not tilt the unit more than 10 degrees to avoid tipping it over. To avoid injury by tipping over. Set the monitor to the lowest position before moving.
  • Page 57: Lcd Monitor Tilt Caution

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-2-3 LCD Monitor Tilt Caution Do NOT put your hand here! Your fingers might be pinched when adjusting angle of the LCD CAUTION monitor. Figure 3-7 LCD Tilt Caution...
  • Page 58: Preparing For Installation

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 3-3 Preparing for Installation 3-3-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered or damaged.
  • Page 59: Completing The Installation

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 3-4 Completing the Installation 3-4-1 Probe (Transducer) Connection 1.) Connect a transducer to the upper transducer receptacle as follows: a.) Ensure that the transducer twist lock lever to the horizontal position.
  • Page 60 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-4-2 Optional Peripherals/Peripheral Connection Depending on the customer order, the peripheral(s) already has been installed onto the scanner before shipment. If it does not, select the proper location to install the peripherals, following the table below.
  • Page 61: Optional Peripherals/Peripheral Connection

    Check if the fixing belts are loosen. If it is, tighten the belts again to secure the peripheral(s). If necessary, remove the peripheral from the device room to tighten the belts. Connect the related cables between the peripheral and LOGIQ S6 if they are NOT Belts connected each other. Belts...
  • Page 62 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-4-2-1 Approved on-board peripherals Refer to Section 5-4 Peripheral Compatibility. Connecting Cables Possible equipment damage if mis-handled. Be sure to use the following recommended connecting CAUTION cables to connect recording devices and a network with LOGIQ™ S6 console.
  • Page 63: Available Probes

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-4-3 Available Probes See in specification in the LOGIQ™ S6 Reference Manual for Probes and intended use. See Chapter 9 - Renewal Parts for Part Numbers to be used when ordering new or replacement probes.
  • Page 64 Japan. Probes which have only one part number are not available in Japan. The following transducers, additional to ones listed in Table 3-21, are supported by BT08 and later LOGIQ S6 models. Table 3-21 List of Transducers supported by BT08 or later Material of...
  • Page 65: Video Specification

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-4-4 Video Specification Table 3-22 Video Specifications 800x600 1280 x 1024 Timing Parameter 75Hz Timing Parameter 60Hz Horizontal Rate [kHz] 46.88 Horizontal Rate [kHz] Horizontal Period [µs] 21.33 Horizontal Period [µs]...
  • Page 66: Installation Paperwork

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 3-5 Installation Paperwork NOTE: During and after installation, the documentation (i.e. Users Manual, Installation Manuals...) for the peripheral units must be kept as part of the original system documentation. This will ensure that all relevant safety and user informations are available during the operation and service of the complete system.
  • Page 67 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-5-1-1 Rear Panel Connector Located on the rear panel are video input and output connectors, audio input and output, camera expose connectors, footswitch connector power connector and control connections for VCR, printer, and service tools.
  • Page 68 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment of S-Video Connector Connector: S-Terminal, 4-pin Table 3-23 Pin Assignment of S-Video Connector Pin No Output/Input Signal Description SVIDEO OUT/IN YG...
  • Page 69 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment of Foot Switch Round 5 pin connector. Table 3-26 Pin Assignment of Mini-Jack for Footswitch Pin No Output Signal SW1-WH SW2-RD...
  • Page 70 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment for Camera B/W Table 3-29 Pin Assignment of Mini-Jack for Controlling B/W Camera Pin No Output Signal PRINT Signal GND NOTE: Output level of control signals indicated in the above tables are TTL level.
  • Page 71 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment of Composite Out, Color BNC Connector Table 3-33 Pin Assignment of BNC Connector Pin No Output Signal Description Composite Video Color...
  • Page 72 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 3-22 Section 3-5 - Installation Paperwork...
  • Page 73: Functional Checks

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose for Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ S6 console, diagnostics by using the built-in service software, and power supply adjustments.
  • Page 74: General Procedure

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 4-2 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
  • Page 75 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-2-1 Power Up (cont’d) 4.) Press the ON/OFF key at the front of the System once. ON/OFF Switch Figure 4-12 Power On/Off Standby Switch Location 5.) After initialization is complete, all lighted buttons on the Control Panel light and the default B-Mode screen or Patient screen (no probes are connected) is displayed on the monitor display.
  • Page 76 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-2-3 Entering the Maintenance Mode (cont’d) 4.) Click on Exit. Figure 4-13 Exit 5.) Enter the proper password to enter the maintenance mode. 6.) Click on OK. Figure 4-14 OK...
  • Page 77 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-2-3 Entering the Maintenance Mode (cont’d) 7.) Click on maintenance. Figure 4-15 Maintenance 8.) Click on Exit to window. Figure 4-16 Exit to Windows Section 4-2 - General Procedure...
  • Page 78: System Restart/Shutdown

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-3 System Restart/Shutdown Purpose: This is a description on how to Shutdown the system. 4-2-3-1 Complete Power Shutdown 1.) Press the ON/OFF key at the front of the System for about two (2) seconds. Refer to Figure 4-12 on page 4-3.
  • Page 79: Using Mod/Dvd Drive

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-4 Using MOD/DVD Drive 4-2-4-1 Using DVD Drive Never move the unit with a disk in the DVD because the drive actuator will not be locked and NOTICE the DVD could break.
  • Page 80 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-4-2 Using MOD Drive 1.) Before installing an MO disk in the MOD, check the MO disk for loose hardware or damaged labels which could jam inside the MO Drive. Also ensure that the slide switch in one corner of the disk is set so that the disk is write enabled (disk hole closed).
  • Page 81: Archiving And Loading Presets For Bt08

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-5 Archiving and Loading Presets for BT08 NOTE: Always save presets before any software reload. This ensures the presets loaded after the software reload are as up–to–date as possible.
  • Page 82 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-5-2 Using Regional Preset Regional Preset is selected during Application Load (Refer to 8-8-13 - Installing R7.x.x Application Software for details). Current Regional Preset is displayed under Utility > System > About .
  • Page 83 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-5-3 Cautions Using Regional Preset • Do not attempt to modify Regional Preset on Upgraded BT08. • Backup/Restore function should be used between the same region systems. Unexpected setting may result if you restore the preset files to another region setting system.
  • Page 84 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-5-5 Archiving Presets to an CD-R/DVD-R Disk 1.) Insert an empty (blank) formatted CD-R/DVD-R disk into the CD-R/DVD-R device. 2.) Access to the Utility Menu on the Touch Panel, and select System>Backup/Restore. The Backup screen will be shown on the monitor.
  • Page 85 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-5-6 Loading Presets from an CD-R/DVD-R disk 1.) Insert the CD-R/DVD-R disk with the archived Presets into the CD-R/DVD-R. 2.) Access to the Utility Menu on the Touch Panel, and select System>Backup/Restore. The Restore sheet will be shown on the monitor.
  • Page 86: Archiving And Loading Presets For R6.X.x

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-6 Archiving and Loading Presets for R6.x.x NOTE: Always save presets before any software reload. This ensures the presets loaded after the software reload are as up–to–date as possible.
  • Page 87 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-6-2 Archiving Presets to an MO/DVD-R Disk 1.) Insert an empty (blank) formatted MO/DVD-R disk into the MO/DVD-R device. 2.) Access to the Utility Menu on the Touch Panel, and select System>Backup/Restore. The Backup screen will be shown on the monitor.
  • Page 88 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-2-6-3 Loading Presets from an MO/DVD-R disk 1.) Insert the MO/DVD-R disk with the archived Presets into the MO/DVD-R. 2.) Access to the Utility Menu on the Touch Panel, and select System>Backup/Restore. The Restore sheet will be shown on the monitor.
  • Page 89: Basic Controls

    4-3-2-1 Recommended Test Phantoms GE Healthcare recommends the RMI 430GS phantom, but it is not required. It is the most current phantom recommended to our field service personnel and provides the necessary targets and extended life necessary for consistent system testing.
  • Page 90: Mode Checks (Qg)

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-3-3 Mode Checks (QG) For a functional check of the system’s different modes, refer to the LOGIQ™ S6 Quick Guide. The Quick Guide will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler.
  • Page 91: Cineloop Check (Qg)

    Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 4-41 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description Verify record/playback capabilities of the VCR.
  • Page 92: Image Management (Qg)

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-3-10 Image Management (QG) This section in the LOGIQ™ S6 Quick Guide talks about several topics: • Clipboard • Printing Images • Browsing and Managing an Exam’s Stored Image •...
  • Page 93 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Table 4-42 Brakes and Direction (swivel) Lock Check Step Task to do Expected Result(s) Shift pedal no.3 to the right. To engage swivel lock Section 4-2 - General Procedure...
  • Page 94 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-3-11-1 Brakes and Direction Locks Checks (cont’d) Swivel Lock Brake / Swivel Lock Figure 4-29 Brake and Swivel Location 4-22 Section 4-2 - General Procedure...
  • Page 95 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 4-3-11-2 Caster and Caster Link Checks Check if: • For Caster: No loose caster with it jacked up. Smooth operation when moving the scanner. • For Caster Link Assy: Effective brake and direction lock function (See section 4-3-11-1).
  • Page 96: Application Turnover Check List

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 4-4 Application Turnover Check List Complete these checks before returning the scanner to customer for use: 4-4-1 Software Configuration Checks Table 4-43 Software Configuration Checks Step Task to do...
  • Page 97: Diagnostics

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 4-5 Diagnostics 4-5-1 Service Software Menu Refer to Section 5-5 - Common Service Platform. 4-5-2 Utility Menu Refer to Section 5-5 - Common Service Platform. Section 4-5 - Diagnostics...
  • Page 98: Power Supply

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 4-6 Power Supply 4-6-1 Power Supply Test Procedure There is no need to perform any special tests on the Power Supplies if there does not seem to be a problem that may be related to the Power Supplies.
  • Page 99: Site Log

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 4-7 Site Log Table 4-44 Site Log DATE SERVICE PERSON PROBLEM COMMENTS Section 4-7 - Site Log 4-27...
  • Page 100 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Table 4-44 Site Log DATE SERVICE PERSON PROBLEM COMMENTS 4-28 Section 4-7 - Site Log...
  • Page 101: Components And Functions (Theory)

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview This chapter explains LOGIQ™ S6’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution System (PDS) and probes.
  • Page 102: Hardware Compatibility Matrix

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 5-2 Hardware Compatibility Matrix 5-2-1 Hardware Comaptibility BT06 or later 5-2-1-1 Plastic Mechanics Figure 5-30 Plastic Mechanics Section 5-2 - Hardware Compatibility Matrix...
  • Page 103 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-2-1-2 Monitor Figure 5-31 Monitor Section 5-2 - Hardware Compatibility Matrix...
  • Page 104 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-2-1-3 Keyboard Figure 5-32 Keyboard Section 5-2 - Hardware Compatibility Matrix...
  • Page 105 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-2-1-4 NEST boards Figure 5-33 NEST boards Section 5-2 - Hardware Compatibility Matrix...
  • Page 106 CRT Connection Label 5316350-2 SVP-BECOMP4 Accessory Includes CRT Connection Label SW / GHOST 5170144-2 LOGIQ S6 Ghost CD for BEP-NP, R6.1.0 Obsolete (Must be R6.2.3 by FMi72272) 5176956 LOGIQ S6 R6.1.2 Application Software CD Obsolete (Must be R6.2.3 by FMi72272) 5194288 LOGIQ S6 Ghost CD for BEP-NP R6.2.2...
  • Page 107 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-2-1-6 Rear I/O Figure 5-35 Rear I/O Section 5-2 - Hardware Compatibility Matrix...
  • Page 108 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-2-1-7 Power Supply Figure 5-36 Power Supply Section 5-2 - Hardware Compatibility Matrix...
  • Page 109 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-2-1-8 Other Mechanics Figure 5-37 Other Mechanics Section 5-2 - Hardware Compatibility Matrix...
  • Page 110 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-2-1-9 Cables Figure 5-38 Cables 5-10 Section 5-2 - Hardware Compatibility Matrix...
  • Page 111 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-2-1-10 Option FRU Figure 5-39 Option FRU Section 5-2 - Hardware Compatibility Matrix 5-11...
  • Page 112: Hardware Identification Tip

    5-2-2 Hardware Identification Tip 5-2-2-1 How to Identify LS6 BT08 with BEP4 Units The flowchart below describes how to identify BT08 with BEP4 unit. LOGIQ S6 BT06 LOGIQ S6 BT06 LOGIQ S6 BT08 LOGIQ S6 BT08 Differentiating BECOMP4 systems (with BECOMP3)
  • Page 113 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-2-2-2 LCD Monitor Differences The flowchart below describes how to differentiate LCD monitor type. BT08 GPN Identification Mark (Bottom of LCD monitor) Figure 5-41 LCD monitor- How to Differentiate...
  • Page 114: Block Diagrams And Theory

    EVISION ERVICE ANUAL Section 5-3 Block Diagrams and Theory 5-3-1 Block Diagram LOGIQ S6 (For forward production of BT08 or later) EUSC I/F & Pulse generator Figure 5-43 Block Diagram LOGIQ S6 (For forward production of BT08 or later) 5-14...
  • Page 115: Block Diagram Logiq S6

    LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-3-2 Block Diagram LOGIQ S6 (For original HW config systems, including initial release, and upgraded BT08 systems) Figure 5-44 Block Diagram LOGIQ S6 (For original HW config systems, including initial release, and upgraded BT08 systems)
  • Page 116: General Information

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-3-3 General Information • LOGIQ™ S6 is a phased and linear array ultrasound imaging scanner. It has provisions for analog input sources like ECG and phono. A dedicated Doppler probe may also be connected and used.
  • Page 117: Patient I/O (Option)

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-3-4 Patient I/O (Option) The optional Patient I/O is mounted at the front of the scanner with its connector panel. Available inputs: • 5-3-5 Top Console The Top Console includes a Stand By/On switch, a keyboard, different controls for manipulating the picture quality, controls for use in Measure &...
  • Page 118: Interconnect Cabling (For Forward Production Of Bt08 Or Later With Bep4)

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-3-8 Interconnect Cabling (For forward production of BT08 or later with BEP4) Figure 5-45 Interconnect Cabling (For forward production of BT08 or later with BEP4) 5-18 Section 5-3 - Block Diagrams and Theory...
  • Page 119: Interconnect Cabling (For Forward Production Of Bt08 Or Later)

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-3-9 Interconnect Cabling (For forward production of BT08 or later) Figure 5-46 Interconnect Cabling (For forward production of BT08 or later) Section 5-3 - Block Diagrams and Theory...
  • Page 120: Interconnect Cabling

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-3-10 Interconnect Cabling Figure 5-47 Internal Cable Connection CRT type for backward production before BT08 (For original HW config systems, including initial release, and upgraded BT08 systems) 5-20...
  • Page 121 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-3-10 Interconnect Cabling (cont’d) Figure 5-48 Internal Cable Connection LCD type for backward production before BT08 (For original HW config systems, including initial release, and upgraded BT08 systems)
  • Page 122 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-3-10 Interconnect Cabling (cont’d) Figure 5-49 USB Connection (For LS6 BECOMP4) 5-22 Section 5-3 - Block Diagrams and Theory...
  • Page 123 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-3-10 Interconnect Cabling (cont’d) Figure 5-50 USB Connection (For CRT/LCD) Section 5-3 - Block Diagrams and Theory 5-23...
  • Page 124: Peripheral Compatibility

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 5-4 Peripheral Compatibility Figure 5-51 List of all supported printers 5-24 Section 5-4 - Peripheral Compatibility...
  • Page 125: Common Service Platform

    The Service Platform contains a set of software modules that are common to all PC backend ultrasound and cardiology systems. This web-enabled technology provides linkage to e-Services, e-Commerce, and the iCenter, making GE’s scanners more e-enabled than ever. The Common Service Platform will increase service productivity and reduce training and service costs.
  • Page 126 GEMS products. 5-5-3-1 Internationalization The user interfaces provided by the service platform are designed for GE personnel and as such are in English only. At this time there is no multi-lingual capability built into the Common Service Interface. 5-5-3-2 Service Login 1.) Touch Utility >...
  • Page 127 Requires a network connection and knowledge of a password.* NOTE: *For a GE Field Engineer, the password changes at specific intervals. Except for iLinq, all access to the service interface is via the Network port. A modem on the scanner is specifically intended for iLinq use.
  • Page 128 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-5-4 Service Home Page When the Common Service Desktop is started, the Service Home Page appears. The Home Page contains the software revision along with the hardware inventory and the results of the latest System Health Information.
  • Page 129 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-5-5 Error Logs Page From the Error Logs page the Log Viewer displays five categories with pull-down sub-menus. The Service Interface has a log viewing tool driven by the following high-level requirements.
  • Page 130 Descriptions System Software bug information for software engineering to debug the software. Power Not used for LOGIQ S6. Infomatics Exam informations. Whenever the New patient or End Current Patient is pressed, they are reported as a log. 5-30 Section 5-5 - Common Service Platform...
  • Page 131 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Table 5-47 Error Log Descriptions Logs Descriptions The error occurred in the temperature with “!!” mark. In the time when the error occurred, the message “Temp is above safe limit. Check air-filter! Shutdown in #seconds.” appears in the status bar and the system is automatically shut down.
  • Page 132 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-5-5-1-1 Log Viewing The Service Platform has a log viewing tool driven by the following high-level requirements. • Simple filtering of the scanner log(s) with filtering capabilities being a function of login access permissions.
  • Page 133 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-5-5-4 Filter Click on he sub-menu, System Logs, to enter Filter menu. Filtering is used to extract select data from the Error Logs. There are four different error levels and 11 packages that can be filtered.
  • Page 134 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-5-6 Diagnostics The Diagnostic page includes various diagnostic tests for troubleshooting. Figure 5-61 Diagnostic Page 5-5-6-1 Diagnostic Reports Diagnostic tests return a report to the Service Platform. The platform retains the report and allows for future viewing of the diagnostic logs.
  • Page 135: Calibration

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-5-7 Image Quality The Image Quality page contains tools for troubleshooting image quality issues. Figure 5-62 Image Quality Page 5-5-8 Calibration The Calibration page is used to calibrate the system.
  • Page 136: Replacement

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-5-10 Utilities The Utilities page contains several miscellaneous tools. These includes access to the problem/solution database at the Online center. Figure 5-65 Utilities Page 5-5-11 Replacement Figure 5-66 Part Replacement Page...
  • Page 137: Remote Software/Option Installation And Updates

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-5-12 Run the Planned/Preventive/Proactive Maintenance tools during such PM visits as running error logs and deleting old test files. Figure 5-67 Planned Maintenance Page 5-5-13 Remote Software/Option Installation and Updates The Common Service Platform provides for remote software downloads including A11 downloads to the scanner.
  • Page 138: Password

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 5-6 Password The following windows request entry of Password. This allows you to be entered into Utility function or Service function with different access and security use levels.
  • Page 139: For Maintenance Access Window

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-6-2 For Service Login Window (cont’d) Figure 5-69 Service Login window 5-6-3 For Maintenance Access Window When you login Windows desktop for maintenance access, this window is open.
  • Page 140: Air Flow Control

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 5-7 Air Flow Control 5-7-1 Air Flow Distribution Figure 5-71 Air Flow Inside the Scanner The four air flow pathes allow the scanner to be cooled down as shown in the figure above.
  • Page 141: Fans

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-7-3 Fans Three fans for Nest Assy One fan for PC box One fan for LV unit One fan for HV unit Figure 5-72 Fans The scanner contains the nine fans at the following positions for producing an air flow.
  • Page 142: Monitor Video Specification

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 5-8 Monitor Video Specification 5-8-1 Input 5-8-1-1 Table 5-50 CRT Specifidations Standard Format Signal Sync Termination H: TTL VESA DDC2 SVGA 800 x 600 / 75 Hz...
  • Page 143: Outputs

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-8-2 Outputs 5-8-2-1 SVGA Table 5-52 SVGA Specifidations Standard Format Signal Sync Termination H: TTL VESA DDC2 SVGA 800 x 600 / 75 Hz RGB: 0 - 700 mV...
  • Page 144: Svhs And Composite Video

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 5-8-3 SVHS and Composite Video 5-8-3-1 Basic DC Parameters Parameters NTSC White relative to blank 714 +/- 7 mV 700 +/- 7 mV Black relative to blank 54 +/- 7 mV...
  • Page 145: Service Adjustments

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this chapter 6 This section describes how to test and adjust the scanner. These tests are optional. You may use them to check the system for errors.
  • Page 146: Caster Brake/Swivel Function Adjustments

    ANUAL Section 6-2 Caster Brake/Swivel Function Adjustments 6-2-1 Brake Function Adjustment The LOGIQ S6 contains front Brake/Swivel lock adjusters. The rear swivel lock function can NOT be adjusted. NOTICE Front Brake/Swivel Lock Adjuster Figure 6-73 Locations of Brake/Swivel Lock Adjusters...
  • Page 147 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-2-1-1 Preparation 1.) Remove the followings to access the brake/swivel lock adjusters. Crescent cover (two rubber caps and screws) Crescent Cover Bracket (two screws) Crescent cover Crescent bracket...
  • Page 148 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-2-1-2 The Pedal does NOT lock even if the pedal is depressed 1.) Loosen the left and right adjuster lock bolts. Adjuster Lock Bolt (each) Figure 6-76 The Left and Right Adjuster Lock Bolts...
  • Page 149 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-2-1-2 The Pedal does NOT lock even if the pedal is depressed (cont’d) 2.) Rotate the right adjusting bolts two turns as shown. Perform this procedure at the left adjusting bolt.
  • Page 150 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-2-1-3 Caster lock/swivel does NOT lock even if the pedal is locked in position (cont’d) 2.) Rotate the right adjusting bolts two turns as shown. Perform this procedure at the left adjusting bolt.
  • Page 151: Reloading The Probe Data

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 6-3 Reloading the Probe Data This procedure will be used for troubleshooting when a image sensitivity is NOT recovered. For details, refer to Section 7, Troubleshooting. This rewrites the probe data (probe delay data for beam forming) saved on the flash memories of DDBF 64 (DDBF128) and BTRAP boards.
  • Page 152 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 6-3 Reloading the Probe Data (cont’d) 5.) Enter the proper password to enter the maintenance mode. 6.) Click on OK. Figure 6-82 Password Entry 7.) Click on maintenance.
  • Page 153 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 6-3 Reloading the Probe Data (cont’d) 8.) Click on Exit to window. Figure 6-84 Exit to Window 9.) The Windows Desktop screen is displayed. Figure 6-85 Windws Desktop 10.)Erase the file, “D:\Pegasus\target\resources\Pegasus\ProbeInfo.res.”...
  • Page 154: Monitor And Lcd Adjustments

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 6-4 Monitor and LCD Adjustments 6-4-1 CRT Monitor Contrast and Brightness Adjustment Before calibrating the Monitor, check your DGVIC Part Number on the Utility Screen. The typical setting value changes with its Part Number.
  • Page 155 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-4-1 CRT Monitor Contrast and Brightness Adjustment (cont’d) 2.) Press the Toggle button (1) for contrast and brightness. Confirm that the contrast (or brightness) indicator is displayed on the monitor. If the brightness is displayed, press the toggle button again.
  • Page 156: Inch Ips-Pro Type Lcd Monitor (5261237) Brightness Adjustment

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 17 inch 6-4-2 IPS-Pro type LCD Monitor (5261237) Brightness Adjustment (BT08 or later and LS6 17 inch IPS-Pro type LCD Upgrade systems) 6-4-2-1 Adjusting Parameters There are 4 parameters to control the brightness of the IPS-Pro type LCD monitor. This section describes the property and adjustment for each parameter.t.
  • Page 157 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-4-2-1 Adjusting Parameters (cont’d) Adjust the Contrast value 6.) Adjust the Contrast value by pressing button (2) and (3). Exit the Contrast Menu 7.) Press the button (1) to go back to the Color menu.
  • Page 158 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-4-2-1 Adjusting Parameters (cont’d) 6-4-2-1-4 Gamma Range: 2.0/2.1/2.2 /2.3/ 2.4 /2.5/ 2.6 Property: Controls the level of transition from black to white from “Rapid” to “Gradual” as the value increases.
  • Page 159: Lcd Monitor (5169935) Contrast And Brightness Adjustment

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL LCD Monitor 6-4-3 (5169935) Contrast and Brightness Adjustment To adjust the contrast and brightness: 12.)Press the Toggle button (1) Once for brightness adjustment. Figure 6-90 Adjustment Buttons a.) Verify that Brightness is 55.
  • Page 160 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-4-4 LCD Touch Panel Adjustment This adjustment must be performed when always replacing: PC Box Assy Keyboard Assy Cable between LCD unit and PC Box 1.) Check the followings:...
  • Page 161 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-4-4 LCD Touch Panel Adjustment (cont’d) 1.) Remove the OP side R Cover to access. 2.) Access to the OSD Menu display button located at backside of the LCD.
  • Page 162: Lcd Touch Panel Adjustment

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-4-4 LCD Touch Panel Adjustment (cont’d) 3.) OSD Menu is shown on the LCD screen. Contrast Adjust Menu Color Adjust Menu Adjust Menu Brightness Position Control Menu Auto Adjust...
  • Page 163 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-4-4 LCD Touch Panel Adjustment (cont’d) 4.) Move the cursor to execute Auto Adjust. 5.) Make sure that the Geometry is selected, then the Menu Selection Button. Figure 6-94 Menu Seliction Button 6.) LCD display Auto Adjustment procedures will be started automatically.The following screen...
  • Page 164 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-4-4 LCD Touch Panel Adjustment (cont’d) 7.) After finishing the Auto Adjustment procedures, the following menu is displayed on the LCD. Figure 6-96 8.) Make sure that the LCD displacement is fixed and shown at the center location.
  • Page 165: Cleaning The Trackball

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 6-5 Cleaning the Trackball 1.) Power OFF the scanner. 2.) Place your fingers onto the notches of the trackball retainer ring. 3.) Rotate the retainer ring counterclockwise until it can be removed from the keyboard.
  • Page 166 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 6-5 Cleaning the Trackball (cont’d) 5.) Wipe off any oil or dust from the trackball using a cleaner or dry cloth. 6.) Wipe off any oil or dust from the trackball housing, rollers, and small ball, using a cleaner or cotton bud.
  • Page 167 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 6-5 Cleaning the Trackball (cont’d) 7.) Wipe off any oil or dust from the two rollers using a cleaner or cotton bud. When cleaning the roller, make sure not to spill or spray any liquid into the trackball housing (Keyboard NOTICE or system).
  • Page 168 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 6-5 Cleaning the Trackball (cont’d) 8.) Insert the trackball into the housing. 9.) Place the trackball and inner retainer into the housing with its stopper facing down.Lift off the inner retainer and trackball from the keyboard.
  • Page 169: Jumper And Dip Switch Setting

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 6-6 Jumper and Dip Switch Setting 6-6-1 Dip Switch Setting Dip switche setting represents the version and revision of the board. This section describes how to read the dip switches in case that a service engineer might need to confirm and communicate the setting of the dip switches.
  • Page 170 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-6-1 Dip Switch Setting (cont’d) Bit Setting for DIP Switch 5 to 8 (Upper 4 bits) 1 (h) 2 (h) 3 (h) 4 (h) 5 (h) 6 (h)
  • Page 171: Jumper Setting

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-6-2 Jumper Setting 6-6-2-1 HDD (Parallel Type) Always set the jumper between MASTER pins. Master Figure 6-106 HDD (Parallel Type) Jumper 6-6-2-2 HDD (SATA Type) Always set the jumper between “1.5Gb/s operation” pins.
  • Page 172: Mitsubishi Md-3000 Vcr Dip Switch Setting

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-6-3 Mitsubishi MD-3000 VCR Dip Switch Setting This section describes Mitsubishi MD-3000 VCR communication setting. Field Service Engineer (FSE) is not allowed to modify dip switch setting. Following setting is documented for purpose of troubleshooting by FSE.
  • Page 173: Cp30D Dip Switch Setting

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-6-3-2 Accessing Dip Switch Module for inspection (cont’d) 2.) Remove the RS232C board from the VCR to access the dip switch. RS232C Board Dip Switch Figure 6-110 Removing Board...
  • Page 174: Daylight Saving Time (Dst) - New Dates

    Issue 1: All Ultrasound products in USA, Canada, Australia, Egypt, and Sri Lanka. Note: This list may be incomplete. Check your local standards for changes to Daylight Saving Time (DST) start and end dates. • Issue 2: All LOGIQ S6 systems 6-7-1 Problem Issue 1: DST start and end dates have changed in some countries.
  • Page 175 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 6-7-4-2 To check or set the system time 1.) From the touch panel, press Utility -> System. 2.) Select Date/Time. 3.) On the Date and Time Properties screen, select the Date & Time tab.
  • Page 176 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 6-32 Section 6-7 - Daylight Saving Time (DST) - New Dates...
  • Page 177: Diagnostics/Troubleshooting

    GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
  • Page 178 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. Section 7-1 - Overview...
  • Page 179 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems. Table 8-60 Contents in Chapter 8...
  • Page 180 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-1-3 Electrostatic Discharge (ESD) Prevention DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE WARNING WARNING NECESSARY ESD PRECAUTIONS: Always connect yourself, via an arm-wrist strap, to the dedicated ground point located on the rear of the scanner (to the left of the power connector) or a proper frame ground.
  • Page 181 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 8-2 Software Loading Procedure Guide According to your system configuration, select the proper section for software loading: Table 8-61 Software Loading Procedure Guide BTxx BT06 BT08 Release Version R6.1.2...
  • Page 182 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. Section 8-2 - Software Loading Procedure Guide...
  • Page 183 ANUAL Section 8-3 Software Loading Procedures (R7.7.x or later with BEP4) This instructions describe the LOGIQ S6 R7.7.x Application Software installation procedures with BEP4. 8-3-1 General This procedure describes re-loading of the base system load and application load. This procedure is NOT intended for upgrades.
  • Page 184 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-4 Software Loading Steering Guide This section and the following pages should provide detailed guide of the system and application installation procedures. Installing OS + Application 8-3-5 PRE-INSTALLATION...
  • Page 185 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-4-1 Processes prior to R7.7.x Installation Important Item Caution Item Check SW Power OFF POWER POWER Save Customer Level USB Devices Setting Figure 8-113 Pre-installation Procedures *1: Loading R7.7.x application software procedures requires, at minimum, C: Drives to be ghosted (as NOTICE opposed to full ghost).
  • Page 186 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-4-3 After installation of R7.7.x Figure 8-115 Post Installation Procedures *3: Verify that Probes are not connected to the scanner. Other peripheral devices such as printers NOTICE should be turned ON at this time.
  • Page 187 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-5 Pre-installation Procedures Loading R7.7.x application software procedures requires basically C drives to be ghosted. In CAUTION this case, patient image data, preset parameters, and customer data will be kept. However , in case full ghost will be required, you MUST back up before starting this procedure as follows.
  • Page 188 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-5-4 Saving Connectivity Before starting upgrading, write down the Connectivity Setting for back-up. 1.) Touch Utility > Connectivity on the touch panel and click the TCP/IP tab. 2.) Write down the following parameters: Remove the Check mark.
  • Page 189 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-5-5 Software Option 1.) Touch Utility > Admin. 2.) Click on System Admin tab and write down HW Number and Installed Option Keys.. Figure 8-118 System Admin Screen...
  • Page 190 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-5-6 Printer Registration 1.) Select Utility > Connectivity > Button. 2.) Write down the printer names which are registered to Print1 through Print4. 3.) Click on >> to add Printflow View.
  • Page 191 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-6 Base System Installation (Ghost) and Application Load This section describes actual system installation and application loading procedures. NOTE: It takes approx. 25 minutes to complete software loading. It is longer than software loading to the BEP3 installed system.
  • Page 192 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-7 Install Base System (cont’d) 4.) The following screen appears. Press B then Enter to initialize the C drive If you select [A], all the archive would be lost.
  • Page 193 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-7 Install Base System (cont’d) 5.) When complete Ghost installation, the following screen appears. Eject the Base System Software Load Image DVD from the drive. The Base System Software Load Image DVD must be stored onto the PC box cover.
  • Page 194 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8 Modify System/Windows Settings After Base System is successfully installed, various parameters must be properly set before proceeding to Application load. The below outlines necessary system/windows setting. Table 8-66 Summery of modification...
  • Page 195 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-1 UPS Setting UPS setting is done by Ghost automatically and no manual setting is necessary. The following window appears. Figure 8-124 Checking Power Supply Click on OK.
  • Page 196 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-2 Drive Letter Setting The following message appears . Figure 8-126 Auto Assign 1.) Click on OK 2.) Check the Drive Letter is set as below from Windows Explorer.
  • Page 197 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-3 Matrox for CRT (Resolution setup) For CRT Monitor System ONLY: 1.) Drag and drop the taskbar into the desktop. Figure 8-127 Drag and Drop Task bar 2.) Click on << in the task tray to show the MatroxPowerDesk-SE icon.
  • Page 198 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-3 Matrox for CRT (Resolution setup) (cont’d) 4.) Select Multi-Desplay Setup... Figure 8-130 Multi-Display Setup... 5.) When the setting dialog is displayed, click on Use advanced Matrox display controls option, then click on Independent mode.
  • Page 199 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-3 Matrox for CRT (Resolution setup) (cont’d) 9.) Click on Yes button on the confirmation dialog. Figure 8-132 Yes 10.)Right-click on the Desktop and select Properties from the popup menu.
  • Page 200 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-3 Matrox for CRT (Resolution setup) (cont’d) 14.)Click on the Advanced button, then click on the Monitor tab on the Advanced Settings dialog. 15.)Verify that Screen refresh rate is 75Hertz.
  • Page 201 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-4 Hardware Device Check: 1.) On the windows desktop, left-click on My Computer, then right-click on Properties. Figure 8-135 My Computer Properties 2.) The System Properties window appears. Select Hardware > Divice Manager...
  • Page 202 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-4 Hardware Device Check: (cont’d) 4.) If X or ! is displayed: Left-click on the device with x mark, then rigtht-click on Enable. Left-click on the device with ! mark, then rigtht-click on Properties to perform the device driver reconfiguration.
  • Page 203 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-5 LAN Parameter Setting (cont’d) 5.) When the following window appears, right-click on Cancel. Figure 8-139 Canceling Local Area Connection Properties 6.) Close all windows. Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
  • Page 204 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-6 TCP/IP Filter Setup 1.) On the desktop, left-click on My Network Places, then right-click on Properties... 2.) Double-click Local Area Connection or Local Area Connection #2 (if you do NOT find "Local Area Connection").
  • Page 205 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-6 TCP/IP Filter Setup (cont’d) 4.) Click Advanced... button. Figure 8-142 Go to Advanced Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4) 8-27...
  • Page 206 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-6 TCP/IP Filter Setup (cont’d) 5.) Click Options tab. Figure 8-143 Options tab 8-28 Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
  • Page 207 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-6 TCP/IP Filter Setup (cont’d) 6.) Select TCP/IP Filtering, then click Properties button. Figure 8-144 TCP/IP Filtering Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
  • Page 208 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-6 TCP/IP Filter Setup (cont’d) 7.) If your system has no TCP port numbers as follows, perform these procedure to add the TCP port numbers. If the TCP port already have been set, skip these procedures. (This depends on the network card installed in the PC box.)
  • Page 209 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-6 TCP/IP Filter Setup (cont’d) 8.) Verify that the proper TCP port numbers are entered in TCP Ports box, then click OK buttons for all opened window to complete the TCP/IP Filter Setup.
  • Page 210 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-7 Date and Time Setting 1.) Select Start > Settings > Control Panel > Date and Time.I 2.) Click on Time Zone tab. Figure 8-148 Date & Time 3.) Select the appropriate time zone.
  • Page 211 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-8-7 Date and Time Setting (cont’d) 5.) Set the current date and time. 6.) Click on Apply. Month and Year Time Figure 8-150 Setting time 7.) Click on OK to close the Date and Time window.
  • Page 212 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-9 Installing R7.7.x Application Software You must install BT08 Base image before Application Software installation. If not, Application shows the NOTICE error message, “Service platform is out of date. You must reghost the C drive and reinstall the software.”...
  • Page 213 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-9-1 Installing R7.7.x Application Software (cont’d) 4.) Double-click on LoadSoftware.bat. Figure 8-152 LoadSoftware.bat 5.) The Installation program will start, and the command prompt screen will be shown on the monitor.
  • Page 214 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10 System Settings After application is successfully installed, various LOGIQ™ S6 system parameters must be properly set. Table 8-67 System Setting Summery Item Method Proper Value Software Version...
  • Page 215 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Table 8-67 System Setting Summery Item Method Proper Value Language (For Japanese Utility Only) \ System \ General \ Language \ Regional Option \ Regional Option (tab) \ Standard and formats...
  • Page 216 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-1 Confirmation of the software version 1.) Touch the Utility button on the Touch Panel. Press Utility button. Figure 8-154 Utility NOTE: The Operator Login window appears. Select adm (Administrator) for the Operator field.
  • Page 217 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-1 Confirmation of the software version (cont’d) 2.) Touch the System button on the Touch Panel. Press System button. Figure 8-156 System 3.) Click on About tab. The software version will be shown on the monitor as shown. Confirm the following: Software Version (R7.7.x)
  • Page 218 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-2 Computer Name and Connection Settings 1.) Touch Utility > Connectivity. 2.) Click on TCP/IP. 3.) Type Computer Name. The computer name is unique. Check the scanner serial Number printed on the label located at the rear NOTICE lower side of the scanner.
  • Page 219 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-2 Computer Name and Connection Settings (cont’d) 7.) Touch Scan on the Touch Panel to return to the scan screen. 8.) Shut down the scanner. 9.) Power ON the scanner, then press Patient key and verify that the following screen appears.
  • Page 220 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-3 Check LCD Monitor Hardware Setting 8-3-10-3-1 Identify LCD Type before adjusting the monitor There are two types of LCD panel, Conventional LCD: GA500 and New LCD: GA500C. To distinguish the LCD type, confirm if the “Identification Label”...
  • Page 221 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-3 Check LCD Monitor Hardware Setting (cont’d) To adjust the contrast and brightness: 1.) Press the Toggle button (1) Once for brightness adjustment. Figure 8-162 Monitor Buttons a.) Verify that Brightness is 55.
  • Page 222 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-3-3 For new LCD Monitor: GA500C To determine the LCD type, refer to 8-3-10-3-1. NOTICE To adjust the brightness: 1.) Press the Toggle button (2) or (3) Once for brightness adjustment.
  • Page 223 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-5 VCR Parameters Setting (for the VCR-equipped system ONLY) 1.) Touch Utility > System. 2.) Click on Peripherals tab. 3.) Set the proper parameters: PAL or NTSC VCR type Figure 8-165 Peripherals 4.) Shut down the scanner, then restart the scanner.
  • Page 224 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-6 Regional Setup (For the system with language other than English ONLY) For Japanese Languages You can set up the keyboard for Japanese. If you select Japanese, the following items display or enter by Japanese.
  • Page 225 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-6 Regional Setup (For the system with language other than English ONLY) (cont’d) Check Do not Change English. Figure 8-166 Regional and Language Options 6.) In Utility> System> General, set the Language as JPN. Save this setting, but do not reboot the system yet.
  • Page 226 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-6 Regional Setup (For the system with language other than English ONLY) (cont’d) 7.) Press Regional Options, under Standards and Formats select Japanese, under Location select Japan. Press Apply.
  • Page 227 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-6 Regional Setup (For the system with language other than English ONLY) (cont’d) 9.) Select the Advanced tab, then select Japanese in the Language for non-Unicode programs pull- down menu.
  • Page 228 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-6 Regional Setup (For the system with language other than English ONLY) (cont’d) 15.)To type Japanese, press Alt+ (right side of the Escape key) to start IME (Input Method Editor).
  • Page 229 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-6 Regional Setup (For the system with language other than English ONLY) (cont’d) For Russian or Greek Languages 1.) In Utility--> System--> General, set the Language as Russian or Greek. Save this setting, but do not reboot the system yet.
  • Page 230 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-6 Regional Setup (For the system with language other than English ONLY) (cont’d) 3.) Select the Language tab, press Details, under Installed Services select the Russian or Greek keyboard, under Default input language select Russian - Russian or Greek - Greek, press Apply, Press OK.
  • Page 231 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-6 Regional Setup (For the system with language other than English ONLY) (cont’d) For non-Russian / Greek Languages 1.) Press Regional Options, select the Language tab, press Details, and comfirm default input Languarge to English (United States) International.
  • Page 232 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-6 Regional Setup (For the system with language other than English ONLY) (cont’d) 2.) Select the Advanced tab, then select the language in the Language for non-Unicode programs pull- down menu.
  • Page 233 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-7 Regional Setup (For the system with English language only) 1.) Go to Utility > System > General. 2.) Press Regional Options button. Figure 8-176 Regional Options 3.) The Regional and Language Options screen appears.
  • Page 234 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-7 Regional Setup (For the system with English language only) (cont’d) 5.) The Text Services and Input Languages window pops up. On BT08 or later version, Default input language becomes English [United States]- United States-International every time after loading the application software.
  • Page 235 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-7 Regional Setup (For the system with English language only) (cont’d) 6.) If Default input language was English [United States]-US before installing the application, change the setting to English [United States]-US and press OK to close the window..
  • Page 236 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-8 Printer Registration 1.) Verify that the power switches of the printer and scanner are turned ON. 2.) Select Utility > Connectivity > Button. 3.) Select the Printer1 button and select the user defined printer which has been recorded in...
  • Page 237 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-8 Printer Registration (cont’d) 5.) Verify that the user defined printer is added in Printflow View.(Example shown: UP-D23MD) 6.) For UP-D23: When upgrading to BT08, updated printer driver may be loaded and assigned. Before assigning Print Button to Printer, check 'DEVICE' Tab for multiple UP-D23 Printer, and check for active printer.
  • Page 238 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-9 Regional Preset For the system other than "NONE" selected as Regional Preset, resetting MUST be required. NOTICE 8.) Touch Utility > System. 9.) Click About tab. 10.)Click Additional About Information.
  • Page 239 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-9 Regional Preset (cont’d) 14.)Click the button Reload **** Preset corresponding you noted. Figure 8-184 Reload buttons 15.)Click OK on confirmation dialog box. Figure 8-185 OK Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
  • Page 240 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-10-9 Regional Preset (cont’d) 16.)Check Preset Region (Requires reboot): is right region you selected. Figure 8-186 Preset Region 8-62 Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
  • Page 241 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-11 Functional Check 8-3-11-1 Service Platform 1.) Click on Utility > Service. It will take about ten (10) seconds for activating. Figure 8-187 Service 2.) Make sure that CAPS is OFF (should be dimmed) for password entry performed later.
  • Page 242 Service Platform has failed. Reload the application software or Base System Software (OS) + application software. Contact a Technical Support for details. 4.) Select GE Service at the “Select User Level” field. 5.) Enter the password for the Service Platform.
  • Page 243 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 8-3-11-2 Functional Checks for Probe Recognition 1.) Touch Scan button on the Touch Panel to return to the scan screen. 2.) Connect each probe to ensure that they are recognized.
  • Page 244 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 8-66 Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
  • Page 245 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Renewal Parts for LOGIQ™ S6. Table 9-68 Contents in Chapter 9...
  • Page 246 BT08 with BECOMP4: Equipment Models Covered in this Chapter Table 9-69 Material List Part Name Part Number Quantity Description LOGIQ S6 R7.7.0 Application Software 5315082-2 LOGIQ S6 Ghost CD for BECOMP4 5305207 LS6 M12L Enable Option 5175356 LS6 M7C Enable Option 5175187 LS6 M3S Enable Option 5175946...
  • Page 247 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Table 9-69 Material List Part Name Part Number Quantity Description PDF CREATOR SOFTWARE LICENSE 2394132 OP CSL TYPE_C JPN LS6 5324059 100V NTSC, Standard Height OP CSL TYPE_C USA-ASIA-CALA120...
  • Page 248 R800 common software for both upgraded BT08 and 5315082-2 pure BT08 Base image (Ghost) for pure BT08 (BEP4 system) LOGIQ S6 Ghost DVD for BECOMP4 5305207 only LOGIQ S6 Ghost CD for BEP-NP, Base image (Ghost) for upgraded BT08 (BEP3 5324919 R7.7.0...
  • Page 249 BT08, LCD: Equipment Models Covered in this Chapter Table 9-71 Material List Part Name Part Number Quantity Description LOGIQ S6 R7.6.0 Application Software 5269591-3 LOGIQ S6 Ghost CD for BEP-NP, 5252046 R7.6.0 LS6 M12L Enable Option 5175356 LS6 M7C Enable Option 5175187 LS6 M3S Enable Option...
  • Page 250 IRECTION EVISION ERVICE ANUAL Table 9-71 Material List Part Name Part Number Quantity Description LOGIQ S6 Software Option Installation 5173830 Instruction PDF CREATOR SOFTWARE LICENSE 2394132 OP CSL TYPE_B JPN LS6 5268701 100V NTSC, Standard Height OP CSL TYPE_B USA-ASIA-CALA120...
  • Page 251 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 9-4-2 BT08 with BECOMP3 Parts Figure 9-192 BT08 with BECOMP3 Parts Table 9-72 BT08 with BECOMP3 Parts Item Part Name Part Number Description for forward production of BT08 or later...
  • Page 252 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Table 9-72 BT08 with BECOMP3 Parts Item Part Name Part Number Description for forward production of BT08 or later. To be used PCI-DGVIC2 Assembly 5301220-2 with BEP 5193189.
  • Page 253 220V PAL, Standard Height JAPAN 5165990 ASIA220V (Standard) 5165990-3 USA (Standard) 5165990-7 ASIA220V (Tall) 5165990-10 USA (Tall) 5165990-2 KOREA 5165990-4 EU (Standard) 5165990-8 ASIA120V 5165990-5 EU (Tall) 5165990-9 CHINA 5165990-6 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 254 220V PAL, Standard Height JAPAN 5165848 ASIA220V (Standard) 5165848-3 USA (Standard) 5165848-7 ASIA220V (Tall) 5165848-10 USA (Tall) 5165848-2 KOREA 5165848-4 EU (Standard) 5165848-8 ASIA120V 5165848-5 EU (Tall) 5165848-9 CHINA 5165848-6 9-10 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 255 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 9-5-2 Plastic Covers Figure 9-193 Plastic Covers Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-11...
  • Page 256 (for standard) Front Bottom Curtain Long 5162514 without plate (for tall) GEL HOLDER L6 5168986 OP REAR COVER TALL L6 5162515 For Tall PROBE HOLDER BRACKET L6 5147525-2 9-12 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 257 USER SW ASSY OF MONITOR 2297050 microphone and switch MON-CABLE-ASSY 2304171 cable NECH-PIPE 5147551 NECK ASSY 2284225 Grill, Left 2279675 Grill, Right 2279676 CRT CONN ASSY 5155394 CB47+CB48+CB49 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-13...
  • Page 258 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 9-5-4 LCD Monitor Figure 9-195 LCD Monitor 9-14 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 259 LCD SPK MIC CABLE CSL L6 5159753 CB57 ARM Assy 5160392 17inch IPS-PRO Type cannot be used. ARM COVER 5170282 NECK PIPE FOR LCD 5160391 ARM LOCK Assy 5170278 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-15...
  • Page 260 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 9-5-5 OP Panel and Keys Figure 9-196 OP Panel and Keys 9-16 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 261 For use with R6.x.x Systems only KBD FOR LCD (S6 LCD) 5170397-2 Full Compatibility (R6/R7) for LCD System SWEDISH AN KEYTOP OPTION 5193888 NORWEGIAN DANISH AN KEYTOP 5193889 OPTION Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-17...
  • Page 262 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 9-5-6 Front End Assemblies Figure 9-197 Front End Assemblies 9-18 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 263 BACK PLANE 5138460 Tools to remove a board in the NEST Assy (Not NEST Board JIG Assy 2315492 Shown) Tx CABLES Assy CSL L7 2305111-3 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-19...
  • Page 264 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 9-5-7 Back End Assemblies Figure 9-198 Back End Assemblies 9-20 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 265 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL LS6 BT08 BECOMP Cable Connection Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-21...
  • Page 266 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL LS6 BECOMP Cable Connection (For backward production before BT08, including the software, LCD field ugprade system to BT08 or later) 9-22 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 267 (including the software,LCD field upgrade system to BT08 or later) for backward LCD production before BT08 PCI VGA (Not Shown) 5121039 (including the software,LCD field upgrade system to BT08 or later) Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-23...
  • Page 268 PROTECTOR, 100V, 120V U0026EK AC EXTERNAL OUTLET Assy 5152282 PEAR BOARD Assy 5149653 REAR CONNECTOR CASE L6 5147515 MAB Assy 5168873 PROTECTOR 220V U0047EK REAR Assy 220V 5170230 9-24 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 269 Board + Filter + Plate + Cable LHV - BP CABLE Assy CSL L6 5168928 Cable Assy + Plate AC PANEL CABLE Assy CSL L6 5168815 Cable + Plate Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-25...
  • Page 270 HV - TRICON CABLE CSL L6 5152274 FUSE 2315585 SSR CABLE Assy CSL L6 5152285 FAN CONNECT CABLE CSL L6 5159703 AC INLET Assy 220V 5169695 FUSE HOLDER CAP 5118614 FUSE HOLDER 5118613 9-26 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 271 (Not Shown) Caster Total Lock 5233774 (Not Shown) Front Pedal Assy 5172590 with 2 lock pin Rear Pedal Assy 5172617 with 1 lock pin Front Crescent Pedal Cover 5167902 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-27...
  • Page 272 Table 9-83 Other Mechanical Assembles Item Part Name Part Number Description Front Pedal Cover (Park) 5172857 Front Pedal Cover (Release) 5172962 Caster Assy 3 2381035-3 Caster Set (Free:1, Swivel:1, Lock:2, Screw:16) 9-28 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 273 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 9-5-11 Misc. Cables Figure 9-202 Misc. Cables Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-29...
  • Page 274 CB25 Foot SW cont CN25B CON1 HD-LED Power Switch CB24 CN24B OP Panel DC in CB23 CN23B CB22 CN22B USB_B2 B/W printer CB21 CN21B ACin CB45 CN45B USB_B1 9-30 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 275 CB25 Foot SW cont CN25B CON1 HD-LED Power Switch CB24 CN24B OP Panel DC in CB23 CN23B CB22 CN22B USB_B2 B/W printer CB21 CN21B ACin CB45 CN45B USB_B1 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-31...
  • Page 276 Foot SW VCR2 VCR1 Service BACK PLANE- BACK REARIO PLANE BACK PLANE- CB34 BACK PLANE- To BEP CN34B BECOMP 4DCtrl HVCtrl HVHin HVLin Tx Cable LVin LVsense PCIBridge MDBRG 9-32 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 277 KEYBOARD AC OUT CABLE CLS L6 5152276 CB42 MAB CABLE Assy CLS L6 5152278 CB26+CB29+CB31 MAB - OP CABLE Assy CSL L6 5152279 CB24+CB25 OP PANEL CABLE Assy CSL L6 5152280 CB22+CB23 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-33...
  • Page 278 USB LAMP CABLE CSL L6 5152286 CB41 MAIN MONITOR CABLE Assy CSL L6 5155276 CB28+CB30 LCD DVI CABLE BOTTOM CSL L6 5159750 CB54 CRT VGA CABLE CSL L6 5165932 CB27 9-34 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 279 Accessories for Parts Setting 5121042 (Not Shown) AC Adapter for Wireless Receiver 5122602 (Not Shown) AC Cable for Wireless Kit LOGIQ7 5121762 (Not Shown) USB Extension Cable LOGIQ7 5121749 (Not Shown) Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06) 9-35...
  • Page 280 5123380 (Not Shown) LOGIQ7 Battery Charger for JPN LOGIQ7 5123382 (Not Shown) LQ-MD800P 5120592 DVD Recorder, 120V AC (PAL/NTSC compatible) LQ-MD800E 5120593 DVD Recorder, 220V AC (PAL/NTSC compatible) 9-36 Section 9-5 - Renewal Parts List for Original LOGIQ S6 (BT06)
  • Page 281 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 9-6 Probes Refer to Section 3-4 - Completing the Installation for available probe list. Section 9-6 - Probes 9-37...
  • Page 282 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 9-38 Section 9-6 - Probes...
  • Page 283 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Chapter 10 Periodic Maintenance Section 10-1 Overview 10-1-1 Purpose of Chapter 10 This chapter describes Periodic Maintenance (PM) on the scanner and its peripherals. These PM procedures are intended to maintain the quality of the ultrasound systems performance. Read this chapter completely and familiarize yourself with the procedures before starting a PM.
  • Page 284 Your GE Service Representative knows your LOGIQ™ S6 best and can provide competent, efficient service. Please contact us for further information and to schedule GE Healthcare Ultrasound to perform this service for you.
  • Page 285 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-3-1How often should PMs be performed? (cont’d) Table 10-2 Periodic Maintenance Schedule Facilities Service at Indicated Time Daily Weekly Monthly Program Notes •* Clean Probes * or before each use •...
  • Page 286 Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs. Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID...
  • Page 287 EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal...
  • Page 288 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-4-2 Special Tools, Supplies and Equipment 10-4-2-1 Specific Requirements for Periodic Maintenance See Chapter 7 Table 10-5 Overview of Requirements for Periodic Maintenance Tool Part Number Comments Digital Volt Meter (DVM)
  • Page 289 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 10-5 System Periodic Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-6 System Preliminary Checks...
  • Page 290 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary. 10-5-2-1 System Checks Table 10-7 System Functional Checks ÷...
  • Page 291 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description Verify record/playback capabilities of the VCR.
  • Page 292 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-5-4-2 Air Filter Cleaning Table 10-11 Air Filter Cleaning - frequency varies with your environment Step Item Description Remove Filter Refer to Chapter 8 for air filter location and removal instructions Cover The filters can be cleaned in sprinkling water, or they can be dusted with a vacuum cleaner.
  • Page 293 The system user manuals and various probe handling cards provide a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details.
  • Page 294 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 10-7 Electrical Safety Tests 10-7-1 Safety Test Overview The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards.
  • Page 295 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-2 GEMS Leakage Current Limits The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed.
  • Page 296 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
  • Page 297 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-4 Grounding Continuity Electric Shock Hazard. The patient must not be contacted to the equipment during this test CAUTION Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
  • Page 298 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-4-2 Dale 600 - Ground Continuity The Dale 600 measures line cord resistance from the third pin of the attachment plug to the meter’s Chassis Cable clamp. Test the grounding continuity of the system to all exposed metal parts in accordance with the IEC 60601-1 procedure as above.
  • Page 299 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-5 Chassis Leakage Current Test 10-7-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
  • Page 300 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-5-3 Dale 600 Meter Procedure When measuring system chassis currents with the Dale 600, always use the CHASSIS selection of the external/chassis function switch. This requires the ground clip lead and changing the meters switches in accordance with the IEC 60601-1.
  • Page 301 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-5-4 Data Sheet for Chassis Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-14. Record all data on the PM Inspection Certificate.
  • Page 302 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-6-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the operating controls such as the lead switch should be operated to find the worst case condition.
  • Page 303 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-6-3 Dale 600 Meter Procedure (cont’d) Patient Cable Figure 10-1 ECG Leakage Current Test 5.) Set the meter's “FUNCTION” switch to LEAD TO GROUND position to measure the patient lead to ground leakage current.
  • Page 304 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-6-3 Dale 600 Meter Procedure (cont’d) 7.) Record the patient lead to ground leakage current measured on the data sheet. 8.) Set the meter's “FUNCTION” switch to LEAD TO LEAD position to measure the lead to lead leakage current.
  • Page 305 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the PM Inspection Certificate.
  • Page 306 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-9 Probe Leakage Current Test 10-7-9-1 Definition This test measures the current that would fGw to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
  • Page 307 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-9-2 Test Equipment (cont’d) Table 10-24 Test Equipment and Accessory Description Dale Part number Accessory Name Picture Description Used on DALE601/601E to measure Vendor MODEL earth resistance and enclosure...
  • Page 308 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-9-3 Generic Procedure for Leakage Current Measurements should be made with the ground open and closed, with power line mains polarity normal and reversed, and with the unit Off and On. For each combination, the probe must be active to find the worst case condition.
  • Page 309 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-9-4 Meter Procedure Using Dale Meter to Measure Leakage Current (cont’d) 11.)Keep a record of the results with other hard copies of Planned Maintenance data. Equipment damage possibility. Never switch the Polarity or the status of the Neutral when the CAUTION Ultrasound unit is powered on.
  • Page 310 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-9-5 Meter Procedure Using Probe Adapter to Measure Probe Isolation (Sink) Current The Dale 600/600E provides a method for testing probes independently from the system. The meter utilizes a probe adapter to apply a test potential commonly to all connector pins.
  • Page 311 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL 10-7-9-5 Meter Procedure Using Probe Adapter to Measure Probe Isolation (Sink) Current (cont’d) 8.) Test every probe and record in Table 10-26 the test results for every probe.
  • Page 312 EVISION ERVICE ANUAL GE Healthcare Leakage Current Limits for LOGIQ™ S6 The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. Measurement limits per IEC 60601-1 Medical Equipment Safety Standards, Table IV. Table 10-27 Chassis Leakage Current Limits—Accessible Metal Surfaces...
  • Page 313 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL Section 10-1 When There's Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
  • Page 314 Applicable Software Options Probe Holders Applicable Hardware Options External I/O Control Panel Wheels, Brakes & Swivel Locks Monitor Cables and Connectors Touch Panel GE Approved Peripherals (VCR, CD-R, MOD, Printers) Measurement Accuracy GE Approved Peripherals COMMENTS: ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ 10-32...
  • Page 315 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe Chassis Source Leakage Current - Wheel...
  • Page 316 GE H EALTHCARE 5162630, R LOGIQ™ S6 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 10-34 Section 10-7 - Electrical Safety Tests...
  • Page 318 GE HEALTHCARE GE Healthcare: Telex 3797371 P.O. Box 414; Milwaukee, Wisconsin 53201, U.S.A. (Asia, Pacific, Latin America, North America) GE Ultraschall:Tel: +49 (0) 212 28 02 208 Deutschland GmbH & Co KG Beethovenstrabe 239, Postfach 11 05 60 D-42655 Solingen, Germany...

Table of Contents