Troubleshooting

General Troubleshooting Tips

The following general troubleshooting tips can be used to help diagnose problems not specifically
discussed elsewhere in this chapter.
Thoroughly inspect the equipment.
l
Disconnected or loose cables, missing hardware, and damaged equipment can cause what
may appear to be unrelated symptoms or equipment failure. For additional information, see
"Inspecting the Device" on page
Verify the equipment has not been modified.
l
Unauthorized modifications to the equipment may cause unexpected results, poor performance,
or system failure.
If the equipment has had unauthorized modifications, contact GE Healthcare Technical Support.
Verify the software has not been updated.
l
Updated software may change system functionality. If the user is unaware of the changes, they
may appear as unexpected results.
If the software has been updated, refer to the revised Operator's Manual to determine whether
the update changed features.
Verify the problem was not caused by operator error.
l