Troubleshooting
8
This section explains error signs and troubleshooting recommendations.
Recorder Troubleshooting
Trouble Indicators
The Event Button LED
flashes yellow once
every second.
An error tone sounds
when you try to start the
recording.
The Hookup Guidance
LED flashes yellow once
every second.
An error tone sounds
when you try to start the
recording.
2067634-077 Revision B
Cause
Battery power is low.
Electrode or ECG cable is not
connected.
SEER™ 1000
8
Recommendation
The battery is not fully charged
or the wrong battery type has
been inserted. Replace with
new alkaline battery or NiMH
rechargeable battery.
Connect the ECG cable or re-
connect loose electrodes.
83