Troubleshooting

System Errors

The table lists messages that you may encounter while using this device.
Message
WARNING: Approaching
<xx>% of ECG storage limit.
Transmit and delete reports to
free memory.
Memory is full. Transmit and
delete reports to free memory.
Memory is full. This ECG
cannot be saved.
Memory is full. New ECGs
cannot be saved.
Battery error. Attach power
cord. Contact Service.
Battery unknown error
5864335-001-1
Cause
The device storage capacity is
approaching 80% or 90%. This
occurs as patient reports are
added to the Files list.
The device storage capacity is
between 99% and 100%.
The device storage capacity is
full.
The device storage capacity is
full.
Power sensor failure
Battery capacity sensor
failure
An unexpected battery error
has occurred.
  MAC
5 A4/MAC
5 A5/MAC
13
Solution
Transmit patient reports to
configured destinations and
delete the transmitted reports
from the Files list to create
additional storage space.
Transmit patient reports to
configured destinations and
delete the transmitted reports
from the Files list to create
additional storage space.
Transmit patient reports to
configured destinations and
delete the transmitted reports
from the Files list to create
additional storage space.
Transmit patient reports to
configured destinations and
delete the transmitted reports
from the Files list to create
additional storage space.
Replace the battery.
Contact your GE Healthcare
Service Support representative
if the error persists.
Replace the battery.
Contact your GE Healthcare
Service Support representative
if the error persists.