Program Completed - Smeg BPW4260 User Manual

Bedpan washers
Table of Contents

Advertisement

On completion of the Reset, FAIL appears.
To eliminate the FAIL message without changing program, acknowledge it by pressing the confirm button
(P5), or select another program using buttons P4 and P6.
9.8

PROGRAM COMPLETED

Program completed successfully: the word END appears (not flashing) on the second line of the display.
Unlock the door with the "padlock" button (P7) to open the door and remove the load.
The cycle can be interrupted in advance at the end of the thermal disinfection phase (once the A
value set
0
for the cycle has been reached) to avoid having to wait for the cooling and hydraulic circuit pumpout phases
to be completed. To do this, press the "padlock" button (P7).
Once thermal disinfection has finished (from the start of the cooling phase, identified by "END" and "COOL"
alternating on the display) the door can be opened. There are 2 possible scenarios:
Case 1) Door opened with cooling (COOL) still in progress:
END – COOL are shown flashing
If the chamber temperature is higher than the temperature set for opening of the door, the chamber
temperature and the
symbol flash on the second line instead of "END" – "COOL": the door does not
unlock until button P5 is pressed.
User Manual
193909105 rev.04
BPW1260 – BPW4260
Page 55 – 107

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bpw1260

Table of Contents