WATER SUPPLY CONNECTIONS
Hot and cold water supply connections are necessary for installation of the device.
If hot water is not available, the user is responsible for checking that the device's performance is
adequate for their specific conditions of use.
FLAMMABLE SUBSTANCES
Do not place flammable substances inside the device. Do not use flammable detergents.
Do not place alcohol or solvents such as turpentine, which might cause an explosion, inside the
device. Do not place materials dirty with ash, wax or paint inside the device.
WARNINGS REGARDING SAFE DISPOSAL OF THE DEVICE AT END-OF-LIFE
At end-of-life the device may be contaminated (e.g. organic material may be present, especially if the
device has ended its working life due to a fault that has caused the failure of the last thermal disinfection
cycle): take appropriate care during decommissioning procedures, complying with the precautions set out
in the "PRECAUTIONS FOR ACCESSING AND REUSING THE DEVICE AFTER AN INCOMPLETE CYCLE AND FOR
ACCESS TO THE DEVICE AT END-OF-LIFE" section
Once decommissioned, the device must be rendered unusable. Cut the power supply cable after removing
the plug / disconnecting the cable from the power socket.

5.2.1 WARNINGS CONCERNING CONNECTION TO AN IT NETWORK

The device can only be connected to an IT network with installation of optional accessory BPW-
LANI, code 905561.
The communication protocol is proprietary and the device can only be connected to an IT network
using the application supplied by the manufacturer and accessed on the network via the
passwords, differentiated by level.
Differentiated passwords for authorised engineers only enable safer management of the
device's operating parameters.
The IT security of devices overall also implies correct behaviour on the part of staff who
interface with them and appropriate control of accesses, in line with the provisions of the MDCG
guidelines (especially Ref. MDCG 2019-16).
PC used for connection, minimum requirements: RJ45 port, Microsoft Windows 32 or 64bit
operating system, Windows 7 or higher with Framework 4.0 or higher installed and updated.
Health operators should adopt a risk management process in line with IT security best
practices in order to maintain overall security:
- physical security to prevent unauthorised physical access to medical devices or network access
points;
- access control measures (e.g. role-related) to guarantee that only authorised, identified
personnel can access network devices, stored data, services and applications;
- network access controls, such as segmentation, to restrict the communication of medical
devices;
- Malware protection to prevent the execution of unauthorised code;
- Training to raise awareness of security.
- ability to reliably establish who has made which changes to the system.