Current Output Check - Honeywell SMARTLINE RM77 Supplementary Instructions Manual

10 ghz radar (fmcw) level meter for distance, level, volume, flow, mass and reflection measurement of liquids
Table of Contents

Advertisement

SMARTLINE® RM77

8.5.2 Current output check

Do a check of the current output value: 4 mA
• Energize the device.
• Enter the Supervisor menu.
• Push [> > > > ], [ ] and [> > > > ] to go to menu item 2.2.1 SET OUTPUT. Push [> > > > ] and then push [ ] or [ ]
to move up or down the list to set the output to 4 mA.
The output will change to the set value after 5 seconds, as an alternative to the measured
i
value. The current output stays at this value while you are in this menu item.
• Measure the current output at the terminals with an ammeter. Does the ammeter show a
current output of 4 mA ± 400 µA?
• Push [^ ^ ^ ^ ] to go back to the menu list.
i
The current output goes back to the measured value.
• End of the procedure.
Do a check of the current output value: 12 mA
• Push [> > > > ] and then push [ ] or [ ] to move up or down the list in menu item 2.2.1 SET OUTPUT
to set the output to 12 mA.
i
The output will change to the set value after 5 seconds, as an alternative to the measured
value. The current output stays at this value while you are in this menu item.
• Measure the current output at the terminals with an ammeter. Does the ammeter show a
current output of 12 mA ± 400 µA?
• Push [^ ^ ^ ^ ] to go back to the menu list.
i
The current output goes back to the measured value.
• End of the procedure.
Do a check of the current output value: 20 mA
• Push [> > > > ] and then push [ ] or [ ] to move up or down the list in menu item 2.2.1 SET OUTPUT
to set the output to 20 mA.
i
The output will change to the set value after 5 seconds, as an alternative to the measured
value. The current output stays at this value while you are in this menu item.
• Measure the current output at the terminals with an ammeter. Does the ammeter show a
current output of 20 mA ± 400 µA?
• Push [^ ^ ^ ^ ] to go back to the menu list.
The current output goes back to the measured value.
i
• End of the procedure.
CAUTION!
If the values do not agree with the tolerances, then the data supplied by device will not be
correct. For a solution to the problem, speak to the supplier.