Honeywell RAEGuard 3 User Manual page 48

Fixed gas detector
Table of Contents

Advertisement

Password settings
Low-level password setting111
o
Advanced password settings: 888
o
Communication settings
Address: 001
o
Baud rate:
o
4800;9600;19200;38400;57600
HART address:00
o
Output crrent lvel
Fault mode
o
1.0 mA;2.0 mA;3.0 mA
Inhibit mode
o
1.0 mA;2.0 mA;3.0 mA
Warning mode
o
1.0 mA;2.0 mA;3.0 mA
Out of range
o
21.0 mA;22.0 mA;23.0 mA
4-20 mA Output calibration
4 mA output calibration
o
20 mA output calibration
o
Restore factory settings
Note: The blackened font in the table above is the factory default value.
Zero & Standard Gas Calibration: See section 8 for details.
Calibration time interval: This menu is to remind users to calibrate regularly.
The default value is 90 days. According to the actual situation, users can set
it in 0-360. If 0 is set, the reminder function will be turned off; if other values
are set, the detector will display
when the calibration interval time is reached.
Bump test: See section 8 for details.
Forced current output: 4 mA by default. It can force 4-20 mA current loop to
output the set current value between 0-25, and the increment and decrease
step is 0.5 mA. At the same time, the detector FAULT LED is always on,
indicating the "Inhibit" mode.
RAEGuard 3 User Manual
and the LCD backlight will be always on
48