Revision 1
ERR SUB A001 - Generator Error Messages
Code
Message text
071/082
E08 Power supply
failure
071/083
072/001
072/003
072/004
072/005
072/006
072/007
E06 Software Error
(gen. CPU)
072/008
072/009
073/001
073/002
073/003
073/004
073/005
073/006
073/007
073/008
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Description
Generator: timeout on signal
Alim_Puis_OK (Inv_Supply_OK) (going
'low' or 'high')
Generator: error reading 'kV law table'
in 0 pt technique (AOP)
Exposure software error: ACQ_NTR or
ACQ_OVR received at start of first trig-
ger
Exposure software error: ACQ_NTR or
ACQ_OVR received at end of first trig-
ger
Exposure software error: unknown ori-
gin of aborted exposure
Exposure software error: aborted by PM
in 2 pt technique
Exposure software error: unknown
exposure technique
One system task (at least) not stable
after a delay of 7 s when a Prep is per-
formed. A Senographe component
(tube, rotor, etc.) or IDC is not ready for
the exposure (prepStatusOk message
not sent to Senographe by IDC).
Exposure software error: unknown task
status
Exposure software error: 0 pt or 1 pt
exposure and Gantry absent
Gene software error (IST, standby):
parameter display problem
Gene software error (IST, standby):
unknown tube type
Gene software error (IST, standby):
alpha=0.0 when calculating value
P_NDac_HTPM
Gene software error (IST, standby):
incorrect thickness for correlation calcu-
lation
Gene software error (IST, standby):
bias calibration task problem
Gene software error (IST, standby):
heater scale calibration task problem
Gene software error (IST, standby): HV/
DAC calibration task problem
Gene software error (IST, standby):
FREQ/HV calibration task problem
Page no. 840
Senographe DS
Recommended actions
Refer to
Job Card TSG A020 - Gener-
ator Power-Up Sequence Flow Charts
on page 1069
This error is either caused by IDC or
Senographe. When a Senographe
sub-system is not ready, there should
be a second error message displayed
relating to the sub-system causing the
error
ERR-SUB-A-001.fm