Job Card D/R A245 - Manual Compression Belt - GE Senographe DS Manual

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1

Job Card D/R A245 - Manual Compression Belt

1
SUPPLIES
Manual Compression Belt
2
TOOLS
Standard Tool Box
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 30 minutes
4
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5
PREREQUISITES
Lower the arm to obtain good access to
the Manual Compression Belt (1).
Switch off the Gantry power.
Remove the compression covers (Refer to
Job Card PHY A044 - Remove/Reinstall
Gantry Covers on page
JC-DR-A-245.fm
Service Information and Procedures Class A 2385072-16-8EN
523).