Honeywell Vertex Edge User Manual page 122

72-point continuous monitor
Hide thumbs Also See for Vertex Edge:
Table of Contents

Advertisement

Clean the Optics
Clean Chemcassette optics annually or whenever optics verification error occurs.
Compressed air is required or per the locations PM schedule.
1. Make sure the Analyzer is out of Monitor Mode.
2. Open the Optics Block Gate.
3. Remove the Chemcassette by releasing and pulling out the analyzer.
4. Open the Vertex Edge side panel.
5. Remove tubing (shown in photo) one at the time and blow out with compressed air.
6. Re-secure side panel and reload the Chemcassette.
NOTE
Remove and clean one port at a time to insure proper orientation of tubing. Do not remove
the capillary tubing (microtubes). After cleaning the Optics, the cleaning counter should be
reset to avoid unnecessary maintenance warning due to Optics cleaning due. (See "Clean
the Optics" above for more information.)
Failure to replace and retighten hardware after service can adversely affect instrument
performance and electromagnetic radiation compliance (EMC). Make certain all fasteners are
reinstalled and firmly fastened.
Vertex Edge System
CAUTION
121
User Manual