GE H
EALTHCARE
D
5394227, 12
IRECTION
4-3-6
Operator panel position adjustment
To avoid injury or damage, make sure nothing is within the range of motion before moving the
CAUTION
operator panel. This includes both objects and people.
4-3-6-1
To raise/lower the operator panel
1.) Hold the elevation lock lever under the operator panel.
2.) Up or Down the operator panel (while holding lever).
3.) Verify the operator panel is free to up/down.
4.) Release the elevation lock lever (disengage).
5.) Verify the operator panel does not up/down.
NOTICE
Do NOT attempt to apply excessive rotating force to the operator panel. The elevation brake is
designed to hold the operator panel in desired position during normal use, and not meant to completely
!! NOTICE:
fix the operator panel position.
PROPRIETARY TO GE
Figure 4-5 Elevation Lock Lever (R1 - R3 System)
Figure 4-6 Elevation Lock Lever (R1 - R3 System)