Table of Contents

Advertisement

Technical
Publication
Direction 5394152
Revision 5
GE Healthcare
LOGIQ™ P6/P6 Pro Service Manual
Operating Documentation
©
Copyright
2011 by General Electric Co.
GE Healthcare

Advertisement

Table of Contents
loading

Summary of Contents for GE LOGIQ P6 Series

  • Page 1 GE Healthcare Technical Publication Direction 5394152 Revision 5 GE Healthcare LOGIQ™ P6/P6 Pro Service Manual Operating Documentation © Copyright 2011 by General Electric Co.
  • Page 2 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Page 1-2...
  • Page 3: Table Of Contents

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
  • Page 4 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL CHAPTER 2 Pre-Installation Overview ............2 - 1 Purpose of chapter 2 .
  • Page 5 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL CHAPTER 3 Overview............3 - 1 Purpose of Chapter 3 .
  • Page 6 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL Paperwork ............3 - 25 Product Locator Installation .
  • Page 7 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL CHAPTER 4 Functional Checks Overview............4 - 1 Purpose of Chapter 4 .
  • Page 8 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL Board Function Checks ..........4 - 37 PI128II Assy function check procedure .
  • Page 9 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview............5 - 1 Block Diagrams and Theory .
  • Page 10 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL Remote Software/Option Installation and Updates ....5 - 34 RFS (Service For Request) ......... . . 5 - 35 General .
  • Page 11 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL CHAPTER 6 Service Adjustments Overview............6 - 1 Purpose of this chapter 6 .
  • Page 12 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview ............7 - 1 Purpose of Chapter 7 .
  • Page 13 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL Noise in Image ..........7 - 31 Unable to Scan .
  • Page 14 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview ............8 - 1 Purpose of Chapter 8 .
  • Page 15 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL Probe Holder ..........8 - 83 Cup Holder .
  • Page 16 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL Printer VCR DVD Fixture Top installation - UP23MD Printer ... . 8 - 169 Color Printer Fixture Top Installation - UP-D25MD Printer ... . . 8 - 171 Printer VCR DVD fixture Top installation - CP900 Printer .
  • Page 17 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL CHAPTER 9 Renewal Parts Overview............9 - 1 Purpose of Chapter 9 .
  • Page 18 Standard GE Tool Kit ........
  • Page 19 • VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES WARNUNG KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN (DE) WURDE. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN. Page ii-i...
  • Page 20 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION • ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
  • Page 21 ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ∆ΥΝΟΥΣ. • EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A FIGYELMEZTETÉS KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
  • Page 22 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION • TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE. • V PŘÍPADĚ, ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTŘEBUJE NÁVOD V JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA ÚKOLEM ZÁKAZNÍKA.
  • Page 23 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION • ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA. • JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ, VERTIMU PASIRŪPINTI TURI KLIENTAS. • NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT ĮSPĖJIMAS...
  • Page 24 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION • ДАННОЕ РУКОВОДСТВО ПО ОБСЛУЖИВАНИЮ ПРЕДОСТАВЛЯЕТСЯ ТОЛЬКО НА АНГЛИЙСКОМ ЯЗЫКЕ. • ЕСЛИ СЕРВИСНОМУ ПЕРСОНАЛУ КЛИЕНТА НЕОБХОДИМО РУКОВОДСТВО НЕ НА АНГЛИЙСКОМ ЯЗЫКЕ, КЛИЕНТУ СЛЕДУЕТ САМОСТОЯТЕЛЬНО ОБЕСПЕЧИТЬ ПЕРЕВОД. • ПЕРЕД ОБСЛУЖИВАНИЕМ ОБОРУДОВАНИЯ ОБЯЗАТЕЛЬНО ОБРАТИТЕСЬ...
  • Page 25 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION (JA) (ZH-CN) (KO) Page ii-vii...
  • Page 26 GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 27 The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
  • Page 28 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Revision History Revision Date Reason for change 24 Aug 2010 Initial Release. 2 Nov 2010 PHI data 2 Mar 2011 Software version update 25 May 2011 Add P95 Series...
  • Page 29: Overview

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the LOGIQ™ P6/P6 Pro scanner. The service provider must read and understand all the information presented in this manual before installing or servicing a unit.
  • Page 30: Typical Users Of The Basic Service Manual

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 1-1-3 Typical Users of the Basic Service Manual • Service Personnel (installation, maintenance, etc.) • Hospital’s Service Personnel • Contractors (Some parts of Chapter 2 - Pre-Installation) 1-1-4 LOGIQ™...
  • Page 31: Important Conventions

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
  • Page 32: Standard Hazard Icons

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
  • Page 33: Product Icons

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-5 Product Icons LABEL/SYMBOL PURPOSE/MEANING...
  • Page 34 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL Table 1-5 Product Icons LABEL/SYMBOL PURPOSE/MEANING LOCATION “OFF/Standby” Indicates the power off/ standby position of the power switch. Adjacent to On-Off/Standby Switch CAUTION This Power Switch DOES NOT ISOLATE Mains Supply “Mains ON”...
  • Page 35: Weee Label

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 1-2-4 WEEE Label The following table describes the meaning of WEEE label and its translation provided on the equipment. Table 1-6 WEEE Label LABEL/SYMBOL LANGUAGE PURPOSE/MEANING This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately.
  • Page 36 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Table 1-6 WEEE Label LABEL/SYMBOL LANGUAGE PURPOSE/MEANING Tämä kuvatunnus ilmaisee, että sillä merkittyä sähkö- ja elektroniikkalaitetta ei saa hävittää lajittelemattomana yhdyskuntajätteenä Finnish (FIN) vaan se on kerättävä talteen erikseen. Ota yhteyttä tuotteen valmistajan valtuuttamaan myyntiedustajaan ja kysy lisätietoja laitteen käytöstä...
  • Page 37 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Table 1-6 WEEE Label LABEL/SYMBOL LANGUAGE PURPOSE/MEANING Questo simbolo indica che i rifiuti derivanti da apparecchiature elettriche ed elettroniche non devono essere smaltiti come rifiuti municipali indifferenziati e Italian (ITA) devono invece essere raccolti separatamente.
  • Page 38 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Table 1-6 WEEE Label LABEL/SYMBOL LANGUAGE PURPOSE/MEANING Acest simbol indică faptul că deşeurile de echipamente electrice şi electronice nu au voie să fie aruncate nediferenţiat ca gunoi menajer şi c...
  • Page 39 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Table 1-6 WEEE Label LABEL/SYMBOL LANGUAGE PURPOSE/MEANING Bu sembol, elektrikli ve elektronik ekipmanların sınıflandırılmamış çöp olarak atılmaması ve ayrı olarak toplanması gerektiğini belirtir. Lütfen Turkish (TUR) ekipmanınızın imhasıyla ilgili olarak üreticinin yetkili temsilcisine baş...
  • Page 40: Safety Considerations

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
  • Page 41: Electrical Safety

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 1-3-3 Mechanical Safety (cont’d) NOTE: Special care should be taken when transporting the unit in a vehicle: • Secure the unit in an upright position. • Lock the wheels (brake) •...
  • Page 42: Label Locations

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 1-3-5 Label Locations Figure 1-1 OUTSIDE MARKINGS OF LOGIQ™ P6/P6 Pro (Rear) 1.) Identification and Rating Plate - USA/Asia 120V Console 2.) Identification and Rating Plate - Europe/Asia/Latin America 220V Console 3.) Identification and Rating Plate - Japan 100V Console...
  • Page 43 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 1-3-5 Label Locations (cont’d) Figure 1-2 OUTSIDE MARKINGS OF LOGIQ™ P6/P6 Pro 1.) Possible Shock Hazard 2.) Label -”Do not use the following devices near this equipment.” 3.) Label - “The equipment weighs approximately 80kg.
  • Page 44: Dangerous Procedure Warnings

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 1-3-6 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT DANGER IN THIS EQUIPMENT.
  • Page 45: Emc, Emi, And Esd

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 1-4 EMC, EMI, and ESD 1-4-1 Electromagnetic Compatibility (EMC) Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface.
  • Page 46: Customer Assistance

    Milwaukee, WI 53219 1-262-524-5698 Customer Answer Center Fax: +1-414-647-4125 Europe Tel: +49 212 2802 208 GE Ultraschall Deutschland GmbH& Co. KG +49 212 2802 207 BeethovenstraBe 239 Postfach 11 05 60, D-42665 Solingen Germany Fax: +49 212 2802 431 Tel: +65 291-8528...
  • Page 47: System Manufacturer

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 1-5-2 System Manufacturer Table 1-8 System Manufacturer Manufacturer FAX Number GE Ultrasound Korea, Ltd. 65-1, Sangdaewon-dong, Jungwon-gu, Seongnam-si, 82-31-740-6436 Gyeonggi-do, 462-120, Korea Chapter 1 - Page 1-19...
  • Page 48 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION This page was intentionally left blank. Page 1-20 Section 1-3 - Safety Considerations...
  • Page 49: Pre-Installation

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Chapter 2 Pre-Installation Section 2-1 Overview 2-1-1 Purpose of chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ™ P6/ P6 Pro.
  • Page 50: General Console Requirements

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 2-2 General Console Requirements 2-2-1 Console Environmental Requirements Table 2-2 Environmental Requirements for LOGIQ™ P6/P6 Pro Scanners Operational Storage Transport 10 - 30 -10 - 60 -40 - 60...
  • Page 51: Electrical Requirements

    Electrical Requirements NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 52 2-2-2-5 Unit Power Plug If the unit arrives without the power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-6 Power Stability Requirements Voltage drop-out Max 10 ms.
  • Page 53: Emi Limitations

    Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit gaskets touch metal RF leakage. Or, if a label has been found in such a position, move the label. Use GE specified harnesses The interconnect cables are grounded and require ferrite beads and other shielding. Also, and peripherals cable length, material, and routing are all important;...
  • Page 54: Scan Probe Environmental Requirements

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 2-2-4 Scan Probe Environmental Requirements Operation:10° to 30° C Storage:-10° to 60° C NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * 9/5) + 32).
  • Page 55: Facility Needs

    All electrical work on these product must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
  • Page 56: Required Features

    2-3-2 Required Features NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 57: Desirable Features

    Emergency oxygen supply • Storage for linens and equipment • Nearby waiting room, lavatory, and dressing room • Dual level lighting (bright and dim) • Lockable cabinet ordered by GE for its software and proprietary manuals. Chapter 2 - Page 2-9...
  • Page 58: Recommended And Alternate Ultrasound Room Layout

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 2-3-4 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a minimal floor plan for ultrasound equipment: Figure 2-1 RECOMMENDED ULTRASOUND ROOM LAYOUT Page 2-10 Section 2-3 - Facility Needs...
  • Page 59: Networking Pre-Installation Requirements

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 2-3-5 Networking Pre-installation Requirements 2-3-5-1 Purpose of DICOM Network Function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network. Examples of DICOM services include the transfer of images to workstations for viewing or transferring images to remote printers.
  • Page 60 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 2-3-5-2 DICOM Option Pre-installation Requirements (cont’d) LOGIQ™ IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default ROUTER1 ROUTER2...
  • Page 61: Overview

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Chapter 3 Installation Section 3-1Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are references to a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
  • Page 62: Average Installation Time

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 3-2Installation Reminders 3-2-1 Average Installation Time Table 3-2 Average Installation Time Description Average Installation Time Comments 0.5 hour Unpacking the scanner 0.5 hour Scanner wo/options Dependant on the configuration that is required 0.5 hour...
  • Page 63: Safety Reminders

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 3-2-3 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE DANGER AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH THE UNIT! Two people should unpack the unit because of its weight.
  • Page 64: Receiving And Unpacking The Equipment

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 3-3 Receiving and Unpacking the Equipment When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1.
  • Page 65 GE H EALTHCARE 5394152, Revision LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont’d) 2.) Lift the TOP COVER up and off. Top Cover Figure 3-3 Removing the top Cover Chapter 3 - Page 3-5...
  • Page 66 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 3-3 Receiving and Unpacking the Equipment (cont’d) 3.) Remove the Top Plate up and off. Top Plate Figure 3-4 Removing the Top Plate Page 3-6 Section 3-1 - Overview...
  • Page 67 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 3-3 Receiving and Unpacking the Equipment (cont’d) 4.) Remove the three PLASTIC JOINTs from the OUTER SLEEVE. 5.) Remove the OUTER SLEEVE. Plastic Joint Outer Sleeve Palette Assy...
  • Page 68 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 3-3 Receiving and Unpacking the Equipment (cont’d) 6.) Remove the MONITOR SUPPORTER and Option Box. 7.) Remove the PLASTIC BAG. 8.) Unlock the front caster and carefully put the console off the PALETTE.
  • Page 69: Moving Into Position

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 3-3-1 Moving into Position Do not lift the unit by the Keyboard. CAUTION Do not tilt the unit more than 5 degrees to avoid tipping it over. To avoid injury by tipping over. Set the monitor to the lowest position before moving.
  • Page 70: Preparing For Installation

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 3-4 Preparing for Installation 3-4-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered or damaged.
  • Page 71: Completing The Installation

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 3-5 Completing the Installation 3-5-1 Power On/Boot Up NOTE: After turning off the system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 72 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 3-5-1-1 Scanner Power On (cont’d) 4.) Press the ON/OFF key at the front of the System once. Figure 3-9 Power On/Off Switch Location 3-5-1-2 Power Up Sequence 1.) The Start Up Screen will be shown on the Monitor display when the system is turned ON.
  • Page 73: Power Off/ Shutdown

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 3-5-1-3 Power Up Sequence NOTE: To enter the Maintenance Mode, select Maintenance button. NOTE: Figure 3-11 on page 3-13 displays when Service Dongle inserted in Rear Panel. Figure 3-11 Start Application Window NOTE: Put in the service password in the Password box.
  • Page 74: Transducer Connection

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 3-5-3 Transducer Connection 1.) Connect a transducer to one of the three rightmost transducer receptacle as follows: A.) Ensure that the transducer twist lock lever to the horizontal position.
  • Page 75: System Configuration

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 3-6System Configuration 3-6-1 System Specifications 3-6-1-1 System Settings Table 3-1 System Configuration Configuration Category Description Enables the user or service personnel to set the date, time, unit, language, basic information about...
  • Page 76 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 3-6-1-2 Physical Dimensions The physical dimensions of the LOGIQ™ P6/P6 Pro unit are summarized in Table 3-2 on page 3-16 . The Size of LOGIQ™ P6/P6 Pro, with monitor and peripherals Table 3-2 Physical Dimensions of LOGIQ™...
  • Page 77: Electrical Specifications

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 3-6-1-3 Weight without Monitor and Peripherals Table 3-3 Weight of LOGIQ™ P6/P6 Pro With Monitor and Without Other Peripherals Model Weight [kg] Weight [lbs] LOGIQ™ P6/P6 Approximately 80 Approximately 176...
  • Page 78: External I/O Connector Panel

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 3-6-4 External I/O Connector Panel Located on the rear panel are video input and output connectors, audio input and output, USB, footswitch connector power connector and control connections for VCR or DVD Recorder, printer, and service tools.
  • Page 79 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 3-6-4-1 External I/O Pin Outs Pin No. Signal Pin No. Signal GREEN SGND BLUE HSYNC RGND VSYNC GGND BGND Table 3-6 Pin Assignments of DB15 connector for External VGA Pin No.
  • Page 80: Video Specification

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 3-6-5 Video Specification Video specifications may be needed to be able to connect the VCR or DVD Video Recorder to the LOGIQ™ P6/P6 Pro. Table 3-9 Video Specifications...
  • Page 81: Setting Vcr (Mitsubishi Md3000) Usb Interface

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 3-6-6 Setting VCR (Mitsubishi MD3000) USB Interface 1.) Turn off the Power of the VCR. 2.) Open the rear panel of the VCR. 3.) Set dip switches. Off 1~6 pins, On 7 pin, and Off 8 pin.
  • Page 82: Available Probes

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 3-7 Available Probes See Specifications in the LOGIQ™ P6/P6 Pro User Manual for probes and intended use. See Chapter 9, for part numbers to be used when ordering new or replacement probes.
  • Page 83: Connectivity Installation Worksheet

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 3-9 Connectivity Installation Worksheet Site System Information Comments: Floor: Site: Dept: Room: Type: LOGIQ SN: REV: CONTACT INFORMATION Name Title Phone E-Mail Address TCP/IP Settings Name - AE Title:...
  • Page 84: Insite Ip Address Configuration

    GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 3-10 Insite IP Address Configuration This information needed to facilitate the OnLine Center engineers in connecting to the system is found in the InSite Installation Manual. Section 3-11...
  • Page 85 3-13-2 User Manual(s) Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. 3-13-2-1 Reference off-board peripherals and options None.
  • Page 86 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION This page was intentionally left blank. Page 3-26 Section 3-12 - Warnings...
  • Page 87 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ P6/P6 Pro scanner diagnostics by using the built-in service software, and power supply adjustments.
  • Page 88 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
  • Page 89 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL 4-3-1-1 Scanner Power On (cont’d) 3.) Press the ON/OFF key at the front of the System once. Figure 4-2 Power On/Off Standby Switch Location Chapter 4 - Page 4-3...
  • Page 90 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-2 Power Off / Shutdown NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 91 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-3 System Features 4-3-3-1 Control Panel for LOGIQ™ P6/P6 Pro Figure 4-4 Control Panel Tour LOGIQ™ P6/P6 Pro 1.) Power ON/OFF switch 2.) Patient Key 3.) Alpha Numeric key 4.) TGC...
  • Page 92 14. Focal Zone Indicator. 2. Patient Name, Patient Identification. 15. TGC. 3. Power Output Readout 16. Body Pattern. 4. GE Symbol: Probe Orientation Marker. 17. Depth Scale. 5. Image Preview. 18. Top Menu 6. Gray/Color Bar. 19. Caps Lock: Lit when On.
  • Page 93 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-4 System B/M-Mode Checks For a basic functional check of the system’s different modes, the Quick Guide will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler.
  • Page 94 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-5 System CFM and PWD Checks Table 4-3 CFM and PWD Functions Adjusts the baseline to accommodate faster or slower blood Baseline flows to eliminate aliasing. Velocity scale determines pulse repetition frequency. If the...
  • Page 95 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-6 Basic Measurements NOTE: The following instructions assume that you first scan the patient and then press Freeze. 4-3-6-1 Distance and Tissue Depth Measurements 1.) Press MEASURE once an active caliper displays.
  • Page 96 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL 4-3-6-3 Worksheets Measurement/Calculation worksheets are available to display and edit measurements and calculations. There are generic worksheets as well as Application specific worksheets. 4-3-7 Probe/Connectors Usage 4-3-7-1 Connecting a probe 1.) Place the probe's carrying case on a stable surface and open the case.
  • Page 97 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-8 Using Cine 4-3-8-1 Activating CINE Press FREEZE, then roll the TRACKBALL to activate CINE. To start CINE Loop playback, press Run/ Stop. To stop CINE Loop playback. press Run/Stop.
  • Page 98 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-10 Using the DVD-R 4-3-10-1 Using the DVD-R Never move the unit with a disk in the DVD-R because the drive actuator will not be locked NOTICE and the DVD-R could break.
  • Page 99 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-10-1 Using the DVD-R (cont’d) 8.) Select the type of Removable CD/DVD on the [To] combo box. 9.) It will do auto-formatting. Figure 4-7 Formatting 10.)After auto-formatting, select some patients on the upper list area and then press the [Transfer] button.
  • Page 100 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-10-1 Using the DVD-R (cont’d) 12.)The selected data is transferred correctly. Figure 4-9 Transfer 13.)Press F3 and select the [Yes and Verify files]. 14.)It will do finalizing. Figure 4-10 Finalizing...
  • Page 101 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-10-1 Using the DVD-R (cont’d) 15.)It will do Verifying. Figure 4-11 Verifying 16.)It will eject the CD/DVD and Go to Utility and reinsert the CD/DVD. 17.)Go to Utility->Connectivity->Removable Media.
  • Page 102 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-11 Backup and Restore Database, Preset Configurations and Images NOTE: When upgrading from R1.0.x to R2.0.x or R3.0.x , preset cannot be transferred. Refer to the table below.
  • Page 103 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Media Verify Format Selection Format Media Label Figure 4-13 Format and Verify Media 5.) The Ultrasound system displays a pop-up menu, as shown in Figure 4-14. When the formatting has been completed, press OK to continue.
  • Page 104 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-11-2 Backup System Presets and Configurations NOTE: Always backup any preset configurations before a software reload. This ensures that if the presets need to be reloaded, after the software update, they will be the same ones the customer was using prior to service.
  • Page 105 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-11-3 Restore System Presets and Configurations The restore procedure overwrites the existing database on the local hard drive. Make sure to insert CAUTION the correct CD-R or DVD-R.
  • Page 106 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-11-4 Archiving Images To export an exam(s) to a compatible Ultrasound system: 1.) Format the removable media (CD-ROM/DVD-R). Label the removable media. Answer Yes/OK to the messages.Press Patient. Deselect any selected patient(s) in the search portion of the Patient screen.
  • Page 107 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-11-4 Archiving Images (cont’d) 4.) In the patient list at the top of the Patient menu, select the patient(s) you want to export. You can use Windows commands to select more than one patient. To select a consecutive list of patients, click the cursor on the first name, move the cursor to the last name, then press and hold down the Shift+right Set key to select all the names.
  • Page 108 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-3-12 ECG Check Out Connect the ECG Harness and check: Table 4-5 ECG Harness Check Step Task to do Expected Result(s) Connect the ECG at the Connector on the Front of the It will display a curve along the bottom edge of the image scanner.
  • Page 109 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 4-5Peripheral Checks Check that peripherals work as described below: Table 4-7 Peripheral Checks Step Task to do Expected Result(s) Press (FREEZE) To stop image acquisition. The image displayed on the screen is printed on B&W or Color printer,...
  • Page 110 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 4-6Monitor Function Checks 4-6-1 17" LCDMON II Assy & LCD Panel with filter Function Validation Procedure 1.) Turn on the system. 2.) Select “Utility Key” on the Key board and then choose “Test Pattern” and “Resolution” marked below in the figure below.
  • Page 111 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-6-2 17" OSD Key Assy Function Validation Procedure 1.) Turn on the system. 2.) Select “OSD Button “marked below in the figure below. OSD button for menu display Figure 4-23 3.) OSD Menu should be displayed within 5sec.
  • Page 112 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 4-7Keyboard Function Checks 4-7-1 Main Keyboard Assy Function Validation Procedure 1.) Boot up the system. 2.) Insert the Service Dongle and go to the maintenance mode. 3.) Click 'Start' and go to 'Run'.
  • Page 113 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-7-1 Main Keyboard Assy Function Validation Procedure (cont’d) 6.) Write the 'scfpapitest.exe' on the command prompt and press 'Enter' button to pop up the scfpapitest program. Figure 4-28 scfpapitest program 7.) Press the keyboard button and rotate rotary button.
  • Page 114 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Table 4-8 Button and Rotaty Table Button or Rotary Name The Name displayed in Log Button or Rotary Name The Name displayed in Log BodyPattern/Ellipse PushEncoder Right Button...
  • Page 115 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-7-2 Sub Keyboard Assy Function Validation Procedure 1.) Boot up the system and run the scfpapitest program. Follow up the procedure 4-7-1 "Main Keyboard Assy Function Validation Procedure" on page 4-26 NOTICE run the scfpapitest program.
  • Page 116 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-7-6 TGC key assy / TGC Knob Set validation 1.) Follow up the procedure 4-7-1 "Main Keyboard Assy Function Validation Procedure" on page 4-26 to run the scfpapitest program.
  • Page 117 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 4-8Mechanical Function Checks 4-8-1 Cover Parts Function Validation Table 4-10 Cover parts of the FRU Item Right and Left cover Front cover Monitor Cover set Rear cover...
  • Page 118 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-8-2 Air filter set Function Validation Procedure Check if the air filter is placed deep inside. Figure 4-30 Air Filter Set Page 4-32 Section 4-8 - Mechanical Function Checks...
  • Page 119 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-8-3 TGC key assy/TGC Knob set, Sub keyboard encoder Knob Set, keycap set, Main keyboard Encoder Knob set Function Validation Procedure • Check if the each button operates properly.
  • Page 120 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-8-4 Cup holder Function Validation Procedure Check if the cup holder can not be removed easily. Figure 4-32 Cup Holder Page 4-34 Section 4-8 - Mechanical Function Checks...
  • Page 121 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-8-5 Front caster/Rear caster Function Validation Procedure • In case of Bi break caster, check all caster locks and caster swivel locks for proper operation. • In case of front caster, check caster locks for proper operation.
  • Page 122 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-8-6 New articulation arm with cover Function Validation Procedure • Check if the lock operates properly. • Check if the swivel motion operates properly. • Check if the up and down motion operates properly.
  • Page 123 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 4-9Board Function Checks 4-9-1 PI128II Assy function check procedure WARNING WARNING When Service Login for FRU Diagnostic, scan mode must be B-mode. After Diagnostic is completed, the system should be rebooted...
  • Page 124 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-1 PI128II Assy function check procedure (cont’d) 10.)Activate probe. Refer to the 4-3-7-2 "Activating the probe" on page 4-10. 11.)Check the system basic functions. Refer to the 4-3-4 "System B/M-Mode Checks" on page 4-7 refer to the 4-3-5 "System CFM and PWD Checks"...
  • Page 125 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-2 DTRX64II Assy Function Validation Procedure 1.) Turn on the system and check the completion of Echoloader loading. 2.) Do the Service Login. Refer to the 5-7-3-2 "Service Login" on page 5-25.
  • Page 126 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-3 Probe Assessment Diagnostic Test Validation Procedure NOTE: This diagnostic test does not represent the result as Passed or Failed. 1.) Clean probe surface that you want to assess and connect the probe to the 1st probe slot of the system.
  • Page 127 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-3 Probe Assessment Diagnostic Test Validation Procedure (cont’d) 7.) If the test is executed without probe connection, the test shows the following result. Figure 4-38 Probe Assessment Diagnostic Test 8.) After finishing the test, shut down the system.
  • Page 128 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-4 CWD64 Assy Function Validation Procedure 1.) Insert the Sector probe in probe port. 2.) Turn on the system. 3.) Press Virtual Convex button to enable Virtual Convex if not already enabled.
  • Page 129 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-4 CWD64 Assy Function Validation Procedure (cont’d) 4.) Press “CW” button to start CWD Mode. Figure 4-41 CW mode 5.) Scan in CWD mode (with Virtual Convex enabled).
  • Page 130 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-4 CWD64 Assy Function Validation Procedure (cont’d) 14.)After All diagnostic function check of CWD64 should be Passed.Refer to the Table 7-5, “FRU Test, CWD64 Diagnostics Menu,” on page 22.
  • Page 131 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-5 SYSCONML Assy Function Validation Procedure 4-9-5-1 L2SYSCON diagnostic Test 1.) Turn on the system and check the completion of echoloader loading. 2.) Do the Service Login. Refer to the 5-7-3-2 "Service Login"...
  • Page 132 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-6 APS II Assy PCB Function Validation Procedure 1.) Turn on the system. 2.) Do the Service Login. Refer to the 5-7-3-2 "Service Login" on page 5-25. 3.) Select 'Diagnostics' => 'LOGIQ P6 Diagnostics' => 'FRU test' => 'APS II' => 'APS II Overall'.
  • Page 133 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-7 SOM Function Validation Procedure 1.) Turn on the system. 2.) Press the 'del' key in A/N key within 3 sec. to enter BIOS setting mode. 3.) Set the system time and the system Date.
  • Page 134 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-9 BP128 Assy Function Validation Procedure 1.) Turn on the system 2.) All of attached ASSY should be tested. 3.) Refer to the 4-9-5 "SYSCONML Assy Function Validation Procedure" on page 4-45.
  • Page 135 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-10 Nest fan Function Validation Procedure 1.) After replace Nest fan assy, turn on the system. Refer to the 3-5-1 "Power On/Boot Up" on page 3- 2.) Confirm the fan rotating with naked eye and check the fan operating by rotating sound. No strange sound should be heard.
  • Page 136 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4-9-12 DTRX Cooler with heatsink Function Validation Procedure 1.) Turn on the system. 2.) Check the basic function of system's different modes. Refer to the 4-3-4 "System B/M-Mode Checks"...
  • Page 137 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 4-10Site Log Table 4-11 Site Log Date Service person Problem Comments Chapter 4 - Page 4-51...
  • Page 138 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION This page was intentionally left blank. Page 4-52 Section 4-9 - Board Function Checks...
  • Page 139 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Chapter 5 Components and Functions (Theory) Section 5-1 Overview This chapter explains LOGIQ™ P6/P6 Pro’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution System (PDS) and probes.
  • Page 140 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 5-2 Block Diagrams and Theory 5-2-1 LOGIQ™ P6/P6 Pro Block Diagram Figure 5-1 LOGIQ™ P6 System Block Diagram, 128CH Page 5-2 Section 5-1 - Overview...
  • Page 141 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-2-1 LOGIQ™ P6/P6 Pro Block Diagram (cont’d) Figure 5-2 LOGIQ™ P6 Front End Block Diagram Chapter 5 - Page 5-3...
  • Page 142 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-2-1 LOGIQ™ P6/P6 Pro Block Diagram (cont’d) Figure 5-3 LOGIQ™ P6 Backend Block Diagram Page 5-4 Section 5-1 - Overview...
  • Page 143 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-2-1 LOGIQ™ P6/P6 Pro Block Diagram (cont’d) Figure 5-4 Others Block Diagram (cont’d) Chapter 5 - Page 5-5...
  • Page 144 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-2-1 LOGIQ™ P6/P6 Pro Block Diagram (cont’d) Figure 5-5 LOGIQ™ P6Pro System Block Diagram, 64CH Page 5-6 Section 5-1 - Overview...
  • Page 145 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-2-1 LOGIQ™ P6/P6 Pro Block Diagram (cont’d) Figure 5-6 P6PRO Front End Block Diagram Chapter 5 - Page 5-7...
  • Page 146 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-2-1 LOGIQ™ P6/P6 Pro Block Diagram (cont’d) Figure 5-7 P6PRO Backend Block Diagram Page 5-8 Section 5-1 - Overview...
  • Page 147 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 5-3 Power Diagrams 5-3-1 Overview The main function of the AC Power assy is to supply powers the various internal subsystems, and to isolate the scanner electrically from the on site Mains Power System. The voltage to peripherals can be confederated to either 110 VAC or 220 VAC.
  • Page 148 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-3-3 APS II (LV Power & HV Power) APS II supply all DC power sources which are for all PWA assy in nest box, nest FANs and trans FAN.
  • Page 149 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-3-4 AC Power Outlet for Peripheral The LP6 ARP II 110V Assy and LP6 ARP II 220V Assy supply the AC power to peripherals. Both of the assemblies are the same except the circuit breaker's current ratings depending on voltage of the system power source (See 'Safety' described below).
  • Page 150 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-3-6 SYSCONML Figure 5-9 SYSCONML Block Diagram This block has 4 main functions: front-end control, mid-processing for MLA2, post processing for display and user interface control. • Front-end Control: This is to control the front beamfomer assys. This includes as: Probe interface, assy version reading interface, scan control for transmit/receiving beamforming.
  • Page 151 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 5-4 Top Console 5-4-1 Keyboard M a in K e yb o a rd M a in K e yb o a rd R A M R A M...
  • Page 152 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-4-1-1 KEYBOARD Signal I/O Main keyboard assy have two main external I/O connectors. One is for USB connection to scan system for main control signal interface, the other is for 5V power supply and momentary switch of scan system with hard drive operation indicator.
  • Page 153 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-4-2 LOGIQ™ P6/P6 Pro Monitor LOGIQ™ P6/P6 Pro system has 17inch LCD monitor with DVI-I interface. Figure 5-12 LOGIQ™ P6/P6 Pro Monitor 5-4-2-1 MONITOR Signal I/O Monitor assy has two main I/O connectors, One is a DVI-I connector, and the other is a 110V/220V power plug.
  • Page 154 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-4-2-2 LOGIQ™ P6/P6 Pro LCDMON control button function LOGIQ™ P6/P6 Pro LCDMON assy has three control buttons: Left button, Menu button, Right button. Figure 5-14 LCDMON control buttons...
  • Page 155 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 5-5 P6 ARP II (Rear Panel) The LOGIQ™ P6/P6 Pro system has the I/O interfaces for peripherals in rear panel. P6 ARP II assy has connectors for peripheral. Composite in/out and S-VHS in/out are analog video interfaces for recording and playback of the VCR or DVD recorder.
  • Page 156 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-5-1 Input/Output Signals for peripheral Table 5-4 Input and Output signals for peripherals Description Signal Name Connection from/via/to SYSCONML -> P6 ARP II -> 7P LAN Standard TCP/IP Ethernet...
  • Page 157 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 5-6 Cable Connection 5-6-1 Overview The cable connection can be separated to two group, internal harness and external cable connection for peripheral equipment. 5-6-2 Internal Cable K B D A S S Y...
  • Page 158 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-6-3 Peripherals connection A DVD recorder, a Black & White Video Printer and a Color Printer may be installed in system as onboard. These devices are connected to rear (P6 ARP II) or front side (bottom of keyboard). LOGIQ™...
  • Page 159 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-6-3-2 DVD Recorder Connect the cables as shown in Figure 5-20.. Audio cable with RCA Jack S-Video Cable S-Video Cable Audio cable with RCA Jack Figure 5-20 DVD Recorder...
  • Page 160 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-6-5 Fans There are 10 fans for ventilation in the system as below a.) DTRX64II Assy: 2 fans in each assy. b.) SYSCONML: 1 fan c.) APS II: 2 fans d.) NESTBOX top: 2 fans...
  • Page 161 The Service Platform contains a set of software modules that are common to all ultrasound and cardiology systems containing a PC backend. This web-enabled technology provides linkage to e- Services, e-Commerce, and the iCenter, making GE’s scanners more e-enabled than ever. The Service Platform will increase service productivity and reduce training and service costs.
  • Page 162 Global Service User Interface (GSUI) 5-7-2-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. 5-7-2-2 Service Login Enable utility menu by pressing the utility key in A/N key.
  • Page 163 GEMS products. 5-7-3-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. At this time there is no multi-lingual capability built into the Common Service Interface. 5-7-3-2 Service Login 1) Touch Utility >...
  • Page 164 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-7-4 Service Home Page When the Service Interface is started, the Service Home Page appears. The Home Page contains the software revision along with the hardware inventory and the results of the latest System Health Information.
  • Page 165 From the Error Logs Tab the Log Viewer displays four categories with pull-down sub-menus and an Exit selection. The Service Interface allows scanner logs to be viewed by all service users. The Filter Error log is keyed to GE Service login access permissions and is not available to customer level analysis.
  • Page 166 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-7-5-1 Logs The seven sub-menus of the Logs category are System, Infomatics, Temperature, Probe, Board, and DICOM. NOTE: Figure 5-26 provides a graphical example of the log entries for the System Logs.
  • Page 167 WARNING WARNING Diagnostics should be performed with probes which have 128 or more elements Detailed Diagnostic information is found in Diagnostics/Troubleshooting. Figure 5-28 GE Service Diagnostic Page NOTE: After finishing Diagnostic test, the system should be rebooted. 5-7-6-1 Diagnostics Execution Diagnostic tests are executable by both local and remote users.
  • Page 168 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-7-7 Image Quality Field is not yet populated. The Image Quality page contains tools for troubleshooting image quality issues. Figure 5-29 Image Quality Page 5-7-8 Calibration Field is not yet populated.
  • Page 169 Figure 5-31 Configuration Page 5-7-9-1 Questra Agent Configuration These feature that allow the customer to register the system to the GE OnLine Center are available for Warranty and Contract customers only. 5-7-9-1-1 Device Information Device name and CRM No. are displayed by default.
  • Page 170 If Proxy is used in the site, it should be filled proper Proxy Server Address and Proxy Server Port. And Press “Submit Changes” button to finish check out process. Figure 5-32 Proxy Configuration 5-7-10 Utilities The Utilities page contains several miscellaneous tools. Figure 5-33 GE Service Utilities Page Page 5-32 Section 5-1 - Overview...
  • Page 171 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-7-11 Replacement The Replacement page is intended to contain the tools used to track replacement parts used in the system. Figure 5-34 Part Replacement Page 5-7-12 The PM page is intended to contain the tools used in periodic maintenance of the system.
  • Page 172 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-7-13 Remote Software/Option Installation and Updates The Service Platform provides for remote software downloads. Any file transferred to the scanner will be in the form of an installation package. The actual content of the installation package is dependent on each product.
  • Page 173 General This describes general RFS functionality. 1.) RFS function is a call for service to be used by customer or FE directly from the system instead of calling GE Cares 5-8-2 System Checkout Procedure If the system is checked out already, skip 5-8-2 - System Checkout Procedure and proceed from 5-8- 3 - Sending RFS Procedure.
  • Page 174 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-8-2 System Checkout Procedure (cont’d) All bold fields should be filled 7.) Device Name and CRM No are pre-populated and can not be changed. 8.) Select Continent and Country.
  • Page 175 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-8-2 System Checkout Procedure (cont’d) 13.)Wait until following screen displays. Figure 5-38 Configuration done 14.)Configuration is complete. 15.)Close the Service Platform. Chapter 5 - Page 5-37...
  • Page 176 Service Platform should be turned off in order to open RFS window. System must be connected to network, configured and checked out to be able to send the RFS. 1.) Right Click on “Connect to GE” icon. Figure 5-39 Contact GE Icon All the fields that are marked with asterisk * are mandatory and should be filled in order to send the RFS request to On Line Center.
  • Page 177 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-8-3 Sending RFS Procedure (cont’d) 7.) Then the “Send” button is enabled. Figure 5-41 RFS send enabled 8.) Verify that “Send” button is enabled. 9.) Click “Send” button and wait for confirmation window with reference number on the screen.
  • Page 178 2.) MRFS function is an automatic call for service from the system to the OLC when system detects high temperature or high or low voltage. 3.) Right Click on “Connect to GE” icon. Figure 5-42 Contact GE Icon 4.) When RFS window pops up, click on “Users” tab.
  • Page 179 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 5-9-2 Enabling MRFS (cont’d) All the fields that are marked with asterisk * are mandatory and should be filled 6.) When fields with asterisk are filled, “Add User” button is enabled.
  • Page 180 This describes Fast Polling feature that will change the polling rate to 15 seconds for 10 minutes. 5-10-2 Enabling Fast Polling 1.) Right Click on “Connect to GE” icon. Figure 5-47 Pressing “OK” will change the polling rate to 15 seconds for 10 minutes.
  • Page 181 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this chapter 6 This section describes how to test and adjust the scanner. These tests are optional. You may use them to check the system for errors.
  • Page 182 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 6-3 Monitor Adjustments 6-3-1 Brightness and Contrast Adjusting the monitor's contrast and brightness is one of the most important factors for proper image quality. If these controls are set incorrectly, the Gain, TGC, Dynamic Range and even Power Output may have to be changed more often than necessary to compensate.
  • Page 183 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 6-3-2 Monitor Arm Locking Flexible LCD arm’s rotation movement can be adjusted by using the LCD arm’s locking mechanism. To unlock the LCD arm: 1.) Pull the LCD arm toggle knob and turn clockwise to unlock the LCD arm.
  • Page 184 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 6-4 Trackball Cleaning The trackball on the keyboard need to be cleaning periodically. 1.) Open the trackball. Figure 6-4 Open Trackball Remove dust inside of bucket Figure 6-5 Trackball Hole...
  • Page 185 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
  • Page 186 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 7-2 Gathering Trouble Data 7-2-1 Overview There may be a time when it would be advantageous to capture trouble images and system data (logs) for acquisition through remote diagnostics (InSite) or to be sent back to the manufacturer for analysis.
  • Page 187 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-2-3 Collect a Trouble Image with Logs If the system should malfunction, press the Alt-D keys simultaneously. This will collect a screen capture of the image monitor, system presets and the following logs:...
  • Page 188 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 7-3 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center through InSite. This is accomplished by first saving the image(s) to the clipboard using a Print Key.
  • Page 189 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-3-1 Check and Record the P2 Key Function (cont’d) 5.) In the “Physical Print Buttons”box, record the followings. Format Image Frames Capture Area Compression 6.) In the “Printflow View”, record the print flow associated with the Print2 Button.
  • Page 190 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-3-3 Capturing a Screen The following is a generic process to capture any screen from the scanner: 1.) Navigate to and display the image/screen to be captured. 2.) Press P2. This will place a snapshot of the screen on the “clipboard” displayed at the bottom of the scan image display.
  • Page 191 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-3-3 Capturing a Screen (cont’d) Figure 7-5 Save Dialog Box 5.) A Save dialog box will be opened. Choose the destination as the archive location to save the image on the CD-R/DVD-R, USB or network storage.
  • Page 192 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 7-4 Diagnostics As described in "Components and Functions (Theory)" on page 5-1, the service platform uses a web- based user interface (UI) to provide access to common service components.
  • Page 193 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 7-5 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center through InSite. This is accomplished by first saving the image(s) to the clipboard using a Print Key.
  • Page 194 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-5-1 Check and Record the P2 Key Function (cont’d) 5.) In the “Physical Print Buttons”box, record the followings. Format Image Frames Capture Area Compression 6.) In the “Printflow View”, record the print flow associated with the Print2 Button.
  • Page 195 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-5-3 Capturing a Screen The following is a generic process to capture any screen from the scanner: 1.) Navigate to and display the image/screen to be captured. 2.) Press P2. This will place a snapshot of the screen on the “clipboard” displayed at the bottom of the scan image display.
  • Page 196 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-5-3 Capturing a Screen (cont’d) Figure 7-10 Save Dialog Box 5.) A Save dialog box will be opened. Choose d:\export folder as the archive location to save the image on the hard disk or CD-R or USB.
  • Page 197 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 7-6 Diagnostics As described in - Components and Functions (Theory), the service platform uses a web-based user interface (UI) to provide access to common service components. The Diagnostic home page displays a listing of test categories consisting of Common Diagnostics, and LOGIQ™...
  • Page 198 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 7-7 Common Diagnostics Common Diagnostics is the first main heading under the GSUI heading of Diagnostics. Figure 7-12 shows the top level menu choices under Common Diagnostics. The subheadings and their respective diagnostic/troubleshooting menu choices are called out and described in the sections below.
  • Page 199 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-7-2 PC Diagnostics (Non-Interactive Tests) 7-7-2-1 CPU Tests • CPU Test (All) - This diagnostic performs a battery of tests on your CPU to ensure it is working properly.
  • Page 200 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-7-2-2 Hard Drive Tests • Hard Drive Test (All) - The Hard Drive test examines your hard disk drive using a series of tests to ensure that your drive can perform its functions.
  • Page 201 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-7-2-4 CD/DVD Drive Test NOTE: Although the global aspect of the diagnostics allow for DVD testing, the LOGIQ™ P6/P6 Pro does not have a DVD drive.(Option) • DVD/CD-ROM Test (All) - The CD-ROM/DVD test examines the CD-ROM/DVD drive to make sure it is working properly.
  • Page 202 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-7-2-6 USB Test (Not yet supported) • USB Test (All) • USB Configuration Test • USB Devices Test 7-7-2-7 PCI Board Configuration Test A bus is a set of circuits, wires or connectors which connect the various components of your system. A PCI (Peripheral Component Interconnect) bus is a fast standard bus which is common in Pentium systems.
  • Page 203 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-7-3 PC Diagnostics (Interactive Tests) 7-7-3-1 Keyboard Test 1.) Click Keyboard Test from the PC Diagnostics interactive tests menu. 2.) Click the Test button. A keyboard layout is displayed on which you can see the keys of your keyboard.
  • Page 204 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-7-4 FRU Test 7-7-4-1 SYSCONML These programs are provided for testing the SYSCONPM/SYSCONCM board. Table 7-2 FRU Test, SYSCONPM/SYSCONCM Diagnostics Menu User Descriptions Menu Level Tools Full Test...
  • Page 205 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-7-4-2 DTRX64II These programs are provided for testing the DTRX64II assy. Table 7-3 FRU Test, DTRX64II Assy Diagnostics Menu User Descriptions Menu Level Tools Full Test Performs all of the DTRX64II board tests listed below None Test EEPROM.
  • Page 206 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-7-4-3 PI128 II (LOGIQ™ P6/P6 Pro) These programs are provided for testing the PI128 II assy. NOTE: A Convex Probe is needed for PI128 II Channel Test. Table 7-4...
  • Page 207 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-7-4-5 P3RLY (LOGIQ™ P6/P6 Pro PRO) These programs are provided for testing the P3RLY board. NOTE: A Convex Probe is needed for RLY Channel Test. Table 7-6 FRU Test, P3RLY Diagnostics Menu...
  • Page 208 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-7-4-7 APS II These programs are provided for testing the APS II board. Table 7-8 FRU Test, APS II Diagnostics Menu User Descriptions Menu Level Tools Full Test...
  • Page 209 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 7-8 LED Descriptions 7-8-1 DTRX64II Assy The DTRX64II Assys are located at the No.1 and No.2 slot in the NEST Assy. Table 7-10 LED on DTRX64II Assy...
  • Page 210 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-8-5 APS II Assy The APS Assy is located at the No. 5 slot of the NEST Assy. Table 7-14 LEDs on APS II Assy Descriptions Normal State...
  • Page 211 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 7-9Trouble Shooting Tree 7-9-1 System Does Not Power On / Boot Up Is the scanner plugged in? Make sure the scanner is not plugged into emergency power (red outlet);...
  • Page 212 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-2 System Does Not Boot System doesn’t boot! Is the AC power cable seated to the Scanner and to the wall power outlet? Connect the AC power cable between the Scanner and the wall power outlet in right way.
  • Page 213 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-3 System Does Not Boot System doesn’t boot! (cont’d) Do the fans start to run? Check; 1.) AC Power 2.) AC Transformer Is the SOM module starting Check/replace the PM SOM/CM to display start-up screens SOM module.
  • Page 214 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-3 System Does Not Boot (cont’d) System doesn’t boot! (cont’d) Are the Control Panel keys lit? 1.) Check the Control Panel cable connection. 2.) Check/replace Control Panel. Is regular scan display...
  • Page 215 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-4 Noise in Image Noise is shown in Image. Is there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices. Check probes.
  • Page 216 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-5 Unable to Scan Unable to scan! Are probes connected in right way? Check the probe connection. Are other operations possible? Check the HV power supply unit. Restart the scanner.
  • Page 217 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-6 Trackball The trackball is low sensitive! Does the trackball move at all? Remove the ball from the Trackball, and clean it. Replace the trackball unit. Figure 7-19 Trackball Troubleshooting...
  • Page 218 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-7 Monitor Troubleshooting NOTE: Reset the monitor settings to the factory defaults prior to troubleshooting the monitor. Comments to remember about the 17” monitor. NOTICE The monitor is the diagnostic monitor and is connected to AC 110/220 VAC.
  • Page 219 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-7 Monitor Troubleshooting (cont’d) Press once the on/off switch. Press and hold on/off switch until Is the shutdown dialog system shut down. box displayed? Shutdown the scanner from a software...
  • Page 220 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-8 Unable to Record to VCR Unable to record to VCR! Is the VCR properly connected? Check that all signal, and power supply cable connections to the VCR unit are OK.
  • Page 221 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-9 Printer Troubleshooting Unable to use Printer(s)! Is the printer(s) properly connected? Check that all signal, and power supply cable connections between Printer and Scanner are OK. Are there print papers left? Insert the printer papers.
  • Page 222 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-9 Printer Troubleshooting (cont’d) Still unable to use the Printer(s)! Is printing unavailable even by manual operation? Check the USB cables Between backplane and usb port. Is there an error message on the monitor display? Replace SOM module.
  • Page 223 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-10 DVD-RW Troubleshooting Unable to record to CD-R or DVD -R! Is a new, unused CD-R disc inserted to the drive? Insert a new, un-used CD-R disc. Are CD-Rs rated at the proper write...
  • Page 224 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-11 Network Troubleshooting 7-9-11-1 No Connection to the Network at All 1.) Check that the network cable between the scanner and the wall network is connected and well seated in both ends.
  • Page 225 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 7-9-11 Network Troubleshooting (cont’d) No Connection to the Network Is the TCP/IP Cable between LOGIQ™ P6/P6 Pro and network outlet connected and well Connect the cable and verify that it is well seated in both ends.
  • Page 226 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION This page was intentionally left blank. Page 7-42 Section 7-1 - Overview...
  • Page 227 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems. Section Description Page Number Overview...
  • Page 228 1 month. All patient information will be permanently deleted at that point. If PHI(Patient Healthcare Information) data needs to be sent to GE employees for service purposes, GE will ascertain agreement from the customer. The patient information shall only be transferred by...
  • Page 229 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-1 17 Inch monitor cover set 8-2-1-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-1-2 Removal procedure 1.) Remove the 17" LCD cable cover.
  • Page 230 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-1-2 Removal procedure (cont’d) 3.) Disconnect power cable and DVI cable. Figure 8-3 Disconnecting power cable and DVI cable 4.) Unscrew 2 screws(5177684, HSH M5x20 WHT) to separate 17" LCD monitor.
  • Page 231 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-1-2 Removal procedure (cont’d) 6.) Unscrew 2 screws to remove the OSD key assy and disconnect the connect. Figure 8-6 Separating the OSD key assy 7.) Separate the 17" LCD tact knob from the OSD key assy.
  • Page 232 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-1-2 Removal procedure (cont’d) 8.) Unscrew 6 screws to remove 17" LCD front cover. Figure 8-8 Removing the 17" LCD front cover 9.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 233 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-2 OSD button 8-2-2-1 Tools • L wrench • Phillips screw driver 8-2-2-2 Preparation Shut down the system and switch off the main breaker 8-2-2-3 17" LCDMON OSD Key Assy Replace Procedure 1.) Remove the LCD Cable cover.
  • Page 234 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-2-3 17" LCDMON OSD Key Assy Replace Procedure (cont’d) 3.) Take the rear cover off using your hands. Figure 8-11 Rear Cover 4.) Unscrew 2 screws and remove OSD Key Assy from the front cover.
  • Page 235 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-2-3 17" LCDMON OSD Key Assy Replace Procedure (cont’d) 6.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-3 Functional Tests...
  • Page 236 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-3 17 INCH LCDMON ASSY 8-2-3-1 Tools • L wrench 8-2-3-2 Preparation • Shut down the system and switch off the main breaker 8-2-3-3 17" LCDMON ASSY Replace Procedure 1.) Remove the LCD Cable cover.
  • Page 237 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-3-3 17" LCDMON ASSY Replace Procedure (cont’d) 4.) Unscrew 2 screw (GEPN:5177684, HSH M5X20WHT) and Remove Spring(GEPN:5177655). Figure 8-16 Removing Spring 5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 238 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4 New Articulation arm with cover installation Procedure 8-2-4-1 Tools • Common Phillips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-4-2 Preparations •...
  • Page 239 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-3 Removal procedure (cont’d) 3.) Disconnect power cable and DVI cable. Figure 8-19 Disconnecting power cable and DVI cable 4.) Unscrew 1 screw (2329677, TAP M4X16) to remove the Fold cover.
  • Page 240 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-3 Removal procedure (cont’d) 6.) Unscrew 2 screws (5177684, HSH M5x20 WHT) to separate the LCD MONITOR and 1 screw (2159625, PH M4x8 W/SP) to remove the Cable bracket.
  • Page 241 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-3 Removal procedure (cont’d) 9.) Unscrew 4 screws (2159634, BH M4x10 WHT) to remove the Arm bottom cover. Figure 8-25 Removing the Arm bottom cover 10.)Unscrew 1 screw (2329677, TAP M4X16) to remove the Lock cover.
  • Page 242 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-3 Removal procedure (cont’d) 12.)Unscrew 2 screws (2159626, PH M4X20) to remove the Arm stopper. Figure 8-28 Removing the Arm stopper 13.) Pull out power cable & DVI cable. Make sure flexible arm not to rotate.
  • Page 243 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-3 Removal procedure (cont’d) 15.)Unscrew 3 screws (2159634, BH M4x10 WHT) after removing the rubber caps to remove neck front cover, neck rear cover. Figure 8-31 Unscrewing 3 screws and removing the neck front cover & neck rear cover 16.)Unscrew 3 screws(5327646, HSH M6X16 WHT) to separate the flexible arm neck pipe.
  • Page 244 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-3 Removal procedure (cont’d) 17.)Unscrew 1 screw (2159625, PH M4x8 W/SP) to remove the OP bracket. Figure 8-33 Removing OP bracket 18.) Unscrew 3 screws (2373562, FH M4x10 YEL) to separate the Flexible arm stopper (A+B).
  • Page 245 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-3 Removal procedure (cont’d) 19.)Pull out power cable & DVI cable and remove the oilless bush after separating the neck pipe. Figure 8-35 Pulling out power cable & DVI cable and removing the stopper...
  • Page 246 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-3 Removal procedure (cont’d) 20.)Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-5 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 247 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-4 Assembly procedure 1.) Insert two of the flexible arm stoppers (A+B) to the Power Cable and DVI cable in the Order illustrated on the figure below. Make sure that chamfered shape of the flexible Arm stopper should face inside.
  • Page 248 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-4 Assembly procedure (cont’d) 4.) Screw 3 screws (2373562, FH M4X10 YEL) to fix the flexible arm stopper after adjusting the holes. Figure 8-39 Fixing the flexible arm stopper 5.) Screw 1 screw (2159625, PH M4X8 W/SP) to install OP bracket.
  • Page 249 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-4 Assembly procedure (cont’d) 7.) Screw 2 screws (5177684, HSH M5X20 WHT) to assemble the LCD monitor with flexible arm Top part. Figure 8-42 Screwing 2 screws 8.) Insert the two cables inside the Flexible arm and Screw 4 screws (2159625, PH M4X8 W/SP) to assemble the Lower arm cable bracket.
  • Page 250 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-4 Assembly procedure (cont’d) 10.)Screw 2 screws (2159626, PH M4X20) to assemble the Arm stopper. Figure 8-45 Installing the Arm stopper 11.)Screw 2 screws (2159625, PH M4X8 W/SP) to assemble the Lower arm guide bracket.
  • Page 251 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-4 Assembly procedure (cont’d) 13.)Place the cables inside the slot and connect power cable and DVI cable. Figure 8-48 Placing the Cables 14.)Assemble the Cable bracket with 1 Screw (2159625, PH M4X8 W/SP) Figure 8-49 Assembling the Cable bracket 15.)Screw 2 screws (2159634, BH M4X10 WHT) to assemble the Cable clamp.
  • Page 252 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-4 Assembly procedure (cont’d) 16.)Place the cables like the figure below. Figure 8-51 Placing the cables 17.)Screw 1 screw (2306565, BH M4X16 WHT) to fix the Power cable.
  • Page 253 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-4 Assembly procedure (cont’d) 18.)Assemble the LCD cable cover. Figure 8-53 Assembling the LCD cable cover Chapter 8 - Page 8-27...
  • Page 254 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-4 Assembly procedure (cont’d) 19.)Screw 3 screws (2159634, BH M4x10 WHT) after placing the neck front cover and neck Rear cover. Figure 8-54 Assembling the neck front cover and the neck rear cover 20.)Screw 1 screw (2159625, PH M4X8 W/SP) to assemble the Lock Knob.
  • Page 255 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-4 Assembly procedure (cont’d) 21.)Screw 1 screw (2329677, TAP M4X16) to assemble the Lock Cover. Figure 8-56 Assembling the Lock Cover Chapter 8 - Page 8-29...
  • Page 256 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-4 Assembly procedure (cont’d) 22.)Screw 2 screws (2159632, BH M4X6 WHT) to assemble the Cam Cover L & R. Figure 8-57 Assembling the Cam Cover L & R 23.)Screw 1 screw (2329677, TAP M4X16) to assemble the Fold Cover.
  • Page 257 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-4 Assembly procedure (cont’d) 24.)Screw 2 screws (2337572, FH M3X6 WHT) to assemble the Arm cover L & R. Figure 8-59 Assembling the Arm Cover L & R 25.)Screw 4 screws (2159634, BH M4X10 WHT) to assemble the Arm bottom Cover.
  • Page 258 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-4-4 Assembly procedure (cont’d) 26.)Attach LCD Arm Caution Label, LCD Arm Rotation Label, LCD Lock Caution Label. LCD Arm Caution Label LCD Arm Rotation Label LCD Lock Caution Label...
  • Page 259 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-5 New Articulation cover set installation Procedure 8-2-5-1 Tools • Common Phillips screwdrivers • Stubby screwdriver (Flat tip and Cross tip) 8-2-5-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 260 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-5-3 Removal procedure (cont’d) 3.) ARM BOTTOM COVER Unscrew 4 screws (2159634, BH M4x10 WHT) to remove the Arm bottom cover. Figure 8-64 Removing the Arm bottom cover 4.) ARM CABLE COVER...
  • Page 261 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-5-3 Removal procedure (cont’d) 5.) ARM LOCK COVER Unscrew 1 screw (2329677, TAP M4x16) to remove the Arm lock cover. Figure 8-66 Removing the Arm lock cover 6.) ARM LOCK KNOB 1.) Unscrew 1 screw (2329677, TAP M4x16) to remove the Arm lock cover.
  • Page 262 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-5-3 Removal procedure (cont’d) 7.) ARM FOLD COVER Unscrew 1 screw (2329677, TAP M4X16) to remove the Arm fold cover. Figure 8-69 Removing the Arm fold cover 8.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 263 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-6 Main Keyboard Assy 8-2-6-1 Tools • Common Phillips screwdrivers 8-2-6-2 Preparations • Shut down the system and switch off the main breaker. 8-2-6-3 Removal procedure 1.) Remove the 3 Probe holders & Cup holder from the keyboard.
  • Page 264 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-6-3 Removal procedure (cont’d) 3.) Unscrew 2 screws from the rear side of the keyboard. Figure 8-72 Unscrews 2 screws 4.) Lift the top cover of the keyboard and disconnect 4 cables from the keyboard.
  • Page 265 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-7 Sub Keyboard Assy 8-2-7-1 Tools • Common Phillips screwdrivers 8-2-7-2 Preparations • Shut down the system and switch off the main breaker. 8-2-7-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the 8-2-6 "Main Keyboard Assy"...
  • Page 266 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-8 Trackball Assy 8-2-8-1 Tools • Common Phillips screwdrivers 8-2-8-2 Preparations • Shut down the system and switch off the main breaker. 8-2-8-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the 8-2-6 "Main Keyboard Assy"...
  • Page 267 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-9 English, Greek, Russian, Swedish, Norwegian A/N Key assy 8-2-9-1 Tools • Common Phillips screwdrivers 8-2-9-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 268 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-9-3 Removal procedure (cont’d) 3.) Unscrew 6 screws from the back side of the A/N key assy. Figure 8-78 Screws on the A/N Key 4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 269 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-10 Freeze key assy 8-2-10-1 Tools • Common Phillips screwdrivers 8-2-10-2 Preparations • Shut down the system and switch off the main breaker. 8-2-10-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the 8-2-6 "Main Keyboard Assy"...
  • Page 270 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-11 TGC key assy / TGC Knob Set 8-2-11-1 Tools • Common Phillips screwdrivers 8-2-11-2 Preparations • Shut down the system and switch off the main breaker. 8-2-11-3 Removal procedure 1.) Remove the CFUNCKEY Assy.
  • Page 271 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-11-3 Removal procedure (cont’d) 4.) Disconnect 1 connector from the PCB. Figure 8-83 TGC Key Connector 5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 272 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-12 Sub Keyboard Encoder Knob Set 8-2-12-1 Tools • Common Phillips screwdrivers 8-2-12-2 Preparations • Shut down the system and switch off the main breaker. 8-2-12-3 Removal procedure 1.) Remove all the Encoder Knobs from the keyboard.
  • Page 273 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-13 Keycap set, Main Keyboard Encoder Knob set 8-2-13-1 Tools • Common Phillips screwdrivers 8-2-13-2 Preparations • Shut down the system and switch off the main breaker. 8-2-13-3 Removal procedure 1.) Remove key caps from the keyboard.
  • Page 274 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-14 Right or Left Side Cover 8-2-14-1 Tools • Common Phillips screwdrivers 8-2-14-2 Preparations • Shut down the system and switch off the main breaker. 8-2-14-3 Removal procedure 1.) Unscrew the four screws on left and right side.
  • Page 275 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-14-3 Removal procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-15 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 276 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-15 Front Cover 8-2-15-1 Tools • Common Phillips screwdrivers 8-2-15-2 Preparations • Shut down the system and switch off the main breaker. 8-2-15-3 Removal procedure 1.) Remove the Left and Right side covers. Refer to the 8-2-14 "Right or Left Side Cover"...
  • Page 277 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-15-3 Removal procedure (cont’d) 4.) Pull the front cover forwards. Figure 8-90 Removal of front cover 5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 278 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-16 Rear Cover 8-2-16-1 Tools • Common Phillips screwdrivers 8-2-16-2 Preparations • Shut down the system, and switch off the main breaker. 8-2-16-3 Removal procedure 1.) Remove the Left and Right side covers. Refer to the 8-2-14 "Right or Left Side Cover"...
  • Page 279 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-17 Middle Cover 8-2-17-1 Tools • Common Phillips screwdrivers • Stubby screwdriver (Flat tip and Cross tip) 8-2-17-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 280 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-18 Top Cover 8-2-18-1 Tools • Common Phillips screwdrivers 8-2-18-2 Preparations • Shut down the system and switch off the main breaker. 8-2-18-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the 8-2-23 "Dummy Cover L / R"...
  • Page 281 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-19 Top Bottom Cover 8-2-19-1 Tools • Common Phillips screwdrivers • Stubby screwdriver (Flat tip and Cross tip) 8-2-19-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 282 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-20 Pole Cover 8-2-20-1 Tools • Common Phillips screwdrivers 8-2-20-2 Preparations • Shut down the system and switch off the main breaker. 8-2-20-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the 8-2-23 "Dummy Cover L / R"...
  • Page 283 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-21 Neck Front 8-2-21-1 Tools • Common Phillips screwdrivers 8-2-21-2 Preparations • Shut down the system and switch off the main breaker. 8-2-21-3 Removal procedure 1.) Remove 1 rubber cap.
  • Page 284 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-21-3 Removal procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-22 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 285 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-22 Neck Rear 8-2-22-1 Tools • Common Phillips screwdrivers 8-2-22-2 Preparations • Shut down the system and switch off the main breaker. 8-2-22-3 Removal procedure 1.) Remove 2 rubber caps (1-2).
  • Page 286 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-22-3 Removal procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-23 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 287 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-23 Dummy Cover L / R 8-2-23-1 Tools • Common Phillips screwdrivers 8-2-23-2 Preparations • Shut down the system and switch off the main breaker. 8-2-23-3 Removal procedure 1.) Remove 2 rubber caps (1-2).
  • Page 288 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-23-3 Removal procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-24 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 289 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-24 EMI Cover L and R 8-2-24-1 Tools • Common Phillips screwdrivers 8-2-24-2 Preparations • Shut down the system and switch off the main breaker. 8-2-24-3 Removal procedure 1.) Remove the Side Left Cover.
  • Page 290 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-25 APS II Assy NOTE: APS Power Supply can not be adjusted. 8-2-25-1 Tools • Common Phillips screwdrivers • Nipper 8-2-25-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 291 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-25-3 Removal procedure (cont’d) 8.) Unplug the J14 connector, and unscrew 2 screws. Pull out the APS II Assy carefully. Note: AC power input connector should be passed through the square hole on chassis bottom.
  • Page 292 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-26 DTRX64II Assy 8-2-26-1 Tools • Common Phillips screwdrivers 8-2-26-2 Preparations • Shut down the system and switch off the main breaker. 8-2-26-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the 8-2-14 "Right or Left Side Cover"...
  • Page 293 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-26-3 Removal procedure (cont’d) 6.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-27 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 294 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-27 P3RLY/PI128II Assy 8-2-27-1 Tools • Common Phillips screwdrivers 8-2-27-2 Preparations • Shut down the system and switch off the main breaker. 8-2-27-3 Removal procedure 1.) Remove the side left and right covers. Refer to the 8-2-14 "Right or Left Side Cover"...
  • Page 295 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-27-3 Removal procedure (cont’d) 6.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-28 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 296 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-28 BP128 Assy 8-2-28-1 Tools • Common Phillips screwdrivers 8-2-28-2 Preparations • Shut down the system and switch off the main breaker. 8-2-28-3 Removal procedure 1.) To replace the BP128 assy, all PWAs should be removed from the nest box.
  • Page 297 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-28-3 Removal procedure (cont’d) 10.) Unscrews 13 screws to remove back plane from nest box. Figure 8-110 BP128 Assy 11.)Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 298 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-29 CWD64 Assy 8-2-29-1 Tools • Common Phillips screwdrivers 8-2-29-2 Preparations • Shut down the system and switch off the main breaker. 8-2-29-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the 8-2-14 "Right or Left Side Cover"...
  • Page 299 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-30 SYSCONML Assy 8-2-30-1 Tools • Common Phillips screwdrivers 8-2-30-2 Preparations • Shut down the system and switch off the main breaker. 8-2-30-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the 8-2-14 "Right or Left Side Cover"...
  • Page 300 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-30-3 Removal procedure (cont’d) Figure 8-114 Separation of ASIR ROM 6.) Mount ASIR ROM to new SYSCONML Assy. Refer to the Figure 8-115. 7.) Confirm mounting direction and pin number.
  • Page 301 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-30-3 Removal procedure (cont’d) 8.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-31 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 302 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-31 SOM Replacement 8-2-31-1 Tools • Common Phillips screwdrivers 8-2-31-2 Preparations • Shut down the system and switch off the main breaker. 8-2-31-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the 8-2-14 "Right or Left Side Cover"...
  • Page 303 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-31-3 Removal procedure (cont’d) 6.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-32 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 304 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-32 RFC Assy 8-2-32-1 Tools • Common Phillips screwdrivers 8-2-32-2 Preparations • Shut down the system and switch off the main breaker. 8-2-32-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the 8-2-14 "Right or Left Side Cover"...
  • Page 305 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-33 Front Caster / Rear Caster NOTE: Both Front Casters and Rear Casters are swivel type in LOGIQ™ P6/P6 Pro. 8-2-33-1 Tools • Common Phillips screwdrivers • Allen/Unbraco wrench...
  • Page 306 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-33-3 Removal procedure (cont’d) 3.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-34 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 307 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-34 Rear Handle 8-2-34-1 Tools • Common Phillips screwdrivers • Allen/Unbraco wrench 8-2-34-2 Preparations • Shut down the system and switch off the main breaker. 8-2-34-3 Removal procedure 1.) Unscrew 1 screw.
  • Page 308 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-35 Rear Hook 8-2-35-1 Tools • Common Phillips screwdrivers 8-2-35-2 Preparations • Shut down the system and switch off the main breaker. 8-2-35-3 Removal procedure 1.) Unscrew 2 screws (1-2) and remove Rear hook.
  • Page 309 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-36 Probe Holder 8-2-36-1 Tools • No need tool 8-2-36-2 Removal procedure 1.) Remove the probe holders from the Keyboard. Figure 8-124 Probe Holders 2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 310 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-37 Cup Holder 8-2-37-1 Tools • No need tool 8-2-37-2 Removal procedure 1.) Remove the cup holder. Figure 8-125 Cup holder 2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
  • Page 311 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-38 ECG Cable Hook 8-2-38-1 Tools • Common Phillips screwdrivers 8-2-38-2 Removal procedure 1.) Unscrew 2 screws (1-2) to remove the ECG Cable Hook. Figure 8-126 Cable hook on the bottom of keyboard 2.) Perform the following functional tests.
  • Page 312 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-39 Air Filter Set 8-2-39-1 Tools • No need tool 8-2-39-2 Preparations • Shut down the system and switch off the main breaker. 8-2-39-3 Removal procedure 1.) Pull out the 2 Air filters from the bottom left and right side of the trans box.
  • Page 313 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-40 DVD R/W Drive 8-2-40-1 Tools • Common Phillips screwdrivers • Clock screw driver 8-2-40-2 Preparations • Shut down the system and switch off the main breaker. 8-2-40-3 Removal procedure 1.) Remove the side Left and Right covers.
  • Page 314 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-40-3 Removal procedure (cont’d) 7.) Unscrew 8 screws to remove the bracket from the DVD Assy. Figure 8-130 Unscrew 8screws and remove the DVD Drive from the bracket 8.) Perform the following functional tests.
  • Page 315 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-41 NEST FAN Assy 8-2-41-1 Tools • Common Phillips screwdrivers 8-2-41-2 Preparations • Shut down the system and switch off the main breaker. 8-2-41-3 Removal procedure 1.) Remove the Left and Right Covers. Refer to the 8-2-14 "Right or Left Side Cover"...
  • Page 316 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-41-3 Removal procedure (cont’d) 6.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-42 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 317 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-42 ARP II Assy 110V/220V 8-2-42-1 Tools • Common Phillips screwdrivers 8-2-42-2 Preparations • Shut down the system and switch off the main breaker. 8-2-42-3 Removal procedure 1.) Remove side right cover and EMI cover. Refer to the 8-2-14 "Right or Left Side Cover"...
  • Page 318 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-42-3 Removal procedure (cont’d) 5.) Disconnect three (3) USB connectors or 5 USB connectors from the rear panel and 2 AC power connector. The 4D controller USB and ECG USB connector also connect to this rear panel PWA.
  • Page 319 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-43 SATA HDD Assy 8-2-43-1 Tools • Common Phillips screwdrivers • Cutter 8-2-43-2 Preparations • Shut down the system and switch off the main breaker. 8-2-43-3 Removal procedure 1.) Remove the side left cover.
  • Page 320 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-43-3 Removal procedure (cont’d) 5.) Disconnect the SATA Power Cable and SATA Cable from HDD Assy after removing 2ea cable ties. And unscrew 4screws to remove the HDD Fixture bracket.
  • Page 321 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-44 ACPC II Assy 8-2-44-1 Tools • Common Phillips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-44-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 322 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-44-3 Removal procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-45 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 323 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-45 AP950 Assy 8-2-45-1 Tools • Common Phillips screwdrivers • Allen/Unbraco wrench 8-2-45-2 Preparations • Shut down the system and switch off the main breaker. 8-2-45-3 Removal procedure •...
  • Page 324 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-45-3 Removal procedure (cont’d) 5.) Disconnect the four connectors form TRANSBOX in Figure 8-143. Figure 8-143 Connectors 6.) Pull out TRANSBOX forwards. Figure 8-144 Pull out the Transbox assy...
  • Page 325 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-45-3 Removal procedure (cont’d) 7.) Perform the following functional tests. If all are successful, include the debrief script provided below. Table 8-46 Functional Tests Service Manual Section Functional Test / Diagnostic Test...
  • Page 326 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-46 Fuse Set 8-2-46-1 Tools • Common Phillips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-46-2 Preparations • Shut down the system and switch off the main breaker.
  • Page 327 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-47 Read and Write Serial Info 8-2-47-1 How to start serial no read/write program 1.) Plug the service dongle on the usb port on rear panel. 2.) Turn on the system.
  • Page 328 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-47-1 How to start serial no read/write program (cont’d) 5.) If the win XP screen appears, run the window explorer. Move the cursor to bottom of the screen. If the task bas appear, click the explorer icon.
  • Page 329 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-2-47-2 How to read Serial No 1.) Check checkboxes of system boards you want to read or System Serial No checkbox. NOTE: You can read System Serial No, L2SYSCON board, DTRX64II_L board, DTRX64II_R board, PI128II board and CWD board information.
  • Page 330 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 8-3 Software Loading Procedure 8-3-1 General This describes a full system software loading (Base Software Load Image + Arirang Application Load Image) procedure. Use this instruction when:...
  • Page 331 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-3-4 Software Loading Steering Guide Figure 8-153 Flow Chart - Installation of Base and Application CD Chapter 8 - Page 8-105...
  • Page 332 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-3-5 Before Starting Software Loading 8-3-5-1 Saving Connectivity Before starting upgrade, write down the Connectivity Setting for back-up. 1.) Go to Utility > Connectivity and click the TCP/IP tab.
  • Page 333 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-3-5-2 Image Management Guide NOTE: Before upgrade to R3.x.x software, perform recommended image backup procedure. Export/Import Data/Images Between Systems - Use this to copy both patient data and images for specified patient(s) from one system to another.
  • Page 334 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-3-5-2 Image Management Guide (cont’d) 4.) In the patient list at the top of the Patient menu, select the patient(s) you want to export. You can use Windows commands to select more than one patient. To select a consecutive list of patients, click the cursor on the first name, move the cursor to the last name, then press and hold down the Shift+right Set key to select all the names.To select a non-consecutive list of...
  • Page 335 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-3-5-2 Image Management Guide (cont’d) Importing Data To import an exam(s) to another Ultrasound system: 1.) At the other Ultrasound system, insert the CD-ROM/DVD. 2.) Press Patient. Press Exam Data Transfer (located at the upper, left-hand corner of the Patient menu).
  • Page 336 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-3-5-2 Image Management Guide (cont’d) 4.) The [TO] combo box is not active. It displays Local Archive. The [FROM] combo box is active. Select the type of removable media CD/DVD Rom. Then please wait until the patient list is visible.
  • Page 337 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 8-4Software Installation 8-4-1 Installing Base Image Software NOTE: While we do not believe it is necessary, It would not hurt to disconnect the system from the network and remove all transducers.
  • Page 338 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-1 Installing Base Image Software (cont’d) 4.) If the BIOS setting value in Primary IDE Slave is ATAPI CDROM, the 1st Boot Device in Boot menu should be Primary Slave. If the 1st Boot Device is not Primary Slave, please change the setting value using Enter key and Arrow key.
  • Page 339 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-1 Installing Base Image Software (cont’d) Figure 8-167 Boot Device Priority 6.) Insert the BT11 Base Image Software Disk into the CD/DVD ROM drive. 7.) After restarting the system, the following screen will be displayed.
  • Page 340 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-1 Installing Base Image Software (cont’d) 8.) Press Enter key and then the following screen will be shown. (If you want to abort installation procedures, press CTRL+C.) Figure 8-169 Base Image Software Installation Menu 9.) Press A key to install the complete disk.
  • Page 341 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-1 Installing Base Image Software (cont’d) 10.)Approx. twenty(20) minutes later, the following screen appears. Eject the Base Image Software Disk from the CD/DVD ROM drive. Figure 8-170 Installation Completion 11.)After removing the disk from the drive, Press ALT + CTRL + DEL key.
  • Page 342 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-2 Verifying Base Image Software 8-4-2-1 Partition Checks 1.) Reboot the SYSTEM and the Desktop screen will be shown as blow. Figure 8-172 Desktop Screen 2.) Click Start button and Run menu, then type explorer and click OK Figure 8-173 Executing Windows Explorer 3.) Right-click My Computer icon on the Desktop and click Manage.
  • Page 343 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-2-1 Partition Checks (cont’d) 4.) Select Storage -> Disk Management and check that only File System, Status, Capacity and Drive Letter of each partition are same as follows.
  • Page 344 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-2-2 Driver Checks 1.) On the window XP desktop, left click on My Computer then right click on Properties. 2.) The System Properties window appears. Select Hardware > Device Manager.
  • Page 345 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-2-3 USB Printer Checks (USB Printer User Only) Without any newly additional user invention, LOGIQ™ P6/P6 Pro will recognize the below 19 kinds of USB printer Printer Name...
  • Page 346 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-2-3 USB Printer Checks (USB Printer User Only) (cont’d) Printer Port Figure 8-177 USB Port - Rear Panel Connector Figure 8-178 USB Port - OP Printer Bracket Back Panel 5.) Make sure that printer properties are proper.
  • Page 347 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-2-4 DMC Checks (DMC User Only) NOTE: DMC Cable is supplied only as an option. 1.) Check DMC Cable (USB to Serial Bridge). 2.) Plug in the Cable as shown on the right side of the USB port.
  • Page 348 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-3 Installing Application Software NOTE: First time SW installation with peripherals on can cause the BlueScreen. NOTE: When connecting peripherals to the system, make sure to turn off the system first in order to avoid the BlueScreen.
  • Page 349 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-3 Installing Application Software (cont’d) 5.) After copying backup files, the following options appear. (Software version can be changed.) Figure 8-183 Select system 6.) Select the product to be installed (P6 or P6 Pro).
  • Page 350 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-3 Installing Application Software (cont’d) 9.) After the installation is completed the system will power off automatically. Figure 8-185 Installation Complete Message NOTE: Some of systems doesn’t power off automatically. In this case, press and hold power button more than 5 sec.
  • Page 351 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4 Standard Confirmation & Functional Checks Configure the system settings and perform functional checks after installation of software. Items for configuration and functional checks include: Confirmation of the software version...
  • Page 352 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4-2 Setting Computer Name 1.) Click on Utility, then select Connectivity. 2.) Click on TCP/IP tab Figure 8-187 TCPIP tab 3.) Type Computer Name The Computer name is unique. Check the scanner serial number printed on the label located at the NOTICE rear lower side of the scanner.
  • Page 353 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4-3 Setting Printer Preferences NOTE: If the 1-3-5 printer setting back-up is NOT performed, ignore this section. 1.) Insert removable disk or media that the printer setting was saved.
  • Page 354 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4-3 Setting Printer Preferences (cont’d) 5.) Select removable disk or media that the printer setting was saved -> Click “Open”. Figure 8-190 Printer set restore 6.) Click the “OK” until all printer windows are closed.
  • Page 355 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4-3 Setting Printer Preferences (cont’d) 9.) Click “Load”. Figure 8-192 Printer set restore 10.)Select removable disk or media which the printer setting was saved -> click “Open. 11.)Click the “OK” until all printer windows are closed.
  • Page 356 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4-4 Functional Checks for Service Platform (Diagnostic Program) The Service platform uses a web-based user interface (UI) to provide access to common service components. NOTE: After perform diagnostics, shut down and restart to make the system working properly.
  • Page 357 ANUAL IRECTION 8-4-4-4 Functional Checks for Service Platform (Diagnostic Program) (cont’d) 4.) Select GE Service at the “Select User Level” field. 5.) Enter the password for the Service Platform. 6.) Click on Okay. Figure 8-195 Select User Level 7.) Verify that the following screen (Service Platform) is displayed on the monitor.
  • Page 358 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4-5 Setting the VCR (if necessary) 1.) Select Utility > System > Peripheral tab. 2.) Under Video Settings, PAL or NTSC. 3.) Under VCR, select Composite or S-Video.
  • Page 359 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4-6 Verifying the VCR (if necessary) 1.) Press Record button, VCR menu shows up at the bottom of the screen. Figure 8-198 VCR Menu 2.) Click Counter Reset, then Press Record ( ) menu to start recording.
  • Page 360 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4-7 Setting Printer Properties in Application Software (if necessary) 1.) Select Utility -> Connectivity -> Service tab 2.) Select Standard Printer, and Add Figure 8-201 Utility > Connectivity > Service 3.) Under Properties, select desired Printer, and Save.
  • Page 361 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4-8 Print Key Assignment and Checkouts necessary) 1.) Select Utility-> Connectivity-> Button tap. 2.) Choose the desired print button (for example, Print1) under Physical Print Button. Figure 8-203 Physical Print Button 3.) Select Standard Print, and press >...
  • Page 362 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4-9 Regional Setup Procedure for Japanese You can set up the keyboard for Japanese. If you select Japanese, the following items display or may be entered in Japanese.
  • Page 363 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4-9 Regional Setup Procedure for Japanese (cont’d) 3.) Select the Language tab, press Details, under Installed Services add the Japanese-Microsoft IME Standard 2002 ver. 8.1 keyboard, under Default input language select Japanese, press Apply, Press OK.
  • Page 364 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-4-4-10 Product Type Check of the System To verify that the correct version of Application software was installed, check the Product Type of the Application software. Verification steps are listed below.
  • Page 365 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 8-5Full Backup / Restore Procedure 8-5-1 General This describes a Full Backup (Database, Image, User preset, Service Data) procedure. Use this instruction when you need to save the data : •...
  • Page 366 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-5-3 Full Backup Procedure (cont’d) 7.) Press “Backup”. Then the below window pops up. Figure 8-210 Emergency Disk Making Wizard 8.) Insert a USB HDD. NOTE: It might take about 30 seconds for the system to recognize the USB HDD.
  • Page 367 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-5-3 Full Backup Procedure (cont’d) 12.)If USB HDD with enough space is not already plugged in, plug in one.Then press Next Figure 8-212 Storage Information 13.)Press next to continue. Then the below windows appear.
  • Page 368 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-5-4 Restore procedure 1.) From Figure 8-140, press “Restore” button. 2.) Rest of the procedure is the same with “Backup” procedure. The system should be rebooted after finishing of Restore procedures.
  • Page 369 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 8-6Software Reloading 8-6-1 General This describes general Software Reloading of C:\ partition functionality. 8-6-2 Software Reloading Procedure 8-6-2-1 Perform Software Reloading in the Service Platform No media (CD/DVD) is needed to perform this function. This can be done in two ways.
  • Page 370 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-6-2-1 Perform Software Reloading in the Service Platform (cont’d) 3.) Press “Time Zone” Figure 8-216 Time Zone 4.) Register the current Time Zone Current Time Zone Table 8-1 Current Time Zone 5.) Press “Cancel”.
  • Page 371 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-6-2-1 Perform Software Reloading in the Service Platform (cont’d) 9.) Select “Reload System Software” Figure 8-217 Reload system Software 10.)Verify that “Local Archive” is available under Dataflow. 11.)Press “ReLoad”.
  • Page 372 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-6-2-1 Perform Software Reloading in the Service Platform (cont’d) 15.)Setup is completed successfully a.) If you do s/w reloading from the system directly, please shut down the system now.
  • Page 373 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-6-2-1 Perform Software Reloading in the Service Platform (cont’d) 17.)If “Disruptive Mode Status” is disabled, Press “Yes” to enable disruptive mode. For the remote reloading, customer has to agree to the diruptive mode. The following window (Figure 8-221) pops up in the system screen and customer has to press “Yes”...
  • Page 374 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-6-2-1 Perform Software Reloading in the Service Platform (cont’d) 19.)Select “Restart System” Figure 8-223 Restart System 20.)Press “Submit” 21.)Press “Confirm” when confirmation message is displayed. 22.)After a while, the system restarts automatically.
  • Page 375 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-6-2-2 Software Reloading 1.) The Base S/W load starts automatically immediately after boot up Figure 8-224 Reloading C:\ partition from Reloading Ghost Image 2.) After finishing C: partition reloading, the system restarts automatically.
  • Page 376 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-6-2-2 Software Reloading (cont’d) 4.) When installation is finished, the system shuts down automatically. 5.) Press Power Button to start the system. 6.) When system fully boots up after the software reloading, please shutdown system again.
  • Page 377 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-6-2-2 Software Reloading (cont’d) 11.)Press Time Zone Figure 8-227 Time Zone 12.)Change the time zone to the local time zone from Table 1. 13.)Press “Apply” 14.)Press “Date & Time” tab.
  • Page 378 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-6-2-2 Software Reloading (cont’d) 15.)Change the Date and Time correctly if needed. 16.)Press “Apply” 17.)Press “OK” 18.)Turn off the system. 19.)Start the system. 20.)Check if the correct date and time are displayed.
  • Page 379 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 8-7Mechanical Option Installation instruction These service instructions describe the installation procedures of following mechanical options to LOGIQ™ P6/P6 Pro. Chapter 8 - Page 8-153...
  • Page 380 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-1 Endo Probe Holder Installation 1.) Check out the 2 screw holes to be used to install Endo Probe Holder. Figure 8-230 Endo Probe Holder Installation 2.) Install Endo Probe holder on the bottom side of the Cable Arm Hook by screwing 2 screws(2159634, BH M4x10 WHT).
  • Page 381 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-2 Microphone Option fixture Installation 1.) Use 2 holes at the bottom of the keyboard. Figure 8-232 Microphone option holes 2.) Screw 2 screws to fix the Microphone fixture.
  • Page 382 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-3 BT11 BW Printer Option Installation 1.) This section describes the installation procedure of the BW Printer option. 2.) Assemble the BW PRINTER to BW PRINTER FIX BRKT with screwing 4 screws (2306562, FH M3X6).
  • Page 383 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-3 BT11 BW Printer Option Installation (cont’d) 4.) Connect the PRT CABLE ASSY 3 (5168835-2) between BW PRINTER and SYSTEM REAR PANEL. Attach the CABLE MOUNT (5399222) on the REAR COVER and fix the cable as the picture below.
  • Page 384 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-4 Drawer Option installation 1.) Push the drawer from the front to the rear direction. Figure 8-238 Drawer Option Installation 2.) Unscrew 2 existing screws from the middle cover.
  • Page 385 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-5 Cable Arm Hook installation 1.) Cable Arm Hook should be installed in prior to installing Probe Cable Hanger, or Endo Probe Holder. 2.) Unscrew 1 existing screw from the rear side of the keyboard.
  • Page 386 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-6 BW Printer Fixture Keyboard installation 1.) Put loctite 290 on the screw(2159634, BH M4X10 WHT)'s edge. Figure 8-242 BW Printer Fixture Keyboard installation 2.) Fix the BW printer brkt beneath the keyboard by the (2159634, BH M4X10 WHT).
  • Page 387 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-6 BW Printer Fixture Keyboard installation (cont’d) 4.) Slide and place the BW printer with printer plate to the BW printer cover and screw 2 screws (5178673, FH M4x6 WHT) (1-2) to fix it.
  • Page 388 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-6 BW Printer Fixture Keyboard installation (cont’d) 6.) Install the BW Printer rear cover a.) Push the BW Printer rear cover to the rear of BW printer cover.
  • Page 389 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-7 BW printer Fixture Middle installation 1.) Install BW Printer on the top of the base bracket by screwing 4 screws. base bracket Figure 8-248 BW Printer Fixture Middle installation 2.) Unscrew 2 existing screw from the middle cover.
  • Page 390 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-7 BW printer Fixture Middle installation (cont’d) 4.) Connect the USB cable and power cable from the printer to the system. Figure 8-250 BW Printer Fixture Middle installation 5.) Install BW hide bracket to the rear side of the BW printer and screw 2 screws(2159633, BH M4x8...
  • Page 391 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-8 Color Printer Fixture Middle installation - UP23MD Printer 1.) Install color printer on the top of the base bracket by screw 4 screws(2306562, FH M3x6 YEL). Figure 8-253 Color Printer Fixture Middle installation 2.) Unscrew 2 existing screws from the middle cover.
  • Page 392 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-8 Color Printer Fixture Middle installation - UP23MD Printer (cont’d) 4.) Connect the USB cable and Power cable from the color printer to the system as shown in the figure below.
  • Page 393 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-9 Color Printer Fixture Middle installation - UP-D25MD Printer 1.) Install color printer on the top of the base bracket by screw 4 screws(2306562, FH M3x6 YEL). Figure 8-258 Color Printer Fixture Middle installation 2.) Unscrew 2 existing screws from the middle cover.
  • Page 394 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-9 Color Printer Fixture Middle installation - UP-D25MD Printer (cont’d) 4.) Connect the USB cable and Power cable from the color printer to the system as shown in the figure below.
  • Page 395 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-10 Printer VCR DVD Fixture Top installation - UP23MD Printer 1.) Unscrew 2 existing screws from the top cover. Figure 8-263 Printer VCR DVD Fixture installation top 2.) Install Fixture DVD bot bracket on the top cover by screwing 2 screws(2327592, FH M4x14 YEL).
  • Page 396 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-10 Printer VCR DVD Fixture Top installation - UP23MD Printer (cont’d) 4.) Screw 4 screws(2159633, BH M4x8 WHT) on both sides to fix the color printer and DVD with top bracket to the color printer top bracket.
  • Page 397 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-11 Color Printer Fixture Top Installation - UP-D25MD Printer 1.) Unscrew 2 existing screws from the top cover. Figure 8-268 Color Printer Fixture Top Installation 2.) Connect the USB cable and Power cable to the Color Printer.
  • Page 398 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-11 Color Printer Fixture Top Installation - UP-D25MD Printer (cont’d) 4.) Install Top Cable BRKT_UPD25MD by screwing 2 screws(2159633, BH M4X8). Figure 8-271 Color Printer Fixture Top Installation 5.) Install Color Printer to the DVD top Bracket by screwing 4 screws (2306562, FH M3X6).
  • Page 399 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-11 Color Printer Fixture Top Installation - UP-D25MD Printer (cont’d) 7.) Connect the USB cable and Power Cable to the system. 8.) Tie the cables using the tie wrap.
  • Page 400 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-12 Printer VCR DVD fixture Top installation - CP900 Printer 1.) Unscrew 2 existing screws from the top cover. Figure 8-275 Printer VCR DVD Fixture Top installation 2.) Install Fixture VCR bot bracket on the top cover by screwing 2 screws(2306565, BH M4x16 WHT).
  • Page 401 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-12 Printer VCR DVD fixture Top installation - CP900 Printer (cont’d) 4.) Screw 2 screws(2159633, BH M4x8 WHT) on each sides to fix the color printer with base bracket to the color printer top bracket.
  • Page 402 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-12 Printer VCR DVD fixture Top installation - CP900 Printer (cont’d) 7.) Install hide bracket by screwing 3 screws(2159633, BH M4x8 WHT). Figure 8-281 Printer VCR DVD Fixture Top installation 8.) Tie the cable properly using the tie wrap.
  • Page 403 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-13 Printer VCR DVD Fixture Top Installation (MD-3000) 8-7-13-1 VCR Remote Port Configuration 1.) LOGIQ™ P6/P6 Pro control the Mitsubishi MD-3000 VCR through USB. Hence the VCR remote control port must be configured as USB 2.) Open the remote control module in the VCR.
  • Page 404 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-13-1 VCR Remote Port Configuration (cont’d) 4.) Set the DIP switch on the VCR remote control panel. The SW 7 in DIP switch should be set to “ON”...
  • Page 405 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-13-2 Bracket Installation 1.) Unscrew 2 existing screws from the top cover. Figure 8-286 Printer VCR DVD Fixture Top installation 2.) Install Fixture Top bracket on the top cover by screwing 2 screws(2306565, BH M4x16 WHT).
  • Page 406 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-13-2 Bracket Installation (cont’d) 4.) Screw 2 screws(2159633, BH M4x8 WHT) on each side of the system to fix the VCR recorder with base bracket to the VCR top bracket.
  • Page 407 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-13-2 Bracket Installation (cont’d) 7.) Connect the cables (Power Cable, USB cable, S-VHS Cable, and Sound cables) from the VCR to the system. Figure 8-292 Printer VCR DVD Fixture Top installation 8.) Connect the cables to the system on the rear panel.
  • Page 408 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-14 Printer VCR DVD Fixture Top Installation (MD-800) 1.) Unscrew 2 existing screws from the top cover. Figure 8-295 Printer VCR DVD Fixture Top installation (LQ MD800) 2.) Install Fixture DVD bot bracket on the top cover by screwing 2 screws(2327592, FH M4x14 YEL).
  • Page 409 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-14 Printer VCR DVD Fixture Top Installation (MD-800) (cont’d) 4.) Install DVD recorder to the base bracket by screwing 2 screws(2159633, BH M4x8 WHT) on each side. Figure 8-298 Printer VCR DVD Fixture Top installation (LQ MD800) 5.) Connect the USB cable and power cable from the DVD to the system.
  • Page 410 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-14 Printer VCR DVD Fixture Top Installation (MD-800) (cont’d) 7.) Install DVD hide bracket by screwing 2 screws(2159633, BH M4x8). Figure 8-301 Printer VCR DVD Fixture Top installation (LQ MD800)
  • Page 411 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-15 Rear Handle installation 1.) Unscrew 2 screws (1-2) and remove Rear hook. Figure 8-302 Rear Handle Installation 2.) Unscrew 2 screws (1-2) and remove the rear handle bracket.
  • Page 412 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-15 Rear Handle installation (cont’d) 3.) Screw 2 screws(5172296, HSH M8x25) from the back of the rear handle. Figure 8-304 Rear Handle Installation 4.) Screw 1 screw(5178673, FH M4x6 WHT) and Install handle dummy bracket.
  • Page 413 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-16 Additional Task Lamp installation 1.) This is the installation instruction of additional task lamp. Figure 8-306 Additional Task Lamp Installation 2.) Connect the USB connector to the location of light bulb mark.
  • Page 414 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-16 Additional Task Lamp installation (cont’d) 4.) Install task lamp bracket to the rear side of keyboard by screwing 2 screws(5178673, FH M4x6 WHT). Figure 8-309 Additional Task Lamp Installation 5.) Fix the task lamp light as shown on the figure below.
  • Page 415 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-17 Protect Cover Installation 1.) Install the prtective covers in rear and front as shown in Figure 8-312 using the double side adhesive tape. Double side adhesive tape...
  • Page 416 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-18 BT11 Drawer Option Installation 1.) This section describes the installation procedure of the drawer option. 2.) Assemble the BW PRINTER to DRAWER FIX BRKT with screwing 4 screws (2306562, FH M3X6).
  • Page 417 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P IRECTION ERVICE ANUAL 8-7-18 BT11 Drawer Option Installation (cont’d) 4.) Screw 2 screws (2159634, BH M4X10) behind the Assembly. Figure 8-315 BT11 Drawer Option Installation 5.) Screw 2 screws (5176744, BH M4X20) to fix the DRAWER Assy to the system.
  • Page 418 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-18 BT11 Drawer Option Installation (cont’d) 6.) Connect the BW Printer cable and fix the cables with CABLE MOUNT. Refer to the pictures below. Figure 8-317 BT11 Drawer Option Installation 7.) Installation Completion.
  • Page 419 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-19 MTZ PROBE HOLDER Installation 1.) This section describes the installation procedure of the MTZ PROBE HOLDER. 2.) Assemble the MTZ PROBE HOLDER FIX BRKT to MTZ HOLDER with 3 screws (2160212, FH...
  • Page 420 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-20 DVR Fixture Top Installation - SONY DVO-1000MD DVD Recorder 1.) This section describes the installation procedure of SONY DVO-1000MD DVD recorder. 2.) Unscrew 2 existing screw from the TOP COVER.
  • Page 421 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-20 DVR Fixture Top Installation - SONY DVO-1000MD DVD Recorder (cont’d) 4.) Assemble the SONY DVO-1000MD to TOP FIXTURE BRKT by screwing 4 screws (2306562, FH M3X6). Figure 8-324 SONY DVR Fixture Top Installation 5.) Assemble the DVO-1000MD with TOP FIXTURE BRKT to FIXTURE DVD BOTTOM BRKT by...
  • Page 422 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-7-20 DVR Fixture Top Installation - SONY DVO-1000MD DVD Recorder (cont’d) 6.) Connect the cable connector with the reference to the picture below. Figure 8-326 SONY DVR Fixture Top Installation 7.) Assemble the TOP CABLE BRKT behind the SONY DVR to cover the Cables by screwing 2 screws...
  • Page 423 It's not allowed to send the PHI data outside of confidential Patient-Doctor environment without encryption of PHI. If PHI needs to be sent to GE employees for troubleshooting purpose, all PHI should be encrypted or protected from unauthorized access by using PKZIP to compress the files with password. And also it can only be sent via a GE Intranet.
  • Page 424 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 8-8-2 ENCRYPTION PROCEDURE 1.) Open PKZip for windows wizard, click on Use PIZIP options Wizard and click on options to do some settings. Figure 8-329 Open PKZip for windows wizard for setting Options...
  • Page 425 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 2.) Switch to Security tab, make below settings, and then click "OK" Figure 8-330 Setting Security tab for mask passwords 3.) Choose Compress files into a new ZIP file, and then click "Next"...
  • Page 426 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 4.) Set storage location and file name Figure 8-332 Set storage location and file name 5.) Select the files you want to compress, the Add to List, click on Finish...
  • Page 427 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 6.) A dialog box will pop out ask for password entry, enter password, confirm it then click OK. Figure 8-334 Confirm password entry 7.) After finish compress, click Done.
  • Page 428 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION This page was intentionally left blank. Page 8-202 Section 8-8 - PHI(Protecting Health Information) DATA COMMUNICATION...
  • Page 429 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Chapter 9 Renewal Parts Section 9-1Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Renewal Parts for LOGIQ™ P6/P6 Pro. Table 9-1 Contents in Chapter 9...
  • Page 430 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-2List of Abbreviations • Assy - Assembly • Ctrl - Control • FRU 1 - Replacement part available in part hub • FRU 2 - Replacement part available from the manufacturer (lead time involved) •...
  • Page 431 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-3MONITOR Figure 9-1 Monitor Chapter 9 - Page 9-3...
  • Page 432 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-3 MONITOR (cont’d) Item Part Name FRU Number Description R1.0.x R2.0.x~ R3.0.x~ LP6 Monitor cover set 5267673 Common LP6 17” LCDMON Assy 5327227 Common New Articulation Arm with Covers...
  • Page 433 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-4 KEYBOARD Figure 9-2 KEYBOARD BLOCK Chapter 9 - Page 9-5...
  • Page 434 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-4 KEYBOARD (cont’d) Item Part Name FRU Number Description R1.0.x R2.0.x~ R3.0.x~ LP6 Main Keyboard Assy 5252353 Common with LP5 Japanese Keyboard P6 BT09 Main Keyboard Assy for JPN...
  • Page 435 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-5 COVERS Figure 9-3 COVERS Chapter 9 - Page 9-7...
  • Page 436 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-5 COVERS (cont’d) Item Part Name FRU Number Description R1.0.x R2.0.x~ R3.0.x~ LP6 Side Left Cover 5327214 Common with LP5 LP6 Side Right Cover 5327215 Common with LP5...
  • Page 437 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-6 NEST Figure 9-4 PCB BOARDS Chapter 9 - Page 9-9...
  • Page 438 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-6 NEST (cont’d) Item Part Name FRU Number Description R1.0.x R2.0.x~ R3.0.x~ Probe Connection PI128 II Assy 5327228 Board 64Ch TX/RX DTRX64II Assy 5252321-2 Beamforming Board 64Ch TX/RX...
  • Page 439 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-7 BODY 511 512 Figure 9-5 BODY BLOCK Chapter 9 - Page 9-11...
  • Page 440 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-7 BODY (cont’d) Item Part Name FRU Number Description Qty FRU R1.0.x R2.0.x~ R3.0.x~ Front Caster 5144523 Common with LP5 Rear Caster 5144524 Common with LP5 Rear Handle...
  • Page 441 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-8 POWER Figure 9-6 POWER BLOCK Chapter 9 - Page 9-13...
  • Page 442 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-8 POWER (cont’d) Item Part Name FRU Number Description R1.0.x R2.0.x~ R3.0.x~ ACPCII Assy 5257219 Common with LP5 AP950 Transbox Assy 110V 5252288 Common with LP5 AP950 Transbox Assy 220V...
  • Page 443 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-9Cables Figure 9-7 Cable Chapter 9 - Page 9-15...
  • Page 444 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-9 Cables (cont’d) Item Part Name FRU Number Description R1.0.x R2.0.x~ R3.0.x~ Mon Pwr Cable, BW PWR Cable, Speaker cables 1. Mon PWR Cable, Connect one side (Label:MON) to LP6 17"...
  • Page 445 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Item Part Name FRU Number Description R1.0.x R2.0.x~ R3.0.x~ Front USB Cable, MIC Cable 1. MIC Cable, Connect one side (Label:MIC) to LP6 Front USB Bracket and another...
  • Page 446 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Item Part Name FRU Number Description R1.0.x R2.0.x~ R3.0.x~ Connect one side (Label CON5) to Label CON5 of Peripheral AC Power Cable 5165244 AC Out Cable and another...
  • Page 447 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-10 Options Item Part Name FRU Number Description R1.0.x R2.0.x~ R3.0.x~ Endo Probe Holder 5171933 Common with LP5 VCR DVD Fixture Top 5171717-2 Common with LP5 BW Printer Fixture Keyboard...
  • Page 448 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-11Software Item Part Name FRU Number Description LP6 CD SET 5252308-3 R1.0.3 LP6 BT09 CD set 5327218 R2.0.X LP6 BT11 CD SET 5395396 R3.0.X Table 9-11 Software...
  • Page 449 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-12 Probes Item Part Name Part Number R1.0.x R2.0.x~ R3.0.x~ 4C Probe 5123455 5CS Probe 5143934 8C Probe 2348094 E8C Probe 2294641 E8CS Probe 47236865 8L Probe...
  • Page 450 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 9-13Peripherals Item Part Name FRU Number Description Panasonic LQ-MD800P LQ-MD800P 5120592 DVR Video Recorder Panasonic LQ-MD800E LQ-MD800E 5120593 DVR Video Recorder VCR Mitsubishi HS-MD3000U 066E9513 VCR Mitsubishi HS-MD3000E...
  • Page 451 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your LOGIQ™ P6/P6 Pro system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
  • Page 452 Your GE Service Representative has an in-depth knowledge of your LOGIQ™ P6/P6 Pro ultrasound scanning system and can best provide competent, efficient service. Please contact us for coverage information and/or price for service.
  • Page 453 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Table 10-2 Customer Care Schedule Quarterl Per Facilities Service at Indicated Time Daily Monthly QA Program Notes •* Clean Probes * or before each use • Clean Probe Holders •...
  • Page 454 Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs. Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID...
  • Page 455 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal...
  • Page 456 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-4-2 Special Tools, Supplies and Equipment 10-4-2-1 Specific Requirements for Care & Maintenance Table 10-5 Overview of Requirements for Care & Maintenance Tool Part Number Comments Digital Volt Meter (DVM)
  • Page 457 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. refer to the system user documentation whenever necessary. Table 10-6 System Checks...
  • Page 458 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. refer to the system user documentation whenever necessary. 10-5-2-1 System Checks Table 10-7...
  • Page 459 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. refer to the User Manual for a list of approved peripherals/options. Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description Verify record/playback capabilities of the VCR.
  • Page 460 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-5-4-2 Air Filter Cleaning Table 10-11 Air Filter Cleaning - frequency varies with your environment Step Item Description Remove Filter refer to Chapter 8 for air filter location and removal instructions.
  • Page 461 The system user manuals and various probe handling cards provide a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details.
  • Page 462 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 10-6 Electrical Safety Tests 10-6-1 Safety Test Overview The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards.
  • Page 463 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-2 GEMS Leakage Current Limits The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed.
  • Page 464 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
  • Page 465 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-4 Grounding Continuity Electric Shock Hazard. The patient must not be contacted to the equipment during this test CAUTION Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
  • Page 466 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-4-2 Dale 600 - Ground Continuity The Dale 600 measures line cord resistance from the third pin of the attachment plug to the meter’s Chassis Cable clamp. Test the grounding continuity of the system to all exposed metal parts in accordance with the IEC 601-1.1 procedure as above.
  • Page 467 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-5 Chassis Leakage Current Test 10-6-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
  • Page 468 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-5-3 Dale 600 Meter Procedure When measuring system chassis currents with the Dale 600, always use the CHASSIS selection of the external/chassis function switch. This requires the ground clip lead and changing the meters switches in accordance with the IEC 601-1.1.
  • Page 469 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-5-4 Data Sheet for Chassis Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-14. Record all data on the PM Inspection Certificate.
  • Page 470 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-6 Isolated Patient Lead (Source) Leakage–Lead to Ground 10-6-6-1 Definition This test measures the current which would flow to ground from any of the isolated ECG leads. The meter simulates a patient who is connected to the monitoring equipment and is grounded by touching some other grounded surface.
  • Page 471 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-6-3 Dale 600 Meter Procedure The Dale 600 provides five snap type ECG buttons for testing patient leads. Snap on all patient leads to the meter and assure that the ground clip is connected to the system’s ground terminal. Select the meter’s LEAD-GND function.
  • Page 472 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Table 10-19 Testing Power Conditions ECG Power Meter’s Polarity Switch Meter’s Neutral Switch REVERSE OPEN NORM CLOSED NORM OPEN REVERSE CLOSED REVERSE OPEN 7.) Record the patient lead to ground leakage current measured on the data sheet.
  • Page 473 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-7 Isolated Patient Lead (Source) Leakage–Lead to Lead reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter’s function selector to the LEAD-LEAD position. Select and test each of the five ECG lead positions (except ALL) on the LEAD selector, testing each to the power condition combinations found in the table.
  • Page 474 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-8 Isolated Patient Lead (Sink) Leakage-Isolation Test reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter’s function selector to the LEAD-ISO. Select the ALL position on the lead selector. Depress the rocker switch to ISO TEST to test lead isolation.
  • Page 475 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-9 Probe Leakage Current Test 10-6-9-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
  • Page 476 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-9-3Meter Procedure Using Probe Adapter (cont’d) Figure 10-10 Transducer Source Leakage Current Test 6.) Set the meter's “FUNCTION” switch to EXTERNAL position. 7.) Add the saline probe and the imaging area of the probe into the saline bath.
  • Page 477 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-9-4 No Meter Probe Adapter Procedure Figure 10-11 Check Without Probe Adapter Follow these steps to test each transducer for leakage current. 1.) Turn the LOGIQ™ P6/P6 Pro unit OFF.
  • Page 478 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION 10-6-9-5 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-15 Table 10- 16. Record all data on the PM Inspection Certificate.
  • Page 479 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION Section 10-7 When There's Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
  • Page 480 Air Filter Applicable Software Options Probe Holders Applicable Hardware Options External I/O Control Panel Wheels, Brakes & Swivel Locks Monitor Cables and Connectors GE Approved Peripherals (VCR, CD-RW, MOD, Touch Panel Printers) Measurement Accuracy GE Approved Peripherals COMMENTS: ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________...
  • Page 481 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe Chassis Source Leakage Current - Caster...
  • Page 482 GE H EALTHCARE 5394152, Revision 5 LOGIQ™ P6/P6 P ERVICE ANUAL IRECTION This page was intentionally left blank. Page 10-32 Section 10-6 - Electrical Safety Tests...
  • Page 484 2011, General Electric Company. © GE Medical Systems, a General Electric Company, going to market as GE Healthcare. 3200 N. Grandview Boulevard Waukesha, Wisconsin 53149 www.gehealthcare.com...

Table of Contents