Figure 11 Terminal Board Replacement - Honeywell OneWireless XYR 6000 User Manual

Temperature and temperature/discrete input transmitter
Hide thumbs Also See for OneWireless XYR 6000:
Table of Contents

Advertisement

Step
Disconnect field wiring from terminal board and label it to ease reconnection.
6
Remove and save the 3 screws that attach the terminal board to the housing. Take care
7
because these screws are not captive. See item 4 in Figure 11.
Remove terminal board by disconnecting cable from back of the terminal board. Do not pull on
8
the wires or you could damage them. Instead, depress the latch while pulling on the connector.
Attach connector to new terminal board. Observe correct polarity of the connector. Verify that
9
the cable is latched to the terminal board.
Fasten terminal board with screws from step 7.
10
Re-connect field wiring.
11
Insert the battery holder into the transmitter. Reattach the screws and tighten to 0,4 – 0,6 N-M
12
(3.5 – 5.3 Lb-in).
Re-connect battery connector.
Honeywell recommends lubricating the end cap O-ring with a Silicone Grease such as Dow
Corning #55 or equivalent before replacing the end cap.
Screw the end cap back on and tighten the M3 locking screw.
13
Rev. 3
OneWireless XYR 6000 Temperature and Temperature/Discrete Input Transmitter User's Manual
6/24/08
4

Figure 11 Terminal board replacement

6.5. Replacing terminal board
Action
3
2
6. Maintenance/Repair
1
45

Advertisement

Table of Contents
loading

Table of Contents