Honeywell Sensepoint XRL Instruction Manual page 60

Fixed gas detector
Hide thumbs Also See for Sensepoint XRL:
Table of Contents

Advertisement

Modbus Register Address Information
40163
Relay On Delay Time
40164
Relay Off Delay Time
40165 to 40166
User configured full scale range
40167 to 40168
Alarm Threshold 1
40169 to 40170
Alarm Threshold 2
40171 to 40172
Target Calibration Gas Concentration
40173
Gas Type
40174 to 40175
Voltage Output
40176 to 40183
Device Serial Number
40184 to 40188
BLE FW Version
40189 to 40200
Reserved
40201
Reset Alarms & Faults
40202 to 40207
Reserved
40208 to 40213
Raw A/D Counts
R/W
Type
Size Description
R
u16
1
R
u16
1
R
f32
2
R
f32
2
R
f32
2
R
f32
2
R
u16
1
R
f32
2
R
string[16]
8
R
string[10]
5
BLE FW Version
W
u16
1
R
u16[6]
6
Notes
Display Range (Full Scale)
0x00: Flammable
0x01: Toxic
0x02: O2
0x03: VOC

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents