International Approvals
8 International Approvals
CE Conformity:
The OM-DAQXL has been designed to meet requirements as outlined
in European Community EMC Directive EN61326. Five ferrite cores
(PN: OM-DAQXL-RF) are included in the OM-DAQXL package. When
using thermocouple temperature sensors with the OM-DAQXL it is
recommended to pass the thermocouple through the supplied ferrite
2 turns before connecting it to the device.
When communicating with the OM-DAQXL via Ethernet it is recommended
to pass the Ethernet cable through a ferrite 2 turns before connecting it to the
OM-DAQXL. It is not recommended to have the AC adapter plugged in during
normal operation. The AC adapter is intended to charge the battery only.
Emissions:
EN 61326-1 Class A
Immunity:
EN 61000-4-3
EN 61000-4-4
EN 61000-4-6