Three Wire Rtd Wiring; Four Wire Rtd Wiring - Omega OM-DAQXL-1 Series User Manual

Multi-channel universal input touch screen data logger
Table of Contents

Advertisement

2
Hardware
2.6.2.2 Three-Wire RTD Wiring
For 3-wire RTD sensors the second common lead must be connected to the
negative terminal of the next consecutive even channel.
Figure 2-14 Three-Wire RTD Wiring Example
2.6.2.3 Four-Wire RTD Wiring
For 4-wire RTD sensors the data logger performs a Kelvin resistance
measurement using 1 channel to supply excitation voltage and the next channel
to sense the voltage across the RTD.
Figure 2-15 Four-Wire RTD Wiring Example
2-8